Journal of Clinical Microbiology, February 2008, p. 834, Vol. 46, No. 2
0095-1137/08/$08.00+0     doi:10.1128/JCM.02144-07
Copyright © 2008, American Society for Microbiology. All Rights Reserved.

LETTER TO THE EDITOR

Plasma Load Discrepancies between the Roche Cobas Amplicor Human Immunodeficiency Virus Type 1 (HIV-1) Monitor Version 1.5 and Roche Cobas AmpliPrep/Cobas TaqMan HIV-1 Assays


arrow
LETTER
 
We read with interest the findings of Damond et al. (1) showing the lack of agreement between plasma viral load measurements performed by the Cobas Amplicor human immunodeficiency virus type 1 (HIV-1) Monitor test, version 1.5 (Roche Molecular Systems, Inc., Branchburg, NJ) and the Cobas AmpliPrep/Cobas TaqMan HIV-1 test (Roche Molecular Systems, Inc.). The authors found generally good overall agreement between the assays, with mean values of 2.99 and 2.51 log10 copies/ml for the Cobas Amplicor HIV-1 Monitor test, version 1.5, and the Cobas AmpliPrep/Cobas TaqMan HIV-1 test, respectively, but substantial differences were noted between the results for some specimens. The reasons for the lack of result agreement in these specimens remain unclear.

Unfortunately, the authors included only a brief statement that routine clinical testing in their laboratory was performed with the Cobas Amplicor HIV-1 Monitor test, version 1.5, following automated nucleic acid extraction using a "MagNA Pure large-volume kit." One wonders if the authors were referring specifically to the MagNA Pure LC instrument (Roche Diagnostics, Indianapolis, IN) and the MagNA Pure LC total nucleic acid isolation kit-large volume (Roche Diagnostics) and using them in conjunction with the Cobas Amplicor assay for (i) routine clinical testing and (ii) analyzing the 160 plasma specimens included in their study. These procedural details are unclear from the authors' description of their routine clinical testing and comparative study methods (1).

There are only two published papers describing the use of automated nucleic acid extraction methods utilizing the MagNA Pure LC instrument in conjunction with either the Cobas Amplicor HIV-1 Monitor test, version 1.5, or the Amplicor HIV-1 Monitor test, version 1.5 (2, 3). One wonders if Damond et al. have adopted or modified either of these methods for use with the Cobas Amplicor assay or validated their own method. Readers would benefit from the authors' provision of appropriate, specific literature references along with additional details of their specific methods used for performing the Cobas Amplicor assay. One can accurately interpret the study findings and conclusions only after being assured that the authors' modifications to the Cobas Amplicor assay procedure did not contribute to the significant discrepancies in HIV-1 RNA levels observed in this study. These procedural details are critical to understanding the implications of the authors' findings for diagnostic laboratories that are considering the use of the Cobas AmpliPrep/Cobas TaqMan HIV-1 test for routine clinical testing.


arrow
REFERENCES
 
    1
  1. Damond, F., B. Roquebert, A. Bénard, G. Collin, M. Miceli, P. Yéni, F. Brun-Vezinet, and D. Descamps. 2007. Human immunodeficiency virus type 1 (HIV-1) plasma load discrepancies between the Roche COBAS AMPLICOR HIV-1 MONITOR version 1.5 and the Roche COBAS AmpliPrep/COBAS TaqMan HIV-1 assays. J. Clin. Microbiol. 45:3436-3438.[Abstract/Free Full Text]
  2. 2
  3. Hölzl, G., M. Stöcher, V. Leb, H. Stekel, and J. Berg. 2003. Entirely automated quantification of human immunodeficiency virus type 1 (HIV-1) RNA in plasma by using the ultrasensitive COBAS AMPLICOR HIV-1 Monitor test and RNA purification on the MagNA Pure LC instrument. J. Clin. Microbiol. 41:1248-1251.[Abstract/Free Full Text]
  4. 3
  5. Lee, B. G., K. R. Fiebelkorn, A. M. Caliendo, and F. S. Nolte. 2003. Development and verification of an automated sample processing protocol for quantitation of human immunodeficiency virus type 1 RNA in plasma. J. Clin. Microbiol. 41:2062-2067.[Abstract/Free Full Text]
Joseph D. C. Yao*
Jeffrey J. Germer

Division of Clinical Microbiology
Mayo Clinic, Su 1-602
200 First Street SW
Rochester, Minnesota 55905

* Phone: (507) 538-1934, Fax: (507) 266-4341, E-mail: jdcyao{at}mayo.edu


Authors' Reply


arrow
LETTER 
 
In response to the commentary of Dr. Yao and colleague regarding our article "Human Immunodeficiency Virus Type 1 (HIV-1) Plasma Load Discrepancies between the Roche COBAS AMPLICOR HIV-1 MONITOR Version 1.5 and the Roche COBAS AmpliPrep/COBAS TaqMan HIV-1 Assays," we would like to clarify the procedural details of our viral load clinical testing.

Routine clinical viral load testing of the 160 plasma specimens for the comparative study was performed using the Cobas Amplicor HIV-1 Monitor test, version 1.5, following automated nucleic acid extraction using the MagNA Pure LC total nucleic acid isolation kit (large volume) (Roche Diagnostics, Meylan, France). This automated extraction was performed using the MagNA Pure LC instrument (Roche Diagnostics, Indianapolis, IN).

The automated nucleic acid extraction method with the MagNA Pure LC total nucleic acid isolation kit (large volume) (Roche Diagnostics) in conjunction with the Cobas Amplicor HIV-1 Monitor test, version 1.5, is similar to the procedure described by Lee et al. according to Roche Diagnostics instructions (Roche Diagnostics, Meylan, France) (1).

We hope that these details clarify the methods section of our paper.


arrow
REFERENCES 
 
    1
  1. Lee, B. G., K. R. Fiebelkorn, A. M. Caliendo, and F. S. Nolte. 2003. Development and verification of an automated sample processing protocol for quantitation of human immunodeficiency virus type 1 RNA in plasma. J. Clin. Microbiol. 41:2062-2067.[Abstract/Free Full Text]
F. Damond*
B. Roquebert
D. Descamps

Laboratoire de Virologie
Hôpital Bichat Claude Bernard
75018 Paris, France

* Phone: 33 1 40 25 61 51 Fax: 33 1 40 25 67 69 E-mail: florence.damond{at}bch.aphp.fr


Journal of Clinical Microbiology, February 2008, p. 834, Vol. 46, No. 2
0095-1137/08/$08.00+0     doi:10.1128/JCM.02144-07
Copyright © 2008, American Society for Microbiology. All Rights Reserved.





This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowReprints and Permissions
Right arrow Copyright Information
Right arrow Books from ASM Press
Right arrow MicrobeWorld
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Yao, J. D. C.
Right arrow Articles by Descamps, D.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Yao, J. D. C.
Right arrow Articles by Descamps, D.