Journal of Clinical Microbiology, March 2008, p. 1050-1059, Vol. 46, No. 3
0095-1137/08/$08.00+0 doi:10.1128/JCM.02227-07
Copyright © 2008, American Society for Microbiology. All Rights Reserved.

Research Program Infection and Cancer, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 242, 69120 Heidelberg, Germany
Received 19 November 2007/ Returned for modification 12 December 2007/ Accepted 4 January 2008
|
|
|---|
1,000 plasmid copies in the presence of 100 ng cellular DNA was obtained. Additionally, we integrated an internal β-globin PCR into both HPV PCR systems, allowing simultaneous DNA quality control without affecting the sensitivity of HPV detection. Furthermore, we describe five additional low-risk HPV probes used in multiplex HPV genotyping (MPG) for simultaneous identification of all 15 high-risk, 3 putative high-risk, and 9 low-risk HPV genotypes. The performance of BSGP5+/6+ multiplexed with β-globin primers was compared to that of standard GP5+/6+ with DNA from 1,112 cervical scrapings. There was 79% overall agreement (kappa = 0.816). BSGP5+/6+ was significantly more sensitive than GP5+/6+ for detection of HPV 30, 39, 42, 44, 51, 52, 53, 68, 73, and 82, detecting 212 additional HPV infections and increasing the proportion of multiple infections from 17.2 to 26.9% in cancer patients. In conclusion, BSGP5+/6+ multiplexed with β-globin PCR provides an improvement in type-specific amplification sensitivity and homogeneity compared to GP5+/6+ and offers simultaneous internal control of DNA quality. BSGP5+/6+-MPG, therefore, is suitable for epidemiologic and also diagnostic applications. |
|
|---|
At present, there are several PCR-based HPV detection methods; one of the most commonly applied uses the GP5+/6+ (GP) primer set (18), which targets conserved sequences within the L1 region of the virus genome flanking highly variable type-specific sequences, allowing detection of a broad range of mucosal HPV types. The HPV genotype can be determined by analysis of the generated PCR product by sequencing, restriction fragment length polymorphism, or hybridization with type-specific probes.
Since the design of GP primers (5), many additional mucosal HPV types have been discovered. It has been shown that the analytic sensitivity of GP PCR for individual genital HPV types differs depending on the number and position of primer mismatches (5). Furthermore, GP PCR detects multiple infections to a lesser extent than other consensus or broad-spectrum PCR systems (9, 13). The selective underamplification of certain HPV types, such as HPV 53, and/or the reduced sensitivity in multiple HPV infections may lead to an underestimation of the prevalence of specific types (1, 2, 13).
Control of template DNA quality is important in the analysis of clinical samples to pinpoint false-negative HPV results. GP PCR is mostly applied to samples prescreened by external β-globin PCR and subsequent PCR product analysis by gel electrophoresis (5, 8). Integration of an internal DNA quality control in a multiplexed fashion has been described for MY09/11 and PGMY09/11 PCR but may decrease the analytic sensitivity for HPV (7). Therefore, an internal DNA control without impairment of the HPV amplification sensitivity is needed not only to reduce assay time and costs but also to provide an internal PCR control for monitoring a failure of the HPV PCR.
Recently, we described a multiplex HPV genotyping assay (MPG) based on GP PCR followed by subsequent detection of the biotinylated products with 22 type-specific oligonucleotide probes covalently coupled to distinct sets of fluorescence-labeled polystyrene beads (Luminex Technology) (17). Meanwhile, MPG allows simultaneous semiquantitative high-throughput analysis for 15 HR, 3 pHR, and 9 LR mucosal HPV genotypes.
To reach a homogeneous analytic sensitivity for all genital HPV types, we designed the novel BSGP5+/6+ (BS) primer set on the basis of mismatch reduction to 48 genital HPV types. Novel β-globin primers were integrated in the BS PCR as well as the standard GP PCR for internal DNA quality control. We compared both PCR-MPG systems with respect to the amplification of 27 genital HPV types in 1,017 cervical samples from a Mongolian population-based study (B. Dondog, G. M. Clifford, S. Vaccarella, T. Waterboer, D. Unurjargal, D. Avirmed, S. Enkhtuya, F. Kommoss, N. Wentzensen, P. J. F. Snijders, C. J. L. M. Meijer, S. Franceschi, and M. Pawlita, submitted for publication) and 95 cervical cancer patients.
|
|
|---|
|
View this table: [in a new window] |
TABLE 1. BSGP5+/6+ primer sequences
|
Plasmid clones. Analytic sensitivity of the BS primers was determined for plasmid clones of HPV types 6, 11, 16, 18, 26, 31, 33, 35, 39, 42, 43, 44, 45, 51, 52, 53, 56, 58, 59, 66, 68, 70, 73, and 82 (Table 2). Plasmid preparations were quantified using NanoDrop ND-1000 (NanoDrop Technologies, Wilmington, DE) or a Hitachi U-1100 spectrophotometer (Hitachi Ltd., Tokyo, Japan). Copy numbers were determined on the basis of the molecular weights of the plasmids. Tenfold endpoint dilution series were prepared in 100 ng/µl of human placenta (HP) DNA in a total volume of 30 µl. Two or three replicates of each dilution were assayed independently.
|
View this table: [in a new window] |
TABLE 2. HPV type detection limits
|
bacteria, transformed by high-copy-number plasmids containing the viral genome, replaced the template DNA. MPG probe design. Amino-modified oligonucleotide probes for HPV 30 (5'-CAC ACA AAC GTT ATC CAC A-3'), 67 (5'-GGA AAA ATC AGA GGC TAC A-3'), and 69 (5'-CAT CTG CCA CTT TTA AAC C-3') were newly designed as described previously (17). Probe sequences for HPV 26 (5'-GTA CAT TAT CTG CAG CAT C-3') and 53 (5'-TGT CTA CAT ATA ATT CAA AGC-3') are described elsewhere (19). For specificity evaluation, cloned HPV genomes were used (for accession numbers, see Table 3).
|
View this table: [in a new window] |
TABLE 3. BSGP5+/6+ and GP5+/6+ sequence alignmentsa
|
The β-globin probe (5'-CTT CTT TTA ATA TAC TTT TTT GTT-3') was designed for the MS3/MS10 PCR product using the reference sequence AY260740 and following the criteria for HPV type-specific probe design (17).
DNA isolation from cervical scrapings. For DNA isolation, 2.0 ml (from a total of 20 ml) of cervical scrapings collected with a cytobrush (Cervex-Brush, Rovers Medical Devices B.V., Oss, The Netherlands) in PreserveCyt solution (Cytyc Corp., Boxborough, MA) were purified with a Roche High Pure PCR template preparation kit according to the manufacturer's instructions. DNA was eluted in 0.2 ml of elution buffer (10 mM Tris, pH 8.5) and stored at –20°C until further use. Scrapes were kept 2 months at 4°C before DNA extraction.
Oligonucleotide synthesis. All primers and probes were purchased from MWG-Biotech AG, Ebersberg, Germany.
GP5+/6+ PCR. GP PCR was performed as previously described (5) with some modifications. Briefly, 10 µl of DNA extracted from cervical scrapings or 1 µl of HPV plasmid dilution was amplified in 50 µl containing 50 mM KCl, 0.8 g/liter Nonidet P40, 10 mM Tris HCl (pH 8.8) (10x PCR buffer; MBI Fermentas GmbH, St. Leon Roth, Germany), 200 µM of each deoxynucleoside triphosphate, 3.5 mM MgCl2 (Biozym Scientific GmbH, Hessisch Oldendorf, Germany), 1 U of DNA AmpliTaq polymerase (Roche Applied Biosystems, Mannheim, Germany) and 500 nM each of the GP5+ and 5'-biotinylated GP6+ primers. In the case of the integrated β-globin-GP5+/6+ PCR, 100 nM each of the β-globin primers MS3 and 5'biotinylated MS10 were added to the PCR mixture. A 4-min denaturation step at 94°C was followed by 40 cycles of amplification with a PCR thermocycler (Gene Amp PCR system 2400; Perkin-Elmer, Wellesley, MA) or a Mastercycler (Eppendorf). Each cycle included a denaturation step at 94°C for 20 s, an annealing step at 38°C for 30 s, and an elongation step at 71°C for 80 s. The final elongation step was prolonged for 4 min further. Ramping rates for the Mastercycler were adjusted as described recently (18): 1.8°C/s from 94°C to 38°C, 2.0°C/s from 38°C to 71°C, and 2,8°C/s from 71°C to 94°C. Each PCR experiment included samples with reference plasmids as positive controls and several samples lacking template DNA as contamination controls.
BSGP5+/6+ PCR. For the initial BS PCR, eight additional forward (BSGP5+-2 to -9) and two additional 5'-biotinylated reverse (BSGP6+-b and -c) primers were added to the GP PCR (Table 1). Two hundred nanomolar of each forward primer (including GP5+), 400 nM of each reverse primer (including GP6+), and 300 nM each of the β-globin primers MS3 and 5'-biotinylated MS10 were used. Otherwise, the PCR buffers, reagents, and temperature profiles were identical to those described above.
Coupling of oligonucleotide probes. 5'-Amino-modified C-12-linked oligonucleotide probes were coupled to carboxylated beads (xMAP; Luminex Corp., Austin, TX) by a carbodiimide-based coupling procedure as described elsewhere (17).
MPG. Following PCR amplification, 10 µl of each reaction was analyzed by MPG as described elsewhere (17) with some modifications. Instead of casein, 0.02% Tween 20 was added to the wash buffer, and casein was also omitted from the 2.0 M tetramethylammonium chloride staining buffer.
Briefly, GP PCR products were generated, denatured, and hybridized to the bead-coupled probes in 96-well plates, allowing PCR products from 96 samples to be processed in parallel. After transfer into wash plates with filter bottoms, unhybridized DNA was removed. Subsequently, biotinylated PCR products were stained by streptavidin-R-phycoerythrin conjugate. After further washing steps, beads were analyzed in a Luminex reader (Luminex Corp.), which contains two lasers to identify the bead set by the internal bead color and to quantify the reporter fluorescence on the bead. The result was expressed as the median fluorescence intensity (MFI) of at least 100 beads per set.
Cutoff definition and statistics. For each probe, MFI values in reactions with no PCR product added to the hybridization mixture were considered background values. Net MFI values were computed by subtraction of 1.1 times the median background value. For all probes, this cutoff value was above the mean background plus three times the standard deviation. Reactions with net MFI values above 5 were defined as positive reactions. Correlation between the two PCR methods was assessed using kappa statistics. The coefficient of variation (CV) was computed to describe assay reproducibility. Fisher's exact test was used to compare dichotomous variables between two groups. P values of <0.05 were considered statistically significant.
|
|
|---|
To assess the efficacy of the novel BS PCR in comparison to GP, plasmids containing HPV L1 sequences from 27 different genotypes were analyzed by both methods. PCR products of the expected size of about 150 bp were obtained from all plasmids, as determined by gel electrophoresis and subsequent genotyping by MPG (data not shown).
The analytic sensitivity of the GP and BS primer sets was compared after integration of the β-globin primers (see next paragraph). PCR was performed at least in duplicate using serial 10-fold dilutions of plasmids containing genomic DNA from 24 HPV types in 100 ng of HP DNA (Table 2). PCR products were analyzed by MPG. The BS primer set detected all 24 HPV genotypes with a sensitivity between 10 and 1,000 copies, while the analytic sensitivity of GP PCR varied between 10 and 1,000,000 copies. For HPV 11, 16, 18, 31, 33, 43, 58, 59, and 66, both PCR primer sets demonstrated the same sensitivity. For HPV 6, 35, 42, 45, 52, 70, and 73 the BS primer set was 10-fold, for HPV 26, 39, 56, 68, and 51 100-fold, and for HPV 44, 53, and 82 1,000-fold more sensitive than GP. These differences were independent of β-globin coamplification (see below).
Development of novel β-globin primers. No internal DNA quality control has been described for the GP PCR. Instead, external β-globin PCR serves to control for DNA integrity, but it cannot control for the PCR efficacy in the HPV PCR itself. To overcome this limitation, novel β-globin primers were integrated into the BS PCR as well as the standard GP PCR. The CO3/5 primer set (5) and seven newly designed primer pairs (MS1-7/MS8-14) were included in various combinations in GP or BS PCR, and 10-fold dilutions of plasmids containing HPV 16, 18, 31, or 33 DNA were amplified in a background of 100 ng of human cellular DNA. β-Globin and HPV PCR products were analyzed by agarose gel electrophoresis (data not shown). The CO3/5 primers failed to coamplify the β-globin sequence, while three combinations of newly designed primers succeeded. These primer sets were further tested with DNA from SiHa and CasKi cells and from clinical specimens. Only the MS3/MS10 primers generated PCR products visible in the gel (data not shown).
Integration of β-globin primers in the HPV PCR. Next, we determined the β-globin primer concentration that maximized β-globin signals after integration in the BS and GP PCR without concurrent competition of HPV PCR.
First, different amounts of MS3/MS10 β-globin primers were tested in the presence of HPV 16 (5,000 copies) and 100 ng of cellular DNA. β-Globin primer concentrations of 100 nM in the GP PCR (Fig. 1) and 300 nM in the BS PCR yielded similar β-globin Luminex signals (data not shown). Titration of cellular DNA resulted in a detection limit for β-globin of less than 10 ng (approximately 1,700 genome equivalents) by GP and BS PCR (data not shown).
![]() View larger version (15K): [in a new window] |
FIG. 1. Dependence of HPV amplification on β-globin coamplification. Twofold dilutions of β-globin primers with stable amounts of GP primers (500 nM) were tested on 100 ng of HP DNA spiked with (empty symbols) or without (filled symbols) 5,000 copies of HPV 16. PCR was performed in duplicate; mean net MFI values and standard deviations are indicated. HPV 16 signals are shown in bars, β-globin signals are shown in circles. No HPV 16 was detected in β-globin singleplex PCR.
|
![]() View larger version (22K): [in a new window] |
FIG. 2. Influence of HPV and β-globin coamplification on HPV amplification sensitivity. HPV 16, 39, 66, and 82 plasmids diluted in 100 ng HP DNA were amplified in duplicate by BSGP5+/6+ PCR with (empty symbols) and without (filled symbols) β-globin primers. HPV 16 signals are shown by bars, and β-globin signals are shown by circles; the cutoff is indicated by the dotted line.
|
![]() View larger version (13K): [in a new window] |
FIG. 3. Analytic sensitivity of BSGP5+/6+ PCR for HPV 82 in different amounts of cellular DNA. Dilution series of HPV 82 plasmids in 100 to 1,000 ng of HP DNA were amplified in the presence and absence of β-globin primers. The cutoff is indicated by the dotted line.
|
For multiplex detection of the HPV (150 nucleotides) and β-globin PCR products (208 nucleotides), a specific β-globin probe was designed and integrated in MPG (Fig. 1 to 4).
![]() View larger version (15K): [in a new window] |
FIG. 4. Reproducibility of BSGP5+/6+ PCR with β-globin coamplification. Tenfold dilution series of plasmids containing genomic HPV 66, diluted in 100 ng of HP DNA, were subjected 32 times to BS PCR with β-globin coamplification. Coamplified HPV 66 and β-globin PCR products were analyzed by MPG. The lower and upper edges of the boxes are the 25th and 75th percentiles, respectively. Median values are shown by the lines in the boxes, the whiskers represent the 5th and 95th percentiles, outliers are indicated by circles, and the cutoff is indicated by the dotted line.
|
Semiquantitative HPV detection, which is independent of β-globin coamplification, was possible for all HPV types and is exemplarily depicted for HPV 16, 39, 66, and 82 (Fig. 2). In addition, quantification was also feasible when three HPV plasmid DNA dilution series were mixed, with 10- to 1,000-fold differences in concentration. The resulting MFI values for each HPV type closely reflected the relative concentrations of targets included in the PCR (data not shown). Altogether, these data demonstrated the interassay consistency of MPG and its ability to quantify several HPV types, especially in multiple infections.
Evaluation of BSGP5+/6+ PCR with clinical samples. We compared the performance of BS plus integrated β-globin PCR against that of GP without β-globin PCR on DNA from exfoliated cervical cells from 1,017 women of the general population in Mongolia (Dondog et al., submitted) and 95 Mongolian cervical cancer patients. To minimize variation, both PCRs were performed in parallel and the products were analyzed on the same plate by MPG. Of the 1,112 samples, 27 were excluded because they were negative for β-globin and HPV amplification by BS; three of the excluded samples strongly tested HPV 16 or 31 positive by GP PCR, while the other 24 were HPV negative.
HPV prevalence in the remaining 1,085 clinical samples is presented in Tables 4 and 5. Of all typing results (1,085 samples and 27 HPV types analyzed), 639 (2.2%) were concordantly positive, 28,378 (96.9%) were concordantly negative, and 278 (0.9%), discussed in detail in the next paragraph, were discordant. These numbers yielded a kappa value of 0.816 (95% confidence interval, 0.797 to 0.836). Identical typing results were obtained for 858 (79.1%) of the 1,085 clinical samples. Despite β-globin coamplification, the overall HPV prevalence was higher (46.0%) by BS than GP (41.4%).
|
View this table: [in a new window] |
TABLE 4. Detection of HPV genotypes in 1,085 clinical samples by BSGP5+/6+ or GP5+/6+ PCR followed by MPG
|
|
View this table: [in a new window] |
TABLE 5. Number of HPV types found in 1,085 clinical samples with BSGP5+/6+ and GP5+/6+ PCR, respectively
|
Between 15% (HPV 73) and 73% (HPV 44) of the HPV infections detected by BS only were single infections. However, the majority of the additionally detected infections was found in multiple infections, resulting in 38.5% multiple infections (192 of 499 total HPV-positive samples) by BS versus 34.7% (156 of 449) by GP (Table 5). Among the general population (n = 992) the proportion of multiple infections was 16.8% by BS versus 14.1% by GP. In cancer patients (n = 93), multiple infections were significantly more frequently found using BS (26.9%, P = 0.007), while the frequency found using GP was not significantly different from that in the general population (17.2%, P = 0.084).
Amplification analysis of HPV types. To further characterize the different affinities of both primer sets to individual HPV types, we took advantage of the quantitative read-out of MPG. For the reactions positive (net MFI of >5) with both primer sets, we computed the mean of the ratio: BS signal over GP signal (Rsignal BS/GP). This ratio allowed us to identify HPV types in two distinct patterns: P1 had an Rsignal BS/GP between 0.5 and 2.0 (positive signals differing less than twofold between PCRs), and P2 had an Rsignal BS/GP above 2.0 (>2-fold-higher signals with BS). According to these criteria, P1 was found for HPV 6, 11, 16, 18, 26, 31, 33, 35, 43, 45, 56, 58, 59, 66, 67, 70, and 73, and P2 for HPV 30, 39, 42, 44, 51, 52, 53, 68, and 82 (Table 4). No Rsignal BS/GP could be computed for HPV 44, grouped in P2, because it was found exclusively by BS. The Rsignal BS/GP closely correlated with the number of additional HPV reactions found by BS or GP (Fig. 5).
![]() View larger version (6K): [in a new window] |
FIG. 5. Relative HPV detection ability and signal strength in positive reactions with BSGP5+/6+ and GP5+/6+. The ratio of additional HPV detections by BSGP5+/6+ to those by GP5+/6+ (x axis) is plotted against the ratio of hybridization signals with BSGP5+/6+ versus GP5+/6+ for reactions positive by both methods (y axis) in a logarithmical scale. Missing HPV types overlap with other types shown; e.g., HPV 6 clusters with 11, 18, 35, and 70, HPV 56 with 45, and HPV 73 with 58.
|
|
|
|---|
Our novel primer set demonstrated an increased sensitivity for at least 10 HPV types, despite the coamplification of β-globin by the MS3/MS10 primer pair. The overall amplification of mucosal HPV types under standardized experimental conditions was more homogeneous, reaching detection limits for all 24 HPV types analyzed between 10 and 1,000 copy numbers. The improved sensitivity for HPV types 30, 39, 42, 44, 51, 52, 53, 68, 73, and 82 was confirmed with clinical samples and correlated with the prediction based on the mismatch analysis with target HPV regions. In addition, BS primers have shown an improved detection of HPV 61 and 84 (I. Sabol, M. Salakova, J. Smahelova, M. Pawlita, M. Schmitt, N. M. Ga
perov, M. Grce, and R. Tachezy, submitted for publication).
The increased sensitivity of BS is entirely due to primer differences, since the same detection system (MPG) was used for both methods. The BS primer pool not only reduced mismatches to HPV types that were not sensitively detected by GP but also reduced competition in multiple HPV infections. However, increased primer concentrations in BS can generate larger amounts of PCR products, which may result in weak cross-reactivities with related probes. Cross-reactivities were determined by colony PCR for all 27 HPV types. These PCR products, originating from millions of HPV copies, revealed a single cross-reactivity; i.e., very strong HPV 51 amplification showed weak hybridization (less than 5% of HPV 51 signal) with the HPV 82 probe (data not shown). Such HPV 82 signals were excluded when HPV positivity was scored.
HPV 31 detection by BS PCR was reduced in clinical samples with low viral loads or multiple infections. Hitherto, we could not find a plausible explanation for this observation. However, increased concentrations of GP5+ and GP6+ primers in the BS mixture resulted in stronger HPV 31 amplification, permitting the detection of 7 out of 13 previously missed HPV 31 infections. Improved amplification was also observed for HPV 66 infections, while the detection of HPV types preferentially amplified by the initial BS PCR protocol remained unchanged. As a consequence, the modified protocol will be used in future studies.
The quantitative read-out of MPG provided a second measure to assess both PCRs abilities to amplify distinct HPV types, including types with a low prevalence in the study. Overall, Rsignal BS/GP values were largely consistent with findings in the clinical samples, confirming the reduced amplification of HPV 31 and 66 and the improved amplification of HPV 30, 39, 42, 44, 51, 52, 53, 68, 73, and 82 by BS (Table 4; Fig. 5).
Likewise, the quantitative data obtained from MPG can be used to better characterize multiple HPV infections. The resulting intensities of MFI signals for each HPV type in a multiple infection faithfully reflected the relative concentrations of targets included in the PCR. This was demonstrated in dilution series of various cloned HPV types. As a result, the dominant HPV type(s) present in an infection, i.e., the type(s) with the highest viral loads, could be easily identified. It has been shown that 20 to 30% of HPV-positive women harbor multiple types that were acquired simultaneously or successively (10). It remains controversial whether an infection with multiple HPV DNA types is a risk factor for HPV persistence and for cervical lesions (14, 16). However, a reliable and unbiased profiling of individual HPV types in multiple HPV infections will be important in evaluating the efficacy of HPV vaccine implementation. In the present study, we demonstrated the improved detection of multiple HPV infection by the novel BS PCR in comparison to GP PCR. A significantly higher prevalence of multiple infections was found in cancer samples than in samples from the general population only by using the BS PCR system. This can be explained by the presence of very high viral loads of the "driving" HR HPV type in cervical cancer. As a consequence, PCR reagents, i.e., GP primers, are preferentially consumed by this type, thereby weakening the amplification of other coinfecting types. Thus, addition of several different BS primers minimizes the risk of amplification bias by a dominant HPV infection, as is in particular the case for cervical cancer samples. Therefore, BS primers may be the better tool for future studies analyzing multiple HPV infections as a risk factor for developing cervical cancer.
More studies are required to show whether the additional detection of some HPV types by the novel BS primers will be of value for predicting the risk of having or developing cervical lesions. However, we believe that the detection of high copy numbers of "rare" HPV types, currently being missed by GP PCR, may have an impact on the management of HPV-related diseases. As such, GP PCR-reverse line blot-based epidemiological studies, summarized in recently published meta-analyses (3, 4), found HPV 39, 51, 66, 68, and 82 in either single or multiple infections in less than 2.9% of high-grade squamous intraepithelial lesions and less than 0.6% of squamous cell carcinoma of the cervix. The distribution of HPV 26 and 53 was not described. The use of MPG in combination with the novel BS PCR, however, suggests that in squamous cell carcinoma of the cervix, all these types may be present more frequently than described so far (unpublished data).
Internal DNA quality controls (e.g., β-globin) possess the advantage of simultaneous HPV and genomic DNA amplification; however, they may also decrease the analytic sensitivity of the HPV PCR (7). This decrease can be explained by highly efficient β-globin amplification, which competes for DNA polymerase and deoxynucleoside triphosphates and exhibits higher sensitivity than the HPV primers. As a consequence, most HPV PCRs, such as GP, lack an internal DNA quality control and require sample prescreening by an external β-globin PCR and subsequent gel electrophoresis (5, 8). However, the use of internal DNA quality controls without concurrent impairment of the HPV PCR is desirable. In addition to gaining time and reducing study costs and sample consumption, an internal DNA quality control can also monitor a failure of the HPV PCR. To this end, we designed a novel β-globin primer pair, MS3/MS10, and showed that its integration had no influence on the sensitivity of the conventional GP PCR or the novel BS PCR. This was achieved by adjusting the β-globin primer concentration to reach an analytic sensitivity of approximately 1,700 genome equivalents, which was purposely lower than for any of the HPV types. Thus, clinical samples containing large amounts of HPV copy numbers showed no or suppressed β-globin signals, while strong HPV values were measured. Nevertheless, this effect is highly desirable, minimizing the risk of false-negative HPV results due to a dominant β-globin coamplification. In contrast, HPV PCR sensitivity decreased with an increasing amount of cellular background DNA, independent of simultaneous β-globin coamplification and the amount of β-globin PCR product obtained.
The analytic sensitivity of a HPV genotyping system is composed of the amplification sensitivity and the detection limit of the read-out method. The selective overamplification of certain HPV types, as well as limiting sensitivity of read-out techniques, however, can skew the reported frequency of mucosal HPV types. Moreover, methodological bias leads to an underestimation of the prevalence of distinct types in cervical specimens (1). Improved HPV genotyping is needed (i) to detect a higher variety of HPV types, (ii) to reduce the number of false-negative and false-positive results, (iii) to better characterize multiple infections, and (iv) to save time and cost. New HPV types have continuously been discovered, and a yet-larger number is expected to exist. Therefore, HPV amplification and genotyping methods that allow specific, sensitive, unbiased, and robust detection of HPV DNA and yet are flexible enough for the integration of additional HPV types are needed. Hitherto, genotyping of GP5+/6+ PCR products has been based on type-specific hybridization with oligonucleotide probes immobilized to membranes (RLB) (19) or to fluorescent beads (MPG) (17). The latter easily allows extension of the probe number to cover all mucosal HPV types with known sequences in the amplified L1 region. The automated read-out of MPG provides semiquantitative analysis of PCR products, avoids mistakes during data entry, and allows genotyping of up to 1,000 PCR products per day.
In conclusion, this study describes the development and validation of a novel BS PCR for the sensitive detection of HPV in single as well as multiple infections. Our data suggest that especially for pHR HPV type 53 and HR HPV types 39, 51, 52, 68, 73, and 82, use of BSGP5+/6+ PCR in clinical practice will lead to more reliable estimates of infection rates with these types. Further studies will clarify whether the use of BSGP5+/6+ PCR in combination with MPG has a positive impact on the prediction of developing cervical lesions caused by these types. We believe that by using the BSGP5+/6+ PCR with internal DNA quality control in conjunction with MPG, researchers will obtain a powerful, highly flexible and sensitive tool for a more homogeneous amplification and genotyping of HPV that could be used in large epidemiological studies, in particular in vaccinated populations.
Published ahead of print on 16 January 2008. ![]()
|
|
|---|