Journal of Clinical Microbiology, March 2008, p. 897-901, Vol. 46, No. 3
0095-1137/08/$08.00+0 doi:10.1128/JCM.01553-07
Copyright © 2008, American Society for Microbiology. All Rights Reserved.

Department of Epidemiology, the University of North Carolina at Chapel Hill, Chapel Hill, North Carolina,1 Division of Infectious Diseases, the University of North Carolina at Chapel Hill, Chapel Hill, North Carolina,2 National Reference Laboratory, Tuberculosis Control Program, Kinshasa, Democratic Republic of Congo,3 Mycobacteriology Unit, Institute of Tropical Medicine, Antwerp, Belgium,4 Ecole de Sante Publique, University of Kinshasa, Kinshasa, Democratic Republic of Congo,5 National Tuberculosis Program, Kinshasa, Democratic Republic of Congo6
Received 3 August 2007/ Returned for modification 21 October 2007/ Accepted 23 December 2007
|
|
|---|
|
|
|---|
Case detection through quality-assured bacteriology is an essential element of the WHO STOP TB strategy (8). Because of a limited culture capacity, many resource-poor countries rely solely upon sputum smear microscopy for the diagnosis of TB. The quality of smear microscopy depends on a network of local laboratories and external quality assessment (EQA) of these laboratories under the supervision of the national reference laboratory (NRL) (9). EQA of smear microscopy in resource-poor settings most often consists of on-site unblinded review by a laboratory supervisor of positive slides and 10% of negative slides. This method, which has not been validated in the field, is labor-intensive and is often a neglected part of national TB programs in resource-poor countries (7).
In an effort to simplify and standardize EQA activities and to prioritize EQA at national TB control programs (NTPs), a practical EQA guideline was developed by an international working group and endorsed in 2002 (2). These international EQA guidelines recommend the use of three methods for the evaluation of laboratory performance: on-site assessment by the use of a standardized questionnaire, panel testing of technician proficiency by the use of centrally prepared slides, and blinded rechecking of a random sample of routine slides from each peripheral laboratory. To decrease the EQA workload, the blinded rechecking recommendations incorporate statistical sampling methods developed for industrial quality control. To date, few countries have implemented the 2002 EQA guidelines and few studies have evaluated the routine implementation of these new guidelines in resource-poor settings (3-6).
We assessed the impacts of an intervention consisting of a 5-day refresher training course for laboratory technicians and the distribution of new microscopes on the performance of smear microscopy. We assessed the feasibility of the implementation of two components of the new EQA guidelines, blinded rechecking and on-site evaluation, in 13 laboratories in Kinshasa, Democratic Republic of Congo (DRC). Panel testing was not evaluated, as it has limited value in assessing routine performance and might best be reserved for testing at the end of training sessions (7).
|
|
|---|
Several steps were implemented as part of the study. First, selected sites were visited for on-site evaluation and the collection of slides for blinded rechecking before the intervention. Second, all laboratory staff performing smear microscopy at the selected sites underwent a 5-day refresher training course at NRL, and new microscopes were distributed to each site. Third, 9 months later, slides were again selected for blinded rechecking.
Site selection. Thirteen diagnostic TB centers were selected from among the 89 diagnostic TB centers in Kinshasa, based on the smear microscopy workload in 2003. The 13 laboratories selected performed 45% of all smears for acid-fast bacilli (AFB) in Kinshasa. The median annual workload at the 13 selected sites was 4,371 sputum smears (range, 1,283 to 17,837 sputum smears) for AFB. Carbol fuchsin staining solutions for the entire city were centrally prepared and quality controlled before distribution to the health care centers. Smear microscopy was performed according to DRC national guidelines. The selected sites agreed not to rotate laboratory technicians during the study period.
On-site evaluation. On-site evaluation of each laboratory was performed by using the comprehensive on-site evaluation checklist proposed by the 2002 EQA guidelines and adapted to reflect the DRC NTP guidelines. The evaluation was completed by the same NRL and research staff for all sites and consisted of both open- and closed-ended questions, simple observations, and limited on-site rechecking of slides.
Intervention. All laboratory technicians employed at the study sites underwent a 1-day theoretical and 4-day practical refresher training course in smear microscopy. The participants were administered a standardized pre- and posttraining test addressing theoretical and practical knowledge related to TB and AFB sputum microscopy. Each center received one or two new microscopes following the training, and the technicians were instructed on their use and care.
Blinded slide rechecking. The blinded slide rechecking process consisted of sample size calculation, slide storage, slide collection, blinded slide rechecking at the NRL before and after restaining, rechecking of slides with discrepant results at a supranational reference laboratory (SRL), and the classification of errors.
In compliance with the 2002 EQA guidelines, the smallest possible sample size that allows solid conclusions to be made about the performance of a laboratory was calculated by using the lot QA system (LQAS) method. The 2003 positivity rate and the total number of slides processed at all laboratories in Kinshasa were provided by the NTP. The maximum number of false-negative errors allowed in a sample was set equal to 0 (acceptance number of 0). The expected performance (sensitivity) of the peripheral laboratory technicians compared to the performance of the controllers was set equal to 80%. The specificity of the peripheral laboratory technicians compared to that of the controllers was set equal to 100% (no false-positive results were tolerated). On the basis of these parameters, the annual sample size needed for the blinded rechecking was 57 slides per site.
Each peripheral laboratory was asked to store the 500 most recently collected slides and have them correctly labeled and cleaned with xylene. The 57 slides for blinded rechecking were selected from among these 500 stored samples. To ensure a random, unbiased, and representative sample of slides from each site for blinded review, sample slides were identified from the register by the research staff and collected by the laboratory technician. If a slide was missing, the next slide identified in the register was selected. The slides selected thus represent slides with negative, positive, and scanty positive results in proportion to their occurrence in the laboratory register.
Randomly selected slides were reviewed in a blinded fashion at the NRL by using the same standard technique employed at the peripheral laboratory. The results were reported according to the standard IUATLD grading scale (Table 1). Following an initial review at the NRL, all slides were also rechecked after they were restained. Discrepant results were resolved at the Institute of Tropical Medicine, Antwerp, Belgium, an SRL. The technician at the SRL was informed of both results but was blinded to which result was from the peripheral laboratory and which one was from the NRL. The decision given at the SRL was considered final.
|
View this table: [in a new window] |
TABLE 1. IUATLD grading scale for AFB found by Ziehl-Neelsen smear microscopy
|
|
View this table: [in a new window] |
TABLE 2. Classification of rechecking errors encountered upon blinded rechecking of smear microscopy results as part of EQA of peripheral laboratories
|
|
|
|---|
|
View this table: [in a new window] |
TABLE 3. Ziehl-Neelsen smear microscopy results pre- and posttraining in 13 public health laboratories in Kinshasa
|
|
View this table: [in a new window] |
TABLE 4. Most common problems reported or observed during on-site evaluation visits of 13 public health laboratories in Kinshasa
|
Blinded slide rechecking. Among the 741 slides collected from the peripheral laboratories and reviewed by the NRL, there were 77 (10.4%) discrepant results. According to the SRL results, 67 (87%) of these discrepant results were attributed to the peripheral laboratory and 10 (13%) were attributed to the NRL.
Prior to the intervention, major errors (i.e., HFP and HFN errors) were observed in 8 of the 13 (61.5%) laboratories, with HFN results being much more frequent than HFP results (Table 5). Minor errors (i.e., LFP and LFN errors and QEs) were observed in the majority (10 or 77%) of peripheral laboratories. After the intervention, the occurrence of major errors remained high, with one or more major errors being observed in 10 (77%) clinics. HFN results were still more likely to occur than HFP results. The number of major errors decreased at five sites, increased at six sites, and remained unchanged at the remaining two sites. The proportion of laboratories where minor errors were detected also remained unchanged from the preintervention to the postintervention assessment. LFN results also remained more common than LFP results.
|
View this table: [in a new window] |
TABLE 5. Results of slide rechecking at 13 primary health laboratories in Kinshasa
|
|
|
|---|
Similar to the findings of studies performed in India, the Philippines, and Mexico, we found blinded rechecking and the LQAS strategy to be operationally feasible (3-6). The blinded rechecking identified those laboratories with a high number of smear microscopy errors and allowed the distinction between those laboratories with optimal and suboptimal performance. Similar to the conclusions of a study performed in Uganda, we conclude that a standardized on-site assessment is a useful tool in quality control and that meaningful changes can be successfully implemented on the basis of the results (1). The on-site evaluation identified potential problems underlying the substandard performance at deficient laboratories and helped with the formulation of corrective actions. On-site evaluation was also a key in motivating laboratory technicians. Implementation of the new EQA method resulted in an important decrease in the number of slides that had to be rechecked, but the use of the comprehensive standardized checklist was time-consuming (1 to 2 h) and many of the deficiencies identified were common, with multiple findings suggesting problems in the same areas. These limitations of the on-site evaluation may be overcome by use of the short checklist included in the 2002 EQA guidelines as a template, and tailoring the questionnaire to the specific problems that a country or a region has identified at sentinel sites may further improve the efficiency of the on-site evaluation.
The greatest challenges encountered were related to the interpretation of the results, the provision of timely feedback, and the formulation of corrective actions. The 2002 EQA guidelines state, "When establishing a rechecking program, it will be important for the NTP to establish standards for acceptable performance, as well as recommended investigation steps and appropriate actions to correct problems" (2). Unfortunately, the directions given in the guidelines on interpretation can lead to confusion. The 2002 EQA guidelines suggest three different interpretations. All interpretations have no errors of any type as the target for optimal performance, and the guidelines recommend that any major error and frequent minor errors trigger an evaluation and the taking of corrective action, if it is needed. According to the most stringent interpretation, any major error may indicate unacceptable performance, while according to the least-stringent interpretation, any HFP result or one or two HFN results may indicate unacceptable performance. While HFP errors should not occur (the specificity is set at 100% in the sample size calculation), an isolated HFP result can be due to a clerical error. More than one HFP result suggests serious microscope malfunction, grossly inadequate technique, or an inability to recognize AFB due to inadequate training or high staff turnover. An HFN result may suggest work overload, poor staining reagents or technique, inadequate microscopes, or incorrect reading. Frequent LFN results should also be addressed for these deficiencies but are most often due to careless work.
In our study, most laboratories had one or more major errors, predominantly false-negative errors. The numbers of laboratories with unacceptable performance preintervention were eight (61.5%) when the example in the guidelines with the most stringent criteria was applied and four (31%) when the least-stringent criteria were applied. Until the working conditions in underresourced public health systems are corrected, the use of the least-stringent criteria may be more efficient, as a higher proportion of laboratories with a problem in need of technical correction would be focused on. Under such conditions, one may even consider reduction of the set point of sensitivity from 80% to 70%. This would allow EQA staff to focus on the provision of timely feedback, problem solving, and supportive supervision at those laboratories with the poorest performance.
Even though the 5-day training course resulted in a significant improvement in the scores on the standardized test, refresher training and the provision of new microscopes did not lead to an observable long-term (9-month) improvement in the quality of smear microscopy. This may be because certain factors were not addressed by the intervention, such as a high workload, poor working conditions, and poor staff motivation; because the sample size in the LQAS method is too small to detect relatively small improvements in performance; or because the training did not have a long-lasting effect.
The logistical complexity of involving both an NRL and an SRL resulted in important delays in the provision of feedback of the results to the peripheral laboratories. This experience suggests that the EQA process will be effective and allow the provision of rapid feedback only if it is completed locally, with final controls at the national or even the regional level and continuous evaluation of the controllers. Timely completion and the provision of feedback with probing for problems are probably far more important than the type of laboratory giving the final verdict. Without proper feedback, other efforts are likely to be in vain. On the basis of the results of this study, the DRC NTP has revised its EQA guidelines and has introduced LQAS method sampling via training sessions, with more emphasis on correct technical execution and follow-up.
In conclusion, this study supports the implementation of the 2002 EQA guidelines in resource-poor settings. The use of the LQAS technique leads to a smaller sample size, thus reducing the EQA workload; and the combination of on-site evaluation and blinding of rechecking allows an unbiased and representative evaluation of the quality of sputum microscopy, the identification of underlying problems, the formulation of feedback and corrective actions, and the motivation of laboratory technicians. Prior to implementation of the EQA system, clear interpretation guidelines must be agreed upon by the NTP; and plans must be in place to provide timely feedback, improve working conditions at suboptimally performing laboratories, and correct underlying problems identified during on-site evaluations.
We acknowledge the work of the laboratory technicians at the participating laboratories and the NRL in Kinshasa. We acknowledge Mourad Gumusboga at the Institute of Tropical Medicine, Antwerp, Belgium, for his work on rereading discrepant slides.
Published ahead of print on 3 January 2008. ![]()
|
|
|---|