Received 29 November 2007/ Returned for modification 2 January 2008/ Accepted 25 January 2008
|
|
|---|
|
|
|---|
Cell culture is still the "gold standard" for the laboratory detection of respiratory viruses. However, cell culture is slow and has a low sensitivity. Therefore, its implementation for routine virus detection is suboptimal. Although rapid antigen detection tests are available for some of the respiratory viruses, these tests have been shown to be less sensitive and less specific than cell culture-based approaches (5, 29). Nucleic acid amplification tests have proven to be rapid, very sensitive, and specific alternatives and can be used in either a monoplex or a multiplex format. Moreover, multiplex assays allow the coamplification of more than one target, thus providing insight into the significance of mixed infections for the prognosis and recrudescence of the respiratory disease. In addition, the incorporation of the ability to detect viruses such as the SARS coronavirus and InfA H5N1 in a multiplex assay would allow monitoring of these viruses and could act as an early built-in detection system (28).
An all-embracing multiparameter test is a prerequisite to reducing the costs involved with such a comprehensive monitoring system (19). Currently, several assays with multiplex formats detect up to nine respiratory viruses in one reaction. For instance, several real-time multiplex assays allow the real-time detection of up to four targets in a reaction, depending on the number of channels available in the real-time PCR machines (13, 30, 34). Based on the Roche LightCycler480 instrument, a real-time TaqMan PCR that detects five different targets simultaneously has been developed (21). Existing multiplex PCR assays that use agarose gel electrophoresis or capillary electrophoresis as the detection system detect five to eight targets per reaction (3, 7, 24). Multiplex PCR assays combined with an enzyme-linked immunosorbent assay currently detect up to nine targets simultaneously (12, 26).
Recently, two multiplex assays that detect respiratory viruses have been evaluated with 360 clinical samples (17). The NGEN respiratory virus analyte-specific assay (Nanogen, San Diego, CA) detects InfA, InfB, PIV type 1 (PIV-1), PIV-2, PIV-3, and RSV on a NanoChip 400 electronic microarray. The ResPlex II assay (Genaco Biomedical Products, Inc., Huntsville, AL) detects InfA, InfB, PIV-1 to PIV-4, RSV, hMPV, rhinovirus, enterovirus, and the SARS coronavirus on a Luminex 100 instrument (Luminex, Austin, TX). Both assays exhibited specificities comparable to those of cell culture and the monoplex real-time reverse transcription (RT)-PCR. However, the sensitivities of both assays were lower than the sensitivity of the monoplex real-time RT-PCR.
The multiplex ligation-dependent probe amplification (MLPA) technology (27) uses a simple-to-perform, multiplex PCR method which is able to amplify up to 45 different targets simultaneously. MLPA is based on the oligonucleotide ligation assay, and identification is accomplished by size by using gel electrophoresis methods. MLPA has three steps: first, an annealing step hybridizes the probes to their target regions; second, a ligation step links the two probes; and third, the final PCR exponentially amplifies the ligated probes with only two primers (Fig. 1). So far, MLPA has mainly been used to detect changes in the copy numbers of specific chromosomal regions (27), in expression profiling studies (6), for the detection of CpG methylation of genes (22), and for the detection of recombination events (16). We developed a multiparameter test (the RespiFinder assay) that is based on the MLPA technology and that can be used clinically to detect infectious disease agents.
![]() View larger version (16K): [in a new window] |
FIG. 1. Overview of the RespiFinder technology. A one-step RT-PCR is performed with specific primers for all targets (steps 1A and 1B). Subsequently, an MLPA reaction is performed with MLPA probes specific for all targets (steps 2, 3, and 4). An MLPA probe consists of two oligonucleotides: one synthetic oligonucleotide and one M13-derived oligonucleotide. The synthetic oligonucleotide contains a universal forward priming site, and the M13-derived oligonucleotide contains a universal reverse priming site. In addition, the M13-derived oligonucleotide contains a unique stuffer sequence. The length of this stuffer sequence is specific for each probe and varies between the different probes. The length of the MLPA probe is the combined length of both oligonucleotides. This length is unique for each probe due to the specific stuffer sequence. The two oligonucleotides hybridize specifically to the target adjacent to each other. Subsequently, the two oligonucleotides are joined by ligation. The ligated oligonucleotides are amplified by one universal primer set. After amplification, the MLPA reaction is analyzed by electrophoresis (step 5). Each MLPA probe can be discerned due to its specific length. Recently, the protocol has been further optimized by combining the probe ligation (step 3) and probe amplification (step 4) in one reaction.
|
The RespiFinder assay was validated with 144 clinical samples, and the results were compared with those of cell culture and an RSV-specific immunochromatography assay (ICA). We demonstrate that the RespiFinder assay provides an excellent tool for the simultaneous detection of 15 respiratory viruses.
|
|
|---|
Clinical samples. A total of 144 nasal wash specimens (from both adults and children) were submitted for the routine diagnosis of viral infection; of these, 110 samples were tested by cell culture on four different cell lines (Virology Department, University Hospital St. Radboud, Nijmegen. The Netherlands). This included adenovirus, hMPV, InfA, InfB, PIV-1 to PIV-4, rhinovirus, and RSV. The samples were not routinely checked for the presence of coronaviruses. A rapid ICA was performed for the detection of RSV in 34 samples from children by using the RSV Respi-Strip from Coris BioConcept (Gembloux, Belgium). All clinical samples were stored at –80°C until further analysis by the RespiFinder assay.
Nucleic acid extraction. Viral RNA or DNA was isolated by using the automated MagNA Pure LC system (Roche Diagnostics, Almere, The Netherlands). The MagNA Pure LC total nucleic acid isolation kit and the total nucleic acid lysis extraction MagNA Pure protocol were applied. Extraction was performed according to the manufacturer's instructions. Briefly, 200 µl of starting material was used, and the purified nucleic acid was eluted in a final volume of 100 µl. Before the start of extraction, 5 µl of the IAC, which contained an encephalomyocarditis virus (EMC) RNA transcript, was added to the lysed samples.
In vitro transcripts of viral targets. Synthetic viral target genes, which contained both the PCR primer regions and the MLPA probe sequence, were constructed by using the software package of DNA Works (version 3.1). The DNA Works software calculates the composition and the number of oligonucleotides required to synthesize the viral target. The oligonucleotides generated are characterized by highly homogeneous melting temperatures and a minimized tendency for hairpin formation. By using a two-step PCR protocol, the synthetic viral target gene can easily be synthesized and amplified (15).
The amplification products were purified with an MSB Spin PCRapace kit (Invitek, Berlin, Germany) and were subsequently cloned in pGEM-T Easy vector system II (Promega, Leiden, The Netherlands). The nucleotide sequences of the cloned viral target genes were sequenced on an ABI 3100 DNA capillary sequencer. All cloned viral targets were linearized and transcribed in vitro from the SP6 promoter by using the Riboprobe in vitro transcription kit (Promega, Leiden, The Netherlands). Residual plasmid DNA was removed by DNase treatment (TURBO DNA-free; Ambion, Huntingdon, United Kingdom), according to the manufacturer's instructions. The concentration of the SP6 RNA transcripts was determined by measuring the absorbance at 260 nM (1 unit of the optical density at 260 nm is equal to 40 µg/ml of RNA), and the absence of residual plasmid DNA was checked by performing a real-time PCR without an RT step. Aliquots of the RNA transcripts were stored at –20°C, and dilutions were made in Tris-EDTA (pH 7.6) containing 1 µg/ml of carrier DNA (herring sperm DNA; Fermentas, St. Leon-Rot, Germany).
IAC. The IAC was a SP6 RNA transcript of the polyproteon gene from EMC. The IAC was produced as described above. Twofold serial dilutions of the IAC were spiked into a clinical sample which previously tested virus negative by the RespiFinder assay. Subsequently, the spiked clinical samples were analyzed by the RespiFinder assay. The optimal spiked concentration was defined as two times the lowest detectable concentration.
Preamplification. A one-step RT-PCR protocol was performed with a OneStep RT-PCR kit (Qiagen, Hilden, Germany). Combined RT and multiplex target amplification PCR were performed in a 25-µl reaction mixture, according to the manufacturer's instructions, in the presence of all preamplification primers (Table 1). The mixture contained 0.2 µM of each primer and 10 µl of the RNA or DNA sample. Amplification was performed on a T1 thermocycler (Biometra, Göttingen, Germany), as follows: an initial cycle of 30 min at 50°C and 15 min at 95°C, followed by 30 cycles of 30 s at 94°C, 30 s at 55°C, and 60 s at 72°C and, finally, a 10-min extended elongation.
|
View this table: [in a new window] |
TABLE 1. Sequences of the preamplification primers
|
|
View this table: [in a new window] |
TABLE 2. Target genes and sequences of virus specific regions of the MLPA probes
|
Analysis of PCR products. The amplified MLPA products can be analyzed on different detection platforms, e.g., by electrophoresis in 2.5% agarose gels, by use of the HT DNA 5000 SE55 kit on a LabChip 90 system (Caliper LifeSciences, Teralfene, Belgium), in 6.5% acrylamide slab gels in a 4300 DNA analyzer (LI-COR Biosciences, Lincoln, NE), by capillary electrophoresis on a MegaBACE DNA analysis system (GE Healthcare Europe GmbH, Diegem, Belgium), and/or in an ABI 3100 genetic analyzer (Applied Biosystems, Nieuwerkerk a/d IJssel, The Netherlands).
For detection on the 4300 DNA analyzer, an IR-700-labeled reverse primer was used in the final PCR. The final PCR mixture was diluted 30 times and was analyzed on the 4300 DNA analyzer according to the manufacturer's instruction. When analysis was performed on a MegaBACE DNA analysis system or an ABI 3100 genetic analyzer, a 6-carboxyfluorescein-labeled forward primer was used in the final PCR. The final PCR mixture was diluted 10 times and was subsequently treated according to the manufacturers' instructions. All clinical samples were analyzed by capillary electrophoresis on a MegaBACE DNA analysis system or an ABI 3100 genetic analyzer.
|
|
|---|
An IAC is included in the assay to discriminate between samples with true-negative results and samples with false-negative results due to PCR failure. This added up to a total of 16 probes. The total lengths of the probes varied from 157 to 492 nucleotides. The sizes of the probes of the related viruses were clustered. Four clusters are distinguished: an influenza virus cluster of about 400 nucleotides, a PIV cluster of about 300 nucleotides, an RSV cluster of about 250 nucleotides, and a coronavirus cluster of about 170 nucleotides. An overview of all probes is presented in Fig. 2.
![]() View larger version (24K): [in a new window] |
FIG. 2. Overview of the RespiFinder probes and their length. nt, nucleotides.
|
The adenovirus group is a very large and heterogeneous group, including 51 different serotypes divided into six subgenera. To cover all six subgenera with one MLPA probe, we combined one long oligonucleotide probe with two synthetic oligonucleotides. This MLPA probe was tested with 15 different adenovirus serotypes covering the six subgenera. Included were serotypes 2, 4, 5, and 7, which are generally more often associated with RTIs (25), and serotypes 8, 9, 12, 19, 30, 31, 35, 37, 40, 48, and 49. All 15 serotypes could be detected by use of this probe (data not shown).
Specificity. To asses the specificity of the primers and the probes, all PCR primers and MLPA probes were tested in either a monoplex or a multiplex MLPA reaction with different samples. No cross-reactivity among the 15 viral probes and the IAC probe was observed when they were tested with culture samples of individual viruses. In addition, clinical samples which tested positive for other pathogens by routine diagnostic analysis by pathogen-specific PCR assays were analyzed by the RespiFinder assay. These samples contained enterovirus, echovirus, coxsackievirus, and herpes simplex virus type 1, as well as the bacterial pathogens Mycoplasma pneumoniae, Chlamydia pneumoniae, Legionella pneumophila, Bordetella pertussis, Staphylococcus warneri, Staphylococcus intermedius, Staphylococcus haemolyticus, Staphylococcus epidermidis, Enterococcus faecalis, Pseudomonas spp., Haemophilus influenzae, Haemophilus parainfluenzae, and Streptococcus spp. No sample showed any cross-reactivity with the RespiFinder assay probes (data not shown).
Internal control. The use of an IAC allows discrimination between a true-negative result and a false-negative result due to a PCR/MLPA failure. However, since the MLPA reaction uses only two primers in the final PCR, the amplification of any respiratory virus-specific MLPA probe competes with the amplification of the IAC MLPA probe. Figure 3 shows the results of the RespiFinder assay with fivefold serial dilutions of a PIV-1 culture, all of which were spiked with an equal amount of IAC. Only a weak IAC signal was detected in the presence of the highest PIV-1 concentration. As the PIV-1 concentration decreased, the IAC signal increased. This demonstrates competition between simultaneously amplified MLPA probes.
![]() View larger version (47K): [in a new window] |
FIG. 3. Detection of the IAC in spiked serial virus dilutions on a 2.5% agarose gel. RespiFinder assay analysis was performed with fivefold serial dilutions of the supernatant of a PIV-1 culture. Lanes: 1, undiluted; 2, diluted 51; 3, diluted 52; 4, diluted 53; 5, diluted 54; 6, diluted 55; 7, diluted 56; 8, blank. Each dilution except the blank was spiked with the IAC.
|
Second, the sensitivity of the RespiFinder assay was compared to that of the monoplex real-time PCR. Nucleic acid preparations from fivefold serial dilutions of culture supernatants of 13 viruses (no InfA H5N1 or adenovirus cultures were available) were tested. The results are shown in Table 3. In the monoplex real-time PCR, 20 µl of an RNA or a DNA sample was used, whereas 10 µl of an RNA or a DNA sample was used in the preamplification step of the RespiFinder assay. The results indicate that the sensitivity of the RespiFinder assay was comparable to that of the monoplex real-time RT-PCR.
|
View this table: [in a new window] |
TABLE 3. Sensitivity of RespiFinder assay compared to that of monoplex real-time RT-PCRa
|
![]() View larger version (8K): [in a new window] |
FIG. 4. Results of the RespiFinder assay with two clinical samples which were analyzed by capillary electrophoresis. Sample 1 represents a clinical sample with a rhinovirus (Rhino) infection. The electropherogram shows a signal at the position of the rhinovirus probe and the IAC. Sample 2 represents a clinical sample with no viral infection. Only the IAC signal was detected in the electropherograms. Contr, internal amplification control.
|
|
View this table: [in a new window] |
TABLE 4. Analysis of concordant results between RespiFinder assay and cell culture for 110 clinical samplesa
|
|
View this table: [in a new window] |
TABLE 5. Analysis of nonconsistent results between RespiFinder assay and cell culturea
|
|
View this table: [in a new window] |
TABLE 6. Comparison of RespiFinder assay and RSV-specific ICA results for 34 clinical samples
|
|
View this table: [in a new window] |
TABLE 7. Sensitivity and specificity of RespiFinder assay compared with those of cell culture and an RSV-specific ICAa
|
|
|
|---|
Here we present a new multiparameter test that is based on the MLPA technology and that is able to detect 15 respiratory viruses simultaneously. In order to obtain the required sensitivity, the MLPA technology was adapted by incorporating a preamplification step. This preamplification step involves an RT step and a 30-cycle target amplification step. Due to the introduction of the preamplification step, the hybridization time in the MLPA reaction could be reduced to 1 h, while the original MLPA protocol required a hybridization time of 16 h. As a result, the whole RespiFinder assay could be completed within a working day.
In the initial experiments, the RespiFinder assay products were analyzed on 2.5% agarose gels. However, due to the complexity of the RespiFinder assay, the resolution of the agarose gel was insufficient to distinguish all products amplified by MLPA. The RespiFinder assay reactions were also analyzed on a LabChip 90 system from Caliper LifeSciences. The resolution of this system was also not sufficient for the nonambiguous identification of all individual products amplified by MLPA. However, due to the clustering of probes of more related viruses, use of the LabChip 90 system allowed the detection of the individual probes for rhinovirus, adenovirus, and hMPV and the four probe clusters for influenza virus, PIV, RSV, and coronavirus. On high-resolution electrophoresis systems, such as the LI-COR 4300 DNA analyzer, the MegaBACE DNA analysis system, and the ABI 3100 genetic analyzer, all individual probes could be clearly distinguished.
Well-designed MLPA probes have the ability to discriminate between single-nucleotide polymorphisms (27). The viral MLPA probes require a more generic design, as viruses, especially RNA viruses, show a great deal of genetic variability. Therefore, a major concern in the design of the MLPA probes was the compatibility of virus-specific probes with as many known viral sequences as possible, including those of the serotypes common among humans. However, this generic design should not compromise the specificity of the probe. The probes were positioned in well-conserved regions. No mismatch within 5 nucleotides from the ligation site was tolerated. With the adenovirus-specific probe two short synthetic oligonucleotides were combined with one M13-derived long oligonucleotide probe to enclose all 51 serotypes of the heterogeneous adenovirus group. No cross-reactivity was detected with any of the probes.
The sensitivity and the specificity of the RespiFinder assay were assessed by comparison of the results of the RespiFinder assay to those of cell culture obtained with 110 clinical samples. All samples were spiked with the IAC. The spiking was done after addition of the lysis buffer in order to prevent degradation of the IAC. For one (1%) sample, the RespiFinder assay failed to produce a result. The RespiFinder assay detected 27 (24.5%) viruses which were missed by cell culture. These included 2 adenoviruses, 4 hMPVs, 2 InfAs, 1 PIV-1, 1 PIV-3, 11 rhinoviruses, and 6 RSVs. This increase in diagnostic yield is in accordance with the findings of other studies that used molecular methods (10, 23). One sample with a false-negative result for PIV-3 was observed by the RespiFinder assay. Among the eight adenoviruses detected, three were of different serotypes (serotypes 2, 3, and 7) representing two different subgenera (subgenera B and C). This illustrates the ability of the generic adenovirus probe to detect different serotypes in clinical samples. The RespiFinder assay found an incidence of coinfection of 7.8%, whereas cell culture did not detect any double infections. The percentage of mixed infections that we detected is in agreement with the percentages detected in other studies, which showed incidences of coinfections of 3.4% (34), 6.1% (21), and 18.6% (3).
Compared to the results of a commercial ICA-based rapid assay for RSV with 34 clinical samples, the RespiFinder assay detected 18 (52.9%) RSV infections, corresponding to a specificity of 82.4% and a sensitivity of 80%. The results for two RSV-positive samples were not confirmed. This included one sample which scored weakly positive by the rapid assay for RSV. The RespiFinder assay detected an InfA infection in this sample. In addition, the RespiFinder assay detected six additional RSV-positive samples. Additional viruses were detected in 52.6% of the ICA RSV-negative samples, including InfA, rhinovirus, Cor-OC43, and Cor-NL63.
In conclusion, we have developed a new multiparameter respiratory assay based on the MLPA technique for the detection of 15 respiratory viruses plus an ICA. The analytical sensitivity of the RespiFinder assay matches the sensitivity of the monoplex real-time RT-PCR. The validation with clinical samples shows that the RespiFinder assay gives an excellent overall performance. The assay does not need specialized equipment but makes use of the PCR and capillary electrophoresis systems generally available in molecular laboratories. The assay can be performed within one working day and is suitable for implementation in a diagnostic setting. Implementation of this assay for routine diagnostics allows the fast analysis of samples for a broad spectrum of pathogens at a reduced cost compared to the costs of individual monoplex real-time RT-PCR assays.
At present, the full analytical power of the RespiFinder assay can be used only when the results are analyzed on a high-resolution system like a capillary electrophoresis system. However, by adapting the design of the MLPA probes, it will also be possible to identify most probes on LabChip systems, like the Caliper LabChip 90 system, the Experion system (Bio-Rad Laboratories, Hercules, CA), or the 2100 bioanalyzer (Agilent Technologies, Foster City, CA). The MLPA technology allows the highly complex analysis of up to 45 fragments. This gives room for future expansion of the number of probes in the RespiFinder assay. Recently, four bacterial probes, including probes for Mycoplasma pneumoniae, Chlamydophila pneumoniae, Legionella pneumophila, and Bordetella pertussis, have been added to the RespiFinder assay. This allows the detection in a clinical sample of 19 pathogens and one IAC in one reaction. In addition, for better implementation in a diagnostic setting, the RespiFinder assay protocol was recently adapted by combining the probe ligation step and the probe amplification step in a single reaction. As a result, the RespiFinder assay protocol consists at present of four steps, i.e., preamplification (one-step RT-PCR), probe hybridization, probe ligation/amplification, and probe detection.
The research described here has been facilitated by a grant provided by the Dutch Ministry of Economic Affairs (grant TSGE3074).
Published ahead of print on 6 February 2008. ![]()