Journal of Clinical Microbiology, April 2008, p. 1510-1513, Vol. 46, No. 4
0095-1137/08/$08.00+0 doi:10.1128/JCM.01694-07
Copyright © 2008, American Society for Microbiology. All Rights Reserved.

ilnik,1
Kristina Fujs Komlo
,1
Ana Doplihar,1
David H. Persing,2 and
Mario Poljak1*
Institute of Microbiology and Immunology, Faculty of Medicine, University of Ljubljana, Ljubljana, Slovenia,1 Cepheid, Sunnyvale, California2
Received 25 August 2007/ Returned for modification 25 October 2007/ Accepted 25 January 2008
|
|
|---|
|
|
|---|
The latest development in the field of the molecular diagnosis of enteroviral meningitis is a fully automated real-time multiplex, reverse transcription-PCR assay, the GeneXpert enterovirus assay (GXEA; Cepheid, Sunnyvale, CA) (10). GXEA is, at present, the only FDA-approved assay for the qualitative detection of enterovirus RNA in CSF. It combines automated nucleic acid sample preparation, amplification, and real-time detection of enteroviral RNA in a disposable, macro/microfluidic cartridge using the GeneXpert Dx system instrument. To date, only one evaluation of GXEA has been published in peer-reviewed journals. A multicenter beta trial with 102 CSF samples obtained from patients with suspected meningitis (34 of whom were enterovirus positive) showed that GXEA had a sensitivity and a specificity of 97.1% and 100%, respectively, and that it is suitable for rapid, on-demand testing (10). In analytical studies, GXEA detected 63 of the enterovirus serotypes for which tests were conducted, with limits of detection ranging from 0.0002 to 200 50% tissue culture infective doses (TCID50s)/ml or 50% lethal doses/ml. It showed no cross-reactivity with 24 different microorganisms known to cause meningitis-like symptoms and tolerated the influence of interfering substances on test performance (10).
We prospectively evaluated GXEA over a 1-year period and compared it with two other PCR-based assays for the detection of enteroviral RNA in CSF. When we assess a new molecular diagnostic assay in the laboratory, the candidate assay is usually first evaluated on proficiency panels, like the Quality Control for Molecular Diagnostics (QCMD) panel. If the candidate assay shows successful performance on the QCMD panels, it is then evaluated prospectively in routine settings by comparison of its results with those of the existing diagnostic methods for a certain period of time (usually 6 to 12 months).
GXEA was first evaluated on three Enterovirus QCMD panels from the years 1999/2000, 2002, and 2004, consisting of 35 panel samples in total. A plasma sample containing coxsackievirus A9 (QCMD 2002) was excluded from the analyses since GXEA is recommended for the testing of CSF only. All 8 enterovirus-negative panel samples tested GXEA negative, while enterovirus RNA was detected by GXEA in 25 of 26 enterovirus-positive panel samples. Enterovirus-positive samples consisted of different 10-fold dilution series of poliovirus 2; coxsackieviruses A9, A16, and B5; ehoviruses 6, 9, and 11; and enterovirus 71 at concentrations ranging from 0.03 to 25,000 TCID50s/ml. The detailed compositions of the three enterovirus QCMD panels included in the study are available on the QCMD home page (http://www.qcmd.org/). A single false-negative GXEA result was obtained when panel sample EV-B07 from the 1999/2000 QCMD panel containing 0.036 TCID50s/ml of coxsackievirus A9 was tested. This result was not unexpected, since this particular sample tested enterovirus negative in all three QCMD reference laboratories and was reported to be enterovirus negative by 40 (81.6%) of 49 participating laboratories (23).
Following the successful performance of GXEA on QCMD panels, the assay was implemented in the daily routine testing of CSF samples. GXEA was evaluated in parallel with an in-house real-time PCR assay, as described previously (8). Briefly, after RNA isolation from 140 µl of CSF by use of the QIAamp viral RNA mini kit (Qiagen, Hilden, Germany), 5 µl of RNA was amplified with previously described primers targeting 120 bp of the 5' noncoding region and with a LightCycler RNA master hybridization probes kit (Roche Applied Science, Mannheim, Germany) on a LightCycler (version 2.0) instrument (Roche Applied Science). Thus, each CSF sample was tested for the presence of enterovirus RNA by using GXEA and an in-house real-time PCR on the same day. Since the in-house real-time PCR assay had been implemented in the daily routine only 7 months before the start of the GXEA evaluation and because of a lack of an internal control for the in-house real-time PCR, all CSF samples were additionally tested once weekly by using the enterovirus consensus kit (Argene SA, Varilhes, France), following the manufacturer's instructions, as described previously (6). Briefly, 10 µl of RNA previously extracted for the in-house real-time PCR and stored for a maximum of 1 week at –70°C was amplified by using a one-step reverse-transcription PCR and primers targeting a 425-bp region of the 5' noncoding region, followed by detection of PCR products in a microtiter plate by use of a biotinylated probe. The in-house real-time PCR and Argene assay were performed by two technicians with substantial experience in molecular diagnostics, while GXEA testing was performed by eight operators who mainly did not have any experience in molecular diagnostics.
A total of 162 CSF samples were tested prospectively for the presence of enterovirus RNA by using three different PCR assays from 15 July 2006 until 14 July 2007. Fifteen of 162 samples had to be retested due to different performance problems in the initial run. Seven samples were retested by using GXEA due to an initial invalid run because of internal control failure, while 11 samples (3 of them retested by GXEA) were retested by the Argene assay due to initial negative internal control results (8 samples) or indeterminate enterovirus results (3 samples). The enteroviral RNA status of all 15 samples with initial performance problems by GXEA and the Argene assay was resolved successfully after repeat testing. As shown in Table 1, 149 (91.9%) of 162 samples initially tested concordantly by all three assays. The initial sensitivity and specificity of GXEA, the Argene assay, and the in-house real-time PCR are shown in Table 2. All samples with discordant results were retested by all three assays. Of five initially Argene assay-negative samples, three samples became positive on repeat testing, while two samples remained enterovirus negative by repeat testing by the Argene assay (Table 1). The results for the last two samples which repeatedly tested negative by the Argene assay but which repeatedly tested positive by GXEA and the in-house reverse transcription-PCR were finally considered to be false negative by the Argene assay. For both patients, these results were supported by clinical diagnoses of enteroviral meningitis; and their CSF samples tested negative for herpes simplex virus DNA, cytomegalovirus DNA, varicella-zoster virus DNA, Epstein-Barr virus DNA, and human herpesvirus 6 DNA. These samples were also negative for immunoglobulin M and RNA tick-borne encephalitis virus. Of eight samples initially negative by GXEA, seven tested enterovirus positive, while one sample remained GXEA negative upon repeat testing (Table 1). The result for the latter sample, which repeatedly tested GXEA negative but which repeatedly tested Argene assay and real-time PCR positive, was finally considered to be a false-negative result by GXEA. This result was supported by a clinical diagnosis of enteroviral meningitis, and CSF and blood samples tested negative for all previously listed viruses. The most surprising finding in this evaluation was that seven initially GXEA-negative samples all became highly positive upon repeat testing. Although these seven samples were initially reported to be enterovirus negative by the GeneXpert Dx instrument, we noticed after the first such discrepant result (due to the nonblinded nature of our comparative evaluation) that the end-point fluorescence in these samples showed some evidence of an enteroviral amplification signal, with the signals ranging from 3 to 15. In contrast, the results for the concordant enterovirus-negative specimens and negative controls invariably showed enterovirus-specific end-point fluorescence values of 0. GXEA incorporates a coamplified control for the detection of inhibition; this comprises an "armored RNA" in-process control which is added at low levels to the sample before it is extracted, in order to control for the efficiency of both extraction and amplification. Lower-than-expected end-point values could be the result of the presence of PCR inhibitors or operator error (pipetting errors, incomplete or incorrect reagent dispensing into the GeneXpert cartridge). The presence of a PCR inhibitor(s) extremely labile to freezing-thawing might be more likely, since repeat testing after a freeze-thaw cycle when freezing lasted at least 1 h resolved the results for all the discrepant samples with end-point fluorescence values above 0 but below the threshold for a positive result. In contrast, repeat testing of these samples on the same day or without a freeze-thaw cycle or sample dilution 1:10 again resulted in end-point fluorescence values above 0 but below the threshold for a positive result. According to our results, all enterovirus-negative samples with end-point fluorescence values above 0 should be checked, and such values should be used as an indication for the need for a repeat test. In our experience, this additional check took less than a minute and could be performed reliably by individuals without a strong background in molecular diagnostics. Further evaluation of GXEA results with lower-than-expected end-point values from other parts of the world (if they are recognized) could be important, since this phenomenon might be due to different lumbar puncture practices in different regions or other factors.
|
View this table: [in a new window] |
TABLE 1. Initial and resolved test results for enterovirus RNA by GXEA, the in-house real-time PCR, and the Argene assay with 162 CSF samples
|
|
View this table: [in a new window] |
TABLE 2. Initial and resolved sensitivities and specificities of GXEA, the in-house real-time PCR, and the Argene assay for detection of enterovirus RNA in 162 CSF samples
|
In conclusion, GXEA, the only FDA-approved assay for the detection of enteroviral RNA in CSF, is an important new tool for the diagnosis of enteroviral meningitis. It delivers enterovirus PCR results on a STAT basis and meets a critical patient need for definitive diagnostic results in the evaluation of meningitis. Our study, the first to be published for a series of European patients, confirms the previously described levels of sensitivity and specificity of GXEA (10). In our experience, further improvements in the performance of GXEA were obtained by manually checking enterovirus end-point fluorescence values. For infrequent results for which enterovirus end-point values are above 0 but below the GXEA threshold for a positive result, we suggest repeating the test after a short freeze-thaw cycle. After 1 year of experience with GXEA in a routine laboratory setting, we consider GXEA to be an excellent system for routine or "on-demand" testing for enteroviral meningitis due to its complete automation and its rapid-result capability.
ka 4, Ljubljana 1000, Slovenia. Phone: 386 1 543 7453. Fax: 386 1 543 7418. E-mail: mario.poljak{at}mf.uni-lj.si
Published ahead of print on 6 February 2008. ![]()
|
|
|---|