Journal of Clinical Microbiology, April 2008, p. 1517-1519, Vol. 46, No. 4
0095-1137/08/$08.00+0 doi:10.1128/JCM.02046-07
Copyright © 2008, American Society for Microbiology. All Rights Reserved.

Laboratory of Clinical Microbiology and Virology, University Hospital Tor Vergata, Viale Oxford, 81-00133 Rome, Italy,1 IRCCS IFO-San Gallicano Institute, Rome, Italy2
Received 19 October 2007/ Returned for modification 25 January 2008/ Accepted 1 February 2008
|
|
|---|
|
|
|---|
Quantification by real-time PCR is based on the determination of the threshold cycle (CT) when the amplified product is detected for the first time (4) and the PCR is still in the exponential phase. In this case the quantification of the viral load is much more accurate and reliable than that measured with endpoint PCR systems (3).
The aim of our study was to evaluate the Abbott RealTime HBV assay for the detection and quantification of HBV-DNA in serum samples and to compare these results with those obtained using the Cobas AmpliPrep/Cobas TaqMan HBV test (Roche Diagnostics). The study was carried out on 128 serum samples collected from patients with chronic HBV infections attending our hepatology clinic at the University Hospital Tor Vergata, Rome, Italy.
Blood samples were collected in a BD Vacutainer tube containing a separation gel. After centrifugation, the serum was divided into two aliquots of 1 ml each and tested by the Abbott and Cobas TaqMan Real-time PCR assays, respectively. Both systems use a hybridization probe with a fluorescent moiety covalently linked to the 5' end of the probe (reporter) and a quenching moiety bound to the 3' end of the probe (quencher). In the presence of target, probe hybridization and primer extension occur simultaneously. The fluorescent signal is generated by removing the reporter through the 5'
3' exonuclease activity of a thermostable Taq DNA polymerase (4, 5). In the TaqMan and Abbott RealTime PCR assays an additional probe is used to detect genotype G or C, respectively. While the Cobas TaqMan system targets the pre-core region, the Abbott RealTime assay amplifies a highly conserved region of the HBV genome which is represented by the N-terminal portion of the S gene.
With the Abbott assay, HBV DNA was extracted from 500 µl of serum by using the m2000sp, an automated sample preparation system designed to use magnetic microparticle-based reagents for the purification of nucleic acids from samples. An internal control DNA is introduced into the sample preparation procedure and is processed along with the calibrators, controls, and specimens. The presence of the amplified HBV-DNA and internal control is detected during the extension and annealing step. The amplification cycle at which a normalized fluorescent signal is detected by the Abbott m2000rt is proportional to the log of the HBV DNA concentration present in the original sample. Each sample is then quantitated by an external calibration curve.
The Cobas AmpliPrep extracts HBV-DNA from 850 µl of serum, and during the extraction a known number of HBV quantitation standard DNA molecules are introduced in each specimen (5). The quantitation of HBV-DNA is performed using the HBV quantitation standard. The system quantifies the amplicons during the exponential phase of amplification. The PCR cycle at which the increase in normalized fluorescence of a sample exceeds the background noise is considered the critical CT and marks the beginning of the exponential growth phase of this signal.
Based on a positive Cobas TaqMan result, 128 serum samples were selected from patients with chronic hepatitis B, including samples with values below the limit of detection (LOD), but with viral target detected (<12 IU/ml). All but one sample (<12 IU/ml as determined by TaqMan) were detected by the Abbott RealTime assay, with a relative sensitivity of 99.2% (127 of 128; confidence interval = 97.69 to 100.00%).
As shown in Table 1, both assays provided a result within their dynamic range for 100 samples, while 14 samples were below the LOD for both assays. Of the remaining eight samples with a TaqMan result below LOD, seven were quantifiable by the Abbott RealTime assay (range, 12 to 35 IU/ml), and one produced a negative result. Of the two samples quantified by Roche TaqMan, but below its linearity range (54 IU/ml), one was quantified (84 IU/ml) and one was <10 IU/ml with the target detected by Abbott. Finally, four samples with a value exceeding the upper limit of linearity with Roche (>8.04 log IU/ml) had reportable results with the Abbott assay, whose upper limit of linearity reaches 9.0 log IU/ml. For three of the samples, the Abbott assay result was consistent with the TaqMan assay off-scale limit.
|
View this table: [in a new window] |
TABLE 1. Comparison between Abbott RealTime HBV and Cobas TaqMan 48 HBV real-time PCR assays
|
![]() View larger version (12K): [in a new window] |
FIG. 1. Linear regression analysis performed on 100 paired quantitative results obtained with the two assays.
|
In order to further verify the assay accuracy, the QCMD HBV06-01 panel (a past program proficiency panel, organized by the Quality Control for Molecular Diagnostics) was tested with the Abbott RealTime HBV assay. The panel consisted of eight members targeted at 3, 4, 5, and 6 log copies/ml (type D) and 3, 4, and 5 log copies/ml (type A) plus a negative sample (Fig. 2).
![]() View larger version (38K): [in a new window] |
FIG. 2. QCMD HBV06-01 panel tested with the Abbott m2000 HBV assay. The Abbott RealTime HBV results (measured) are compared to expected sample concentrations (target) and the mean results obtained by all 109 participants (consensus).
|
The assay reproducibility was assessed by testing in five separate different runs two four-member panels created by diluting two high-titer patient samples.
The coefficient of variation ranged from 0.77 to 2.93% (log copies/ml) for concentrations between 4.27 and 8.12 log copies/ml. The corresponding standard deviation (range, 0.06 to 0.19) was always below 0.25 log copies, a value that is recognized as the maximum allowed for analytical variability.
In conclusion, this evaluation demonstrated that in comparison to the Cobas AmpliPrep/Cobas TaqMan, the newly developed Abbott RealTime HBV-DNA assay has at least equivalent sensitivity and has a slightly wider dynamic range, and its results correlate extremely well. A small bias between the two assays was observed, most probably due to a different standardization process. Both assays, by coupling an exquisite sensitivity to a large linear range, are useful in monitoring the therapy of patients with chronic hepatitis B.