Journal of Clinical Microbiology, April 2008, p. 1523-1525, Vol. 46, No. 4
0095-1137/08/$08.00+0 doi:10.1128/JCM.02100-07
Copyright © 2008, American Society for Microbiology. All Rights Reserved.

Clinical Microbiology-Immunology Laboratories, University of North Carolina Hospitals, and Departments of Microbiology-Immunology and Pathology-Laboratory Medicine, University of North Carolina School of Medicine, Chapel Hill, North Carolina 27514
Received 31 October 2007/ Returned for modification 11 December 2007/ Accepted 25 January 2008
|
|
|---|
|
|
|---|
Given the emergence of a highly virulent C. difficile strain in the United States, a recent report by Ticehurst and colleagues was a cause for concern (18). They showed that another widely used EIA, the C. difficile Tox A/B II assay (TechLab, Blacksburg, VA), which has performance characteristics similar to those of the A/B EIA (11, 12, 19), had a sensitivity of only 36% for C. difficile detection when the results were compared to those obtained by the use of a two-step algorithm in which specimens were first tested for the presence of the glutamate dehydrogenase (GDH) antigen by a solid-phase EIA (C. Diff Chek-60; TechLab) and the positive results were confirmed by the CTN assay. The test for the GDH antigen proved to be an excellent screening test, with a sensitivity of >95% and a negative predictive value of >99% compared with the results of the CTN assay (18). In practical terms, that meant that GDH-negative specimens could be reported to be negative without further testing. By using the CTN assay as a reference method, it was found that the assay for the GDH antigen had a positive predictive value of 53%, indicating that a confirmatory test was required to detect specimens containing C. difficile toxins.
I applied the two-step algorithm approach using a modification of the GDH antigen assay, which used a recently developed membrane-bound lateral flow device, C. Diff Quik Chek (TechLab), and compared its performance characteristics to those of the A/B EIA with a membrane-bound lateral flow device (Tox A/B Quik Chek; TechLab) for the detection of toxins A and B (the LF-A/B assay).
Three hundred sixty-eight clinical specimens submitted between April and June 2007 for the detection of C. difficile toxin by the A/B EIA were studied. The inclusion criteria were as follows: a clinician had to order the detection of C. difficile toxin in a specimen, the specimen had to be less than 24 h old, the specimen had to take the form of the specimen container, and the specimen volume had to be sufficient to allow performance of the four assays that were used in this study. All specimens were stored at 4°C prior to assay. The A/B EIA, the C. Diff Quik Chek assay (the GDH assay), the LF-A/B assay, and the CTN assay with the TechLab C. difficile toxin B assay were all done according to the manufacturer's instructions, with the exception that the results for specimens with optical densities at 450/630 nm of 0.1 to 0.5 by the A/B EIA were confirmed by the CTN assay. In-house data have shown that only 30% of specimens with optical density values between 0.1 and 0.5 by the A/B EIA could be confirmed to be positive by the CTN assay. Positive controls were run on each day of the assay for all tests. In addition, a negative control was run for the A/B EIA, while each of the GDH and LF-A/B assays has an internal negative control. The individual performing the GDH and the LF-A/B assays was blinded to the A/B EIA results.
For the purposes of comparisons of the algorithm with the A/B EIA and the LF-A/B assays, specimens were assigned a true-positive result by the two-step algorithm if they were both GDH assay positive and CTN assay positive; they were assigned a true-negative result if they GDH assay negative or GDH assay positive but CTN assay negative. These designations are based on the high sensitivity and the high negative predictive value previously reported for this algorithm (18). Confidence intervals were determined by the use of EP Evaluator software (David G. Rhoads Associates, Kennett Square, PA). This study was approved by the Institutional Review Board of the University of North Carolina.
Of the 368 specimens studied, 70 were positive and 298 were negative by the GDH assay.
Of the 70 GDH assay-positive specimens, 37 were confirmed to contain C. difficile toxin by the CTN assay. Only 22 of the 37 specimens confirmed to have positive results were positive by the A/B EIA, resulting in an A/B EIA sensitivity of 59.5%. The sensitivity was even lower for the LF-A/B assay (43%). The positive predictive values, the negative predictive values, and the specificities of the EIAs were all very good to excellent (Table 1). Two specimens were found to be GDH assay negative but A/B EIA positive. Both specimens were negative by the CTN assay. The results for five specimens found to be positive by the LF-A/B assay could not be confirmed by the CTN assay; two of the five specimens were also positive by the GDH assay.
|
View this table: [in a new window] |
TABLE 1. Performance characteristics of A/B EIA and the LF-A/B assay compared with those of a two-step algorithm using C. Diff Chek-60 and C. difficile toxin B assays
|
What are some of the practical observations made during this study? First, the LF-A/B assay had performance characteristics very similar to those reported for the C. Diff Chek-60 assay (16, 18) in that it had a positive predictive value of only 53% but a sensitivity of 100% compared to tissue culture results. Because the results for all negative specimens were not confirmed by the CTN assay, the specificity and the negative predictive value of this GDH assay cannot be stated with certainty. However, 293 of 298 of the GDH assay-negative specimens were also negative by the two other immunoassays that were studied, with the A/B EIA having been shown to be highly specific and to have a high negative predictive value in several studies (9, 11, 12, 15, 19). The seven GDH assay-negative specimens that were positive by one of the two other EIAs were found to be negative by the CTN assay. These data suggest that the LF-A/B assay does have a very high negative predictive value, a crucial characteristic of any screening test, but because of the study design, this cannot be proven conclusively.
Second, the GDH screening test is easily performed and the results are easily interpreted. The test takes approximately 20 min to perform and is very simple, and the filters clogged for less than 5 of 368 specimens tested. There was one failed run during the study because the reagents were not added in the proper order. This was determined by failure of the internal control. I used positive controls on each day of use. Except for the failed run, the positive control gave an appropriate reaction each time, suggesting that positive controls need to be tested only once per test lot. My laboratory can currently offer this testing three times per day, and stat testing is also easily accomplished. My laboratory can offer testing by the A/B EIA only once a day, and stat testing is inconvenient and is rarely done.
Third, this approach is more expensive, and for an estimated 20% of specimens that require the CTN assay, the turnaround time is delayed compared to the turnaround time for the EIA. The material costs per specimen for the A/B EIA were approximately $5.00, while the current per specimen cost for the LF-A/B assay is $9.00. In addition, 20% of specimens require the CTN assay, which has an approximate cost of $8.00. These specimens require both centrifugation and filtration, which are labor-intensive and which thus add further to the cost of this algorithm. The overall material cost per specimen for the two-step algorithm is $10.60. In order to provide the best possible turnaround time, testing by the CTN assay must be offered 7 days/week.
A major question concerning this study is the explanation for the discrepancy between the high sensitivities (>90%) for the A/B EIA reported recently (11, 20) and the comparatively modest sensitivity found in this study (59.5%). It is possible that there were systematic errors in the performance of this assay during the study period. However, quality control failures have been rare, and the lack of proficiency test failures since the institution of this assay in my laboratory makes this type of error less likely. Alternatively, there may have been a regional shift in the antigenic nature of the toxin, resulting in the EIA becoming less sensitive. I did notice that several of the GDH assay- and CTN assay-positive but A/B EIA-negative specimens were CTN positive only after 48 h of incubation, suggesting the presence of low levels of toxin and perhaps explaining the inability of the EIA to detect the toxin. However, a similar number of EIA-negative specimens were GDH and CTN assay positive at 24 h, suggesting that a low toxin level was not the entire explanation for the relative insensitivity of the A/B EIA.
The data presented here and by others (2, 6, 9, 16, 18) indicate that GDH detection is the most sensitive and convenient method by which C. difficile can be detected in stools. The assay also has a high negative predictive value. Because of this test's modest positive predictive value of approximately 50%, a rapid, convenient confirmatory test for GDH-positive specimens is needed before this test is likely to gain wide acceptance in clinical laboratories. The data presented here and by others (18) indicate that two solid-phase EIAs, the C. difficile Tox A/B II EIA (TechLab) and the A/B EIA (Meridian Bioscience), as well as a single-test lateral flow immunochromatographic test, the LF-A/B assay (TechLab), lack sufficient sensitivity compared with the results of the CTN assay to be used as confirmatory tests. For laboratories wishing to implement screening for C. difficile by the GDH assay as part of a testing algorithm, for now it is necessary to use either the CTN assay or toxigenic culture (5, 6, 14) as a confirmatory test for GDH assay-positive specimens.
I thank TechLab for providing the kits used in this study.