Journal of Clinical Microbiology, June 2008, p. 2088-2091, Vol. 46, No. 6
0095-1137/08/$08.00+0 doi:10.1128/JCM.00126-08
Copyright © 2008, American Society for Microbiology. All Rights Reserved.

APHP, Hôpital Bichat—Claude Bernard, Laboratoire de Virologie, Paris, 75018 France,1 INSERM, U897.ISPED, Université Victor Segalen Bordeaux 2, Bordeaux, F-33076, France,2 AIDS Reference Laboratory, Université Catholique de Louvain, Louvain-la-Neuve, Belgium,3 Medical Research Council Laboratories, Banjul, Gambia,4 Institute of Virology, Bonn, Germany,5 Laboratório de Biologia Molecular, CHLO—Hospital Egas Moniz, Lisboa, Portugal,6 Department of Infectious Diseases, Hospital Carlos III, Madrid, Spain,7 Department of Virology, Swedish Institute for Infectious Disease Control, Solna, and Department of Microbiology, Tumor and Cell Biology, Karolinska Institutet, Stockholm, Sweden,8 Swiss HIV Cohort Study, Zurich, Switzerland,9 Centre for Virology, Royal Free & University College London Medical School, London, United Kingdom,10 APHP, Hôpital Bichat—Claude Bernard, Service de Maladies Infectieuses et Tropicales, Paris, France,11
Received 22 January 2008/ Accepted 14 April 2008
|
|
|---|
|
|
|---|
|
View this table: [in a new window] |
TABLE 1. Characteristics of the different quantification assays assessed by the ACHIEV2E Collaboration in 2006
|
The accuracy and reproducibility of assays were estimated for the three theoretical concentrations. A quantification assay was defined as accurate if at least 9 of the 10 measurements fell within a clinically acceptable interval based on the NIHz theoretical concentrations. The interval was defined as one-third of the theoretical viral load to three times the theoretical viral load for theoretical viral loads of 3.0 and 2.3 log10 copies/ml and as 0 to three times the theoretical viral load for a theoretical viral load of 1.7 log10 copies/ml.
Reproducibility of assays was evaluated by using the intralaboratory coherence coefficient (ILCC) and the coefficient of variation (CV) at each of the three theoretical concentrations (1). A quantification heterogeneity was defined as an ILCC greater than the theoretical value L, extracted from a reference table for p laboratories and n repetitions (1).
Each of the participating laboratories quantified 35 coded and randomized aliquots. All laboratories reported an undetectable viral load for the five HIV-negative aliquots.
At a theoretical level of 3.0 log10 copies/ml, median quantifications varied from 2.9 log10 copies/ml in laboratory 6 to 4.2 log10 copies/ml in laboratory 8 (Fig. 1A). Five laboratories reported accurate measurements (i.e., 9 of the 10 quantifications fell in the interval from 2.52 to 3.48 log10 copies/ml). The four other assays overquantified viral loads (laboratories 1, 3, 7, and 8) (Fig. 1A). Seven laboratories reported reproducible measurements (CV = 3 to 6%; ILCC = 0.5 to 0.9; L = 1.35) (laboratories 1, 2, 3, 6, 7, 8, and 9). Laboratories 2, 6, and 9 reported both accurate and reproducible measurements. At a theoretical level of 2.3 log10 copies/ml, the accuracy and reproducibility of the tests at laboratories 4 and 5 could not be evaluated since 10 and 9 negative results were reported, respectively. Median quantifications varied from 2.4 log10 copies/ml in laboratory 9 to 3.4 log10 copies/ml in laboratory 8 (Fig. 1B). Three laboratories reported accurate measurements (i.e., 9 of the 10 quantifications fell in the interval from 1.82 to 2.78 log10 copies/ml), and four laboratories overquantified viral loads (laboratories 1, 3, 7, and 8) (Fig. 1B). Six laboratories reported reproducible measurements (CV = 5 to 8%; ILCC = 0.5 to 1.1; L = 1.35). Only two laboratories, 2 and 9, reported both accurate and reproducible measurements.
![]() View larger version (17K): [in a new window] |
FIG. 1. Accuracy of quantification assays evaluated by the ACHIEV2E Collaboration in 2006. Quantification results are reported for each participating laboratory. The accuracy interval is represented by the white area for each of the three theoretical viral loads used. Panels: A, theoretical viral load = 3.0 log10 copies/ml; B, theoretical viral load = 2.3 log10 copies/ml; C, theoretical viral load = 1.7 log10 copies/ml.
|
Our study is the first international evaluation of the validity of HIV-2 RNA quantification assays. We found a fair homogeneity of results of assays evaluated at a theoretical level of 3.0 log10 copies/ml and a decreasing accuracy and reproducibility of assays with decreasing viral loads, as previously shown in a quality control study of HIV-1 RNA quantification assays (6). Overall, four laboratories systematically overquantified viral loads, no. 1, 3, 7, and 8. Two laboratories, 4 and 5, reported 18 negative results out of 20 quantifications at theoretical viral loads of both 2.3 and 1.7 log10 copies/ml. The thresholds of these two assays had been evaluated at 2.7 and 2.3 log10 copies/ml, respectively. The cutoff values of the methods used by laboratories 5 and 7 may also explain their results for the 1.7-log10-copy/ml aliquots. One laboratory (no. 6) reported both accurate and reproducible results at a theoretical viral load of 3.0 log10 copies/ml but heterogeneous quantifications at 2.3 log10 copies/ml and negative results at 1.7 log10 copies/ml. Two laboratories, 2 and 9, reported both accurate and reproducible results at theoretical viral loads of 3.0 and 2.3 log10 copies/ml and accurate but relatively poorly reproducible quantifications at 1.7 log10 copies/ml. The aliquot were made from the NIHz HIV-2 supernatant, which is the standard used for the calibration of the laboratory 2 assay. That may have improved the results reported by this laboratory and decreased those observed in other laboratories such as no. 8, which has not yet been evaluated with calibrated specimens. However, laboratory 9, which did not use NIHz as a standard, reported results very similar to those obtained by laboratory 2.
This study provides an overview of the performance of different HIV-2 RNA quantification assays. If the HIV-2 supernatant counted by electron microscopy is accepted as a standard, only two laboratories reported accurate and reproducible measurements at both 3 and 2.3 log10 copies/ml. This heterogeneity may yield difficulties in the comparison of results between the different cohorts of the ACHIEV2E Collaboration. The ACHIEV2E network is working toward the standardization of quantification assays to improve the interpretation of results for case management and collaborative clinical trials.
The members of the ACHIEV2E study group in Belgium are Patrick Goubau and Jean Ruelle (AIDS Reference Laboratory, Université Catholique de Louvain). Those in France are Françoise Brun-Vezinet, Pauline Campa, Florence Damond, Diane Descamps, Sophie Matheron (Bichat—Claude Bernard Hospital, Paris), François Simon (Saint-Louis Hospital, Paris), Antoine Benard, Geneviève Chene, Audrey Taieb (INSERM U897, Bordeaux2 University, Bordeaux), and Brigitte Autran (Pitié-Salpétrière Hospital, Paris). Those in Gambia are Abraham Alabi, Matt Cotten, Assan Jaye, Kevin Peterson, and Sarah Rowland-Jones (Medical Research Council Laboratories). Those in Germany are Bernd Kupfer (Institute of Virology, Bonn), Jürgen Rockstroh, and Carolynne Schwarze-Zander (Department of Internal Medicine I, University of Bonn, Bonn). Those in The Netherlands are Frank De Wolf, Ard Van Sighem (National AIDS Therapy Evaluation Center, AMC, Amsterdam), Peter Reiss, Maarten Schim Van Der Loeff (AMC-CPCD, Amsterdam; MRC, Gambia), and Martin Schutten (Department of Virology, Erasmus MC, Rotterdam). Those in Portugal are Ricardo Camacho, Perpetua Gomes, Kamal Mansinho (Egas Moniz Hospital, Lisbon), Francisco Antunes, Luis Franca, and Emilia Valadas (Clinica Universitaria de Doenças Infecciosas, Lisbon). Those in Spain are Berta Rodes, Carlos Toro, and Vicente Soriano (Department of Infectious Diseases, Hospital Carlos III, Madrid). Those in Sweden are Jan Albert (Department of Virology, Swedish Institute for Infectious Disease Control, Solna), Katarina Gyllensten, Anders Sonnerborg (Divisions of Clinical Virology and Infectious Diseases, Huddinge University Hospital, Karolinska Institute, Stockholm), Aylin Yilmaz, and Magnus Gisslén (Department of Infectious Diseases, Sahlgrenska University Hospital, Göteborg). Those in Switzerland are Jürg Böni, Alexandra Calmy, and Martin Rickenbach (Swiss HIV Cohort Study). Those in the United Kingdom are Deenan Pillay, Bridget Ferns, Jeremy Garson (Centre for Virology, Royal Free & University College London Medical School, London), Jennifer Tosswill (Centre for Infections, Health Protection Agency, London), Jane Anderson (Centre for the Study of Sexual Health and HIV, Homerton University Hospital, London), and David Chadwick (James Cook University Hospital, Middlesbrough).
Published ahead of print on 23 April 2008. ![]()
|
|
|---|