Journal of Clinical Microbiology, July 2008, p. 2327-2333, Vol. 46, No. 7
0095-1137/08/$08.00+0 doi:10.1128/JCM.00075-08
Copyright © 2008, American Society for Microbiology. All Rights Reserved.
Direct Comparison of the BD Phoenix System with the MicroScan WalkAway System for Identification and Antimicrobial Susceptibility Testing of Enterobacteriaceae and Nonfermentative Gram-Negative Organisms
J. W. Snyder,1,2*
G. K. Munier,2 and
C. L. Johnson2
Department of Pathology and Laboratory Medicine, University of Louisville School of Medicine,1
University of Louisville Hospital, Louisville, Kentucky 402022
Received 14 January 2008/
Returned for modification 12 March 2008/
Accepted 13 May 2008

ABSTRACT
The Phoenix automated microbiology system (BD Diagnostics, Sparks,
MD) is designed for the rapid identification (ID) and antimicrobial
susceptibility testing (AST) of clinically significant human
bacterial pathogens. We evaluated the performance of the Phoenix
instrument in comparison with that of the MicroScan WalkAway
system (Dade Behring, West Sacramento, CA) in the ID and AST
of gram-negative clinical strains and challenge isolates of
Enterobacteriaceae (
n = 150) and nonfermentative gram-negative
bacilli (NFGNB; 45 clinical isolates and 8 challenge isolates).
ID discrepancies were resolved with the API 20E and API 20NE
conventional biochemical ID systems (bioMerieux, Durham, NC).
The standard disk diffusion method was used to resolve discordant
AST results. The overall percentages of agreement between the
Phoenix ID results and the MicroScan results at the genus and
species levels for clinical isolates of
Enterobacteriaceae were
98.7 and 97.7%, respectively; following resolution with conventional
biochemical testing, the accuracy of the Phoenix system was
determined to be 100%. For NFGNB, the levels of agreement were
100 and 97.7%, respectively. Both systems incorrectly identified
the majority of the uncommon nonfermentative nonpseudomonal
challenge isolates recovered from cystic fibrosis patients;
these isolates are not included in the databases of the respective
systems. For AST of
Enterobacteriaceae, the rate of complete
agreement between the Phoenix results and the MicroScan results
was 97%; the rates of very major, major, and minor errors were
0.3, 0.2, and 2.7%, respectively. For NFGNB, the rate of complete
agreement between the Phoenix results and the MicroScan results
was 89.1%; the rates of very major, major, and minor errors
were 0, 0.5, and 7.7%, respectively. Following the confirmatory
testing of nine clinical isolates initially screened by the
MicroScan system as possible extended-spectrum-β-lactamase
(ESBL)-producing organisms (seven
Klebsiella pneumoniae isolates
and two
Escherichia coli isolates), complete agreement was achieved
for eight isolates (one ESBL positive and seven negative); one
false positive was obtained with the Phoenix instrument. The
MicroScan system correctly detected the 10 ESBL challenge isolates,
versus the 6 detected by the Phoenix system. Overall, there
was good correlation between the Phoenix instrument and the
MicroScan system for the ID and AST of
Enterobacteriaceae and
common NFGNB. The Phoenix system is a reliable method for the
ID and AST of the majority of clinical strains encountered in
the clinical microbiology laboratory. Until additional performance
data are available, results for all
Klebsiella pneumoniae or
Klebsiella oxytoca and
E. coli isolates screened and confirmed
as ESBL producers by any automated system should be confirmed
by alternate methods prior to the release of final results.

INTRODUCTION
Clinical microbiology laboratories are under increasing pressure
and scrutiny from clinicians and administrators to provide rapid,
accurate, and timely bacterial identification (ID) and antimicrobial
susceptibility testing (AST) results. The primary goals in providing
such information are to aid clinicians in the diagnosis and
treatment of infectious diseases, influence the appropriate
selection of antimicrobial therapy, monitor community and institutional
resistance patterns, and contribute to a reduction in health
care-associated costs. To accomplish these goals, many laboratories
have implemented the use of semiautomated microbial ID and AST
instruments that have been designed to reduce turnaround times,
increase efficiency, and improve cost-effectiveness (
14). The
Phoenix automated microbiology system (BD Diagnostic Systems,
Sparks, MD) is the most recent instrument to receive clearance
from the Food and Drug Administration (FDA) to provide rapid
ID and AST of gram-negative and gram-positive bacteria recovered
from human specimens.
In this study, we performed a side-by-side direct evaluation of the Phoenix instrument and our in-house automated system, the MicroScan WalkAway 96 system (Dade Behring, West Sacramento, CA). The objectives of the study were threefold: (i) to evaluate the performance and accuracy of the Phoenix instrument for the ID and AST, under daily routine laboratory operating conditions, of gram-negative clinical isolates from the family Enterobacteriaceae and nonfermentative gram-negative bacilli (NFGNB) that reflect our organism mix; (ii) to challenge the respective systems by testing a set of previously characterized NFGNB clinical isolates that are not listed in the databases of the respective systems; and (iii) to screen and detect extended-spectrum-β-lactamase (ESBL)-producing Escherichia coli and Klebsiella pneumoniae or Klebsiella oxytoca isolates, including a set of challenge ESBL isolates previously characterized and provided as unknowns by an independent laboratory.
(The findings of this study were presented in part at the 106th General Meeting of the American Society for Microbiology, Orlando, FL, 2006 [15].)

MATERIALS AND METHODS
Bacterial isolates.
A total of 195 fre
sh clinical isolates (150 strains of
Enterobacteriaceae and 45 strains of NFGNB) recovered from routine cultures in
the University of Louisville Hospital clinical microbiology
laboratory, Louisville, KY, were tested. To ensure that an appropriate
mix of species was tested, the numbers of
E. coli and
Pseudomonas aeruginosa isolates were restricted to 30 for each organism.
The following genera and species made up the bacterial mix:
E. coli (25 isolates),
Klebsiella pneumoniae (45 isolates),
Klebsiella oxytoca (9 isolates),
Enterobacter cloacae (25 isolates),
Enterobacter aerogenes (11 isolates),
Serratia marcescens (10
isolates),
Proteus mirabilis (18 isolates), and
Proteus vulgaris,
Morganella morganii,
Salmonella sp.,
Citrobacter freundii,
Pantoea (
Enterobacter)
agglomerans, and
Yokenella (
Koserella)
regensburgei (1 isolate each); representative NFGNB consisted of
Pseudomonas aeruginosa (26 isolates),
Acinetobacter baumannii (8 isolates),
Stenotrophomonas maltophilia (8 isolates), and one isolate each
of
Chromobacterium violaceum,
Kingella sp., and
Pseudomonas fluorescens or
Pseudomonas putida. A challenge set of eight
previously well characterized NFGNB isolates with low frequencies
of occurrence that had been recovered primarily from cystic
fibrosis patients consisted of
Ralstonia mannitolilytica (two
isolates),
Burkholderia cenocepacia (three isolates), and one
isolate each of
Burkholderia multivorans,
Burkholderia gladioli,
and
Pandoraea pnomenusa (
4). In addition to clinical isolates
of
E. coli and
Klebsiella pneumoniae, a set of 10 ESBL-producing
challenge isolates previously characterized by appropriate biochemical
or molecular methods, including isoelectric focusing, substrate
profiling, inhibitor profiling, plasmid isolation, recombinant
DNA techniques, sequencing, and transformation, were tested
by the respective automated systems to examine the abilities
of the systems to detect ESBL production. These isolates were
previously confirmed to harbor and produce β-lactamases
known to cause false-positive and false-negative Clinical and
Laboratory Standards Institute (CLSI) confirmatory test results,
such as high-level AmpC, SHV-1, and K1 β-lactamases and
class A carbapenemases, as well as multiple other β-lactamases
(up to five enzymes) (
11,
18).
All isolates were concurrently processed by the Phoenix and MicroScan systems for genus and species ID and the determination of the antimicrobial susceptibility category. Inoculum suspensions for the respective systems were processed on the same day using the same fresh 18- to 24-h-subculture plate. Separate purity plates were inoculated from each system's basic inoculum, and the resulting subcultures were used for discrepancy resolution (for discordant ID and AST results) and the assessment of culture purity.
Phoenix ID and AST.
Phoenix NMIC/ID-108 combo panels (combined ID and AST cards) were inoculated and incubated according to the manufacturer's recommendations. BDExpert software versions 3.34A and 3.54A were used in this evaluation. The bacterial ID method employs modified conventional fluorogenic and chromogenic substrates in a system of 45 microwells containing dried biochemical substrates and 2 fluorescent control wells (2, 8). A suspension of each isolate, following adjustment to a 0.5 McFarland standard by using a CrystalSpec nephelometer (BD Diagnostics), was poured into the designated ID chamber of the Phoenix panel after a 25-µl aliquot was removed for AST. The AST section of the combination panel consists of 84 microwells containing dried antimicrobials and one growth control well. The method is a broth-based microdilution test that utilizes a redox indicator to enhance the detection of organism growth. As previously indicated, 25 µl of the standardized ID broth suspension was transferred into the AST broth, which resulted in final concentration of approximately 5 x 105 CFU/ml. The Phoenix ESBL test is based on the detection of growth in the presence of cefpodoxime, ceftazidime, ceftriaxone, and cefotaxime with or without clavulanic acid. The BDExpert system (versions 3.34A and 3.54A) consists of a series of rules, which are activated by MIC data, by the identity of the bacterial species, or by the result of the ESBL test. At the time of this study, only rule no. 1505, "isolate is confirmed positive for extended-spectrum beta-lactamase; consultation with an infectious disease practitioner is recommended," was incorporated into the program, but interpretative rules for Citrobacter, Enterobacter, Morganella, Proteus, and Serratia were not available in the United States as a result of the FDA's restriction of testing for ESBL production to E. coli, Klebsiella pneumoniae, and Klebsiella oxytoca. All panels were sealed with a panel closure and loaded into the instrument. Quality control and maintenance were performed according to the manufacturer's recommendations. All organism preparations were logged and loaded into the instrument within the required 30-min timeline. Final results are available in 2 to 12 h for ID and 4 to 16 h for AST. AST results are expressed as an MIC with an accompanying category (susceptible, intermediate, or resistant) interpretation.
MicroScan ID and AST.
The MicroScan WalkAway 96 system served as the primary reference method for ID and AST. ID and AST (using breakpoints with category interpretations) were performed with Neg BP combo panel type 30. All procedures were performed according to the manufacturer's recommendations. The screening of ESBL-producing microorganisms and interpretation were based on rules contained within the LabPro Expert system (version 2.0), which uses growth in the presence of cefpodoxime (4 µg/ml) and ceftazidime (1 µg/ml) at concentrations recommended by the CLSI for ESBL screening (3). The FDA limits screening for ESBLs by this system to E. coli, Klebsiella pneumoniae, and Klebsiella oxytoca isolates, as outlined in the CLSI guidelines (3). Other members of the family Enterobacteriaceae, such as Citrobacter spp., Enterobacter spp., Serratia spp., and members of the Proteus group (for which FDA approval and CLSI guidelines for testing were added in 2007), are known to harbor AmpC β-lactamase enzymes and may additionally produce ESBLs that give a positive ESBL screening result. However, the expert system does not support the detection of derepressed AmpC β-lactamases and ESBL production in these organisms and does not alert the user to the possibility of ESBL production (20), although organisms with this type of β-lactamase would test as resistant.
The CSLI four-disk diffusion confirmatory method (including cefotaxime and ceftazidime with and without clavulanic acid) was used for the confirmation of ESBL phenotypes of the relevant clinical and challenge isolates of Klebsiella pneumoniae, Klebsiella oxytoca, and E. coli (3).
Discrepancy resolution.
The ID results from the Phoenix and MicroScan systems were compared to determine the agreement at the genus and species levels. The ID was considered to be correct when the results from the two systems agreed at both levels. If the genus- and/or species-level IDs from the two systems were not in agreement, the isolate was retested in duplicate by both systems. If the discrepancy was not resolved, the organism was tested with the API 20E or API 20NE (internal arbitration). For the strains for which the Phoenix or MicroScan result was concordant with the third (internal-arbitration) result, this ID was considered to be correct. If the third result (from internal arbitration) did not agree with either of the original discrepant IDs, the isolate in question was excluded from the study.
Discrepant AST results were resolved by retesting the isolate in parallel with both systems, and the respective expert systems were applied for category concordance for the repeat results. If the AST results remained discrepant, the isolate was tested by the CLSI disk diffusion method (3). Rates of errors, categorized as very major errors (VMEs), major errors (MEs), and minor errors (mEs), were calculated for each system, but only VMEs and MEs were resolved.

RESULTS
Bacterial ID.
Table
1 shows the results of testing of 150 isolates of
Enterobacteriaceae and 45 NFGNB. The Phoenix and MicroScan systems showed percentages
of agreement at the genus and species levels of 98.7 and 97.7%,
respectively, for the various isolates of
Enterobacteriaceae.
Two isolates of the
Enterobacteriaceae family required the resolution
of results by testing with the API 20E. One isolate was initially
identified by the MicroScan system as
Yokenella (
Koserella)
regensburgei and as
Hafnia alvei by the Phoenix system. The
MicroScan system identified the second isolate as
Serratia fonticola or
Enterobacter aerogenes, versus the Phoenix ID as
Enterobacter aerogenes or
Enterobacter cloacae. Following confirmatory testing,
the Phoenix system was found to have correctly identified both
isolates with discrepant results, which gave an overall accuracy
of the Phoenix system of 100% for the ID of
Enterobacteriaceae to the genus and species levels.
For the 45 NFGNB isolates, two discrepant IDs by the Phoenix
system were noted. One isolate, a mucoid strain of
Pseudomonas aeruginosa, was misidentified as
Pseudomonas putida or
Pseudomonas aeruginosa and
Pseudomonas fluorescens or
Pseudomonas putida.
Following confirmatory testing, the isolate was identified as
Pseudomonas aeruginosa. The second discrepant isolate, following
confirmatory testing, was found to have been correctly identified
by the Phoenix system as
Kingella kingae, versus an incorrect
ID by the MicroScan system as
Moraxella sp. The overall accuracy
of the Phoenix system for the ID of NFGNB was 100% at the genus
level and 97.7% at the species level.
Each system had difficulty in providing precise IDs of the eight challenge NFGNB isolates (Table 2). Both systems correctly identified the two strains of Ralstonia mannitolilytica to the genus level but misidentified both strains as Ralstonia pickettii. Of the five Burkholderia isolates, the MicroScan and Phoenix systems correctly identified one and three isolates, respectively, to the genus level; all five isolates were misidentified at the species level by both systems. The Phoenix system yielded a "no ID" message for the strains of Burkholderia multivorans and Burkholderia cenocepacia, whereas the MicroScan system generated a "no ID" result for two strains of Burkholderia cenocepacia. Pandoraea pnomenusa was incorrectly identified as Alcaligenes xylosoxidans and Moraxella sp. by the MicroScan and Phoenix systems, respectively. Burkholderia gladioli was misidentified by the MicroScan system as Acinetobacter baumannii, while the Phoenix system identified the isolate as Burkholderia sp. or Ralstonia sp.
Antimicrobial susceptibility.
Antimicrobial susceptibility results for the
Enterobacteriaceae and NFGNB isolates tested against individual antibiotics are
summarized in Table
3. Category agreement (CA) was defined as
matching classifications as susceptible, intermediate, or resistant
by the two systems. The level of CA for the
Enterobacteriaceae was 96.5%. The mE, ME, and VME rates were 2.4, 1.4, and 6.0%,
respectively. The majority of VMEs were associated with β-lactam
antibiotics. β-Lactam antibiotics also accounted for the
majority of MEs. Based on the indications of the MicroScan ESBL
screening software (version 2.0), nine clinical isolates (seven
Klebsiella pneumoniae and two
E. coli isolates) were further
screened as possible ESBL producers (Table
4). Following confirmatory
testing, six
Klebsiella pneumoniae isolates were determined
to be negative for ESBL production and one was found to be positive;
both
E. coli isolates were negative. The Phoenix system correctly
detected the one ESBL-producing
Klebsiella pneumoniae isolate
and incorrectly detected one of the two
E. coli isolates as
an ESBL-producing organism (Table
4). Of the 10 previously characterized
ESBL-producing challenge isolates (5
Klebsiella pneumoniae,
3
Klebsiella oxytoca, and 2
E. coli isolates), the Phoenix instrument
correctly identified 6 isolates, with a "no alert" message recorded
for 4 isolates; the MicroScan system correctly classified all
10 isolates as possible ESBL producers with the message "?ESBL,"
indicating the need for confirmatory testing.
Overall complete agreement between the two systems following
AST of NFGNB was 93.2%; the mE, ME, and VME rates were 8.0,
2.9, and 0%, respectively. Gentamicin and trimethoprim/sulfamethoxazole
accounted for the two MEs.

DISCUSSION
Overall, the Phoenix instrument compared favorably to the MicroScan
system for the ID and susceptibility testing of gram-negative
bacilli. We purposely restricted the number of
E. coli and
Pseudomonas aeruginosa isolates to ensure a balanced representation of clinical
isolates, especially those with low frequencies of occurrence
in our laboratory. Although the range and number of isolates
tested were not as broad as those described in comparable published
reports, they do reflect a realistic representation of the organisms
encountered by a service-oriented clinical microbiology laboratory.
The overall agreement for the ID of common members of the family
Enterobacteriaceae was similar to that reported by others, although
in the previously reported studies, the performance of the Phoenix
instrument was compared directly to conventional ID systems
(
1,
2,
12). In contrast, O'Hara tested a set of challenge isolates
consisting of biochemically typical and atypical stock culture
representatives of the
Enterobacteriaceae and reported that
the Phoenix system correctly identified only 89.9% to the genus
and species levels (
13). The majority of errors reported in
her study were associated with the genus
Salmonella, but no
such errors were encountered in the present study. In the study
by Carroll et al. (
2), an
Enterobacter (
Pantoea) species, specifically
Pantoea agglomerans, was misidentified as
Shigella flexneri,
one isolate of
E. coli was misidentified as
Salmonella sp.,
and all
Shigella isolates were misidentified as
E. coli. In
our study, only two isolates required the resolution of results
with the API 20E, and in these cases, the MicroScan system misidentified
Hafnia alvei as
Yokenella (
Koserella)
regensburgei and
Enterobacter aerogenes as
Serratia fonticola or
Enterobacter aerogenes. Following
the resolution of the two discrepancies, the Phoenix instrument
was found to have achieved an overall accuracy of ID of
Enterobacteriaceae at the genus and species levels of 100%. However, it is important
for microbiologists to be aware of the shortcomings of both
automated and nonautomated ID systems, especially for isolates
identified as
Shigella or
Salmonella spp., which require confirmatory
ID by additional testing (agglutination or the detection of
characteristic phenotypic features following growth on MacConkey
agar), in addition to knowledge of the source of the specimen.
Following the resolution of two discrepancies, the Phoenix system was found to have accurately identified all of the NFGNB clinical isolates, including the Kingella kingae isolate, which was misidentified as Moraxella sp. by the MicroScan system, and the mucoid strain of Pseudomonas aeruginosa, even though Pseudomonas putida was part of the selection. The results for the ID of this group of organisms are similar to those reported by Endimiani et al. (7), Menozzi et al. (10), and Stefaniuk et al. (16). In contrast, O'Hara reported that of 57 isolates of NFGNB, 48 (84.2%) were correctly identified by the Phoenix system to the genus and species levels (12).
Both systems encountered considerable difficulty in identifying unusual isolates that are occasionally recovered, albeit in low frequencies, from cystic fibrosis patients, including Burkholderia species other than Burkholderia cepacia and Ralstonia and Pandoraea spp. Given that these organisms are rare and that Pseudomonas aeruginosa is the most common bacterium isolated from cystic fibrosis patients, these results are not surprising since these organisms are not included within the databases of the respective instruments. In fairness to both systems, Burkholderia cenocepacia is part of the Burkholderia cepacia complex and the organisms of this complex are difficult to distinguish biochemically. Other investigators have reported similar observations and concerns in utilizing automated systems for the ID of such organisms and have emphasized the importance of confirmatory testing of these traditionally difficult-to-identify organisms (1, 7, 12, 19). Our data also illustrate the potential problems that may be encountered when using these systems alone for the ID of unusual NFGNB and the need to perform confirmatory testing using more reliable phenotypic or genotypic methods.
In our study comparing AST results from the MicroScan and Phoenix systems, the CAs for both groups of organisms were excellent. A significant weakness of this study was the lack of sufficient numbers of resistant isolates to accurately determine the true rates of MEs and VMEs. In a service-oriented clinical microbiology laboratory in which clinical isolates are recovered from clinical specimens, like the one in which this study was conducted, some types of bacterial resistance occur at very low frequencies, which compromises the ability to generate data that are essential for calculating accurate error rates. This problem is reflected in the VME rates that exceed the accepted standard of
3% (6). Of particular note was the observation that the majority of VMEs involved β-lactam antibiotics. Menozzi et al. reported similar observations and noted, as was observed in our study, nearly equal numbers of errors for the various organisms tested (10). Similar findings were also reported by Carroll et al. for the Enterobacteriaceae (2).
Currently, five studies have extensively evaluated the ability of the Phoenix system to detect ESBLs (2, 10, 16, 17, 20). In the most recent study, Wiegand et al. (20) performed a side-by-side comparison of the MicroScan, Vitek 2, and Phoenix systems by testing 150 putative ESBL producers that were blindly distributed to three participating laboratories following the biochemical and molecular characterization of β-lactamases, which served as the reference method. A total of 85 isolates were identified as ESBL producers by the reference method. The Phoenix system demonstrated the highest sensitivity (99%) for the detection of ESBLs, followed by the Vitek 2 (86%) and the MicroScan (84%) systems. Carroll et al. demonstrated that the Phoenix expert rules system correctly detected the six ESBL producers, three E. coli and three Klebsiella spp. isolates, encountered in their study (2). Sturenburg et al. (17) also reported a high degree of sensitivity (100%) for the Phoenix system following the testing of 34 ESBL-producing challenge strains. Similar observations were reported by Sanguinetti et al. (14), who determined the Phoenix system to have 100% sensitivity and 98.9% specificity for the detection of ESBLs in 11 species of Enterobacteriaceae that had been characterized previously by phenotypic and genotypic methods. In a study of 74 ESBL-producing clinical isolates of E. coli and Klebsiella pneumoniae and 17 genotypically characterized control strains, Leverstein-van Hall et al. (9) compared the Phoenix system to the Vitek 1 and Vitek 2 systems, with the accuracies of the three systems for detecting ESBLs reported to be 89, 78, and 83%, respectively. In our study, both systems performed very well in screening for and/or detecting ESBL production in clinical isolates of Klebsiella pneumoniae and E. coli. However, the MicroScan system demonstrated higher levels of sensitivity and specificity in screening the ESBL-producing challenge isolates of E. coli, Klebsiella pneumoniae, and Klebsiella oxytoca that contained other β-lactamases. The higher-level performance of the Phoenix system reported by Wiegand et al. (20) and Sanguinetti et al. (14) was most likely due to the availability in Europe of expanded expert rules software (Phoenix software version 4.05W and MicroScan software version 1.12) that included organisms other than E. coli and Klebsiella pneumoniae, which is not available in the United States and is awaiting FDA clearance and CLSI recommendations for the testing of other members of the Enterobacteriaceae family, with the exception of the recent addition of guidelines for Proteus mirabilis.
Overall, the Phoenix system compared favorably with the MicroScan WalkAway system and is an acceptable alternative for the ID and susceptibility testing of commonly encountered species of Enterobacteriaceae and non-glucose-fermenting gram-negative bacilli. However, caution should be used in relying on either system for use in the ID and susceptibility testing of less commonly encountered clinical isolates, such as those that are occasionally recovered from patients with cystic fibrosis.

ACKNOWLEDGMENTS
This study was supported in part by a grant from BD Diagnostics,
Inc.
We thank Ellen Smith-Molan and Ken Thomson for providing the ESBL challenge isolates and Peter Gilligan for providing the NFGNB challenge isolates. We also thank the University of Louisville Hospital microbiology laboratory staff for their cooperation in this study.

FOOTNOTES
* Corresponding author. Mailing address: Department of Pathology and Laboratory Medicine, University of Louisville Hospital, 530 S. Jackson St., Louisville, KY 40202. Phone: (502) 852-1777. Fax: (502) 852-1771. E-mail:
jwsnyd01{at}gwise.louisville.edu 
Published ahead of print on 21 May 2008. 

REFERENCES
1 - Brisse, S., S. Stefani, J. Verhoef, A. Van Belkum, P. Vandamme, and W. Goessens. 2002. Comparative evaluation of the BD Phoenix and Vitek 2 automated instruments for identification of isolates of the Burkholderia cepacia complex. J. Clin. Microbiol. 40:1743-1748.[Abstract/Free Full Text]
2 - Carroll, K. C., B. D. Glanz, A. P. Borek, C. Burger, H. S. Bhally, S. Henciak, and D. Flayhart. 2006. Evaluation of the BD Phoenix automated microbiology system for the identification and antimicrobial susceptibility testing of Enterobacteriaceae. J. Clin. Microbiol. 44:3506-3509.[Abstract/Free Full Text]
3 - Clinical and Laboratory Standards Institute. 2005. Performance standards for antimicrobial susceptibility testing: 15th informational supplement. M100-S15. CLSI, Wayne, PA.
4 - Coenye, T., T. Spiker, R. Reik, P. Vandamme, and J. J. LiPuma. 2005. Use of PCR analysis to define the distribution of Ralstonia species recovered from patients with cystic fibrosis. J. Clin. Microbiol. 43:3463-3466.[Abstract/Free Full Text]
5 - Reference deleted.
6 - Elder, B. L., S. A. Hanson, J. A. Kellogg, F. J. Marskik, and R. J. Zabransky. 1997. Cumitech 31, Verification and validation of procedures in the clinical microbiology laboratory. ASM Press, Washington, DC.
7 - Endimiani, A., F. Luzzaro, A. Tamborini, G. Lombardi, V. Elia, R. Belloni, and A. Toniolo. 2002. Identification and antimicrobial susceptibility testing of clinical isolates of non-fermenting gram-negative bacteria by the Phoenix automated microbiology system. New Microbiol. 25:323-329.[Medline]
8 - Kiska, D. L., A. Kerr, M. C. Jones, J. A. Caracciolo, B. Eskridge, M. Jordan, S. Miller, D. Hughes, N. King, and P. H. Gilligan. 1996. Accuracy of four commercial systems for identification of Burkholderia cepacia and other gram-negative nonfermenting bacilli recovered from patients with cystic fibrosis. J. Clin. Microbiol. 34:886-891.[Abstract]
9 - Leverstein-van Hall, M. A., A. C. Fluit, A. Paauw, A. T. A. Box, S. Brisse, and J. Verhoef. 2002. Evaluation of the Etest ESBL and the BD Phoenix, VITEK 1, and VITEK 2 automated instruments for detection of extended-spectrum beta-lactamases in multiresistant Escherichia coli and Klebsiella spp. J. Clin. Microbiol. 40:3703-3711.[Abstract/Free Full Text]
10 - Menozzi, M. G., U. Eigner, S. Covan, S. Rossi, P. Somenzi, G. Dettori, C. Chezzi, and A.-M. Fahr. 2006. Two-center collaborative evaluation of performance of the BD Phoenix automated microbiology system for identification and antimicrobial susceptibility testing of gram-negative bacteria. J. Clin. Microbiol. 44:4058-4094.
11 - Moland, E. S., N. D. Hanson, J. A. Black, A. Hossain, W. Song, and K. S. Thomson. 2006. Prevalence of newer β-lactamases in gram-negative clinical isolates collected in the United States from 2001 to 2002. J. Clin. Microbiol. 44:3318-3324.[Abstract/Free Full Text]
12 - O'Hara, C. M. 2006. Evaluation of the Phoenix 100 ID/AST system and NID panel for identification of Enterobacteriaceae, Vibrionaceae, and commonly isolated nonenteric gram-negative bacilli. J. Clin. Microbiol. 44:928-993.[Abstract/Free Full Text]
13 - O'Hara, C. M. 2005. Manual and automated instrumentation for identification of Enterobacteriaceae and other aerobic gram-negative bacilli. Clin. Microbiol. Rev. 18:147-162.[Abstract/Free Full Text]
14 - Sanguinetti, E., B. Posteraaro, T. Spanu, D. Ciccaglione, L. Romano, B. Fiori, G. Nicoletti, S. Zanett, and G. Fadda. 2003. Characterization of clinical isolates of Enterobacteriaceae from Italy by the BD Phoenix extended-spectrum β-lactamase detection method. J. Clin. Microbiol. 41:1463-1468.[Abstract/Free Full Text]
15 - Snyder, J. W., G. K. Munier, and C. L. Johnson. 2006. Abstr. 106th Gen. Meet. Am. Soc. Microbiol., abstr. C-016. American Society for Microbiology, Washington, DC.
16 - Stefaniuk, E., A. Baraniak, M. Gniadkowski, and W. Hryniewicz. 2003. Evaluation of the BD Phoenix automated identification and susceptibility testing system in clinical microbiology laboratory practice. Eur. J. Clin. Microbiol. Infect. Dis. 22:479-485.[CrossRef][Medline]
17 - Sturenburg, E., I. Sobottka, H.-H. Feucht, D. Mack, and R. Laufs. 2003. Comparison of BD Phoenix and VITEK 2 automated antimicrobial susceptibility test systems for extended-spectrum beta-lactamase detection in Escherichia coli and Klebsiella spp. clinical isolates. Diagn. Microbiol. Infect. Dis. 45:29-34.[CrossRef][Medline]
18 - Thomson, K. S., N. E. Cornish, S. G. Hong, K. Hendrick, C. Herdt, and E. S. Molan. 2007. Comparison of Phoenix and VITEK 2 extended-spectrum-β-lactamase detection tests for analysis of Escherichia coli and Klebsiella pneumoniae isolates with well-characterized β-lactamases. J. Clin. Microbiol. 45:2380-2384.[Abstract/Free Full Text]
19 - van Pelt, C., C. M. Verduin, W. H. F. Goessens, M. C. Vos, B. Tummler, C. Segonds, F. Reubsaet, H. Berbrugh, and A. Van Belkum. 1999. Identification of Burkholderia spp. in the clinical microbiology laboratory: comparison of conventional and molecular tests. J. Clin. Microbiol. 37:2158-2164.[Abstract/Free Full Text]
20 - Wiegand, L., H. K. Geiss, D. Mack, F. Sturenburg, and H. Seifert. 2007. Detection of extended-spectrum beta-lactamases among Enterobacteriaceae by use of semiautomated microbiology systems and manual detection procedures. J. Clin. Microbiol. 45:1167-1174.[Abstract/Free Full Text]
Journal of Clinical Microbiology, July 2008, p. 2327-2333, Vol. 46, No. 7
0095-1137/08/$08.00+0 doi:10.1128/JCM.00075-08
Copyright © 2008, American Society for Microbiology. All Rights Reserved.
This article has been cited by other articles:
-
Segonds, C., Clavel-Batut, P., Thouverez, M., Grenet, D., Le Coustumier, A., Plesiat, P., Chabanon, G.
(2009). Microbiological and Epidemiological Features of Clinical Respiratory Isolates of Burkholderia gladioli. J. Clin. Microbiol.
47: 1510-1516
[Abstract]
[Full Text]