Previous Article | Next Article ![]()
Journal of Clinical Microbiology, August 2008, p. 2681-2685, Vol. 46, No. 8
0095-1137/08/$08.00+0 doi:10.1128/JCM.00449-08
Copyright © 2008, American Society for Microbiology. All Rights Reserved.

Medical Microbiology Laboratory, University of Alberta Hospital, Edmonton, AB, Canada,1 Clinical Microbiology Laboratory, Duke University Hospital, Durham, North Carolina,2 University of Illinois Medical Center, Chicago, Illinois3
Received 6 March 2008/ Returned for modification 16 June 2008/ Accepted 18 June 2008
|
|
|---|
|
|
|---|
The science has evolved over many years. Classical methods gave way to substrate-based identification of these microorganisms involving overnight and then shorter incubation. More recently molecularly based identification has started to change the landscape (2, 5, 7, 8, 12), but for most laboratories that lack such sophisticated technical capabilities, the ability to use either a manual or automated system that will give a high level of correct identifications is sufficient for most purposes.
For identification of Haemophilus and Neisseria species, a variety of commercial methods have been available on the market for many years. For Neisseria species these have included Neisseria-Kwik, Gonogen, Gonochek II, RIM-N, API QuadFERM-Plus, Minitek, Identicult-Neisseria, and several others (1, 6, 16, 18). All of these are based on colorimetric changes in miniaturized substrates, with either enzymatic or growth end points. The ability to identify and separate Haemophilus species from nonpathogenic Neisseria species, particularly those from the respiratory tract, then led to the addition of Haemophilus identification in these systems. RapID NH, API NH strip, and Haemophilus Identification Test Kit were developed to separate Haemophilus species (3, 7, 9-11, 13-15, 17). These systems all perform with reasonable accuracy in the clinical laboratory. Depending on the method used and on the number and variety of strains tested, correct identifications varied between 73 and 99% for species in either genus.
The aforementioned systems are designed primarily for smaller laboratories. With changes and consolidations of laboratories, the use of automated systems for the identification of clinical isolates has become commonplace, and commercial manufacturers have now attempted to expand the capabilities of their identification systems to optimize identification for better patient care.
bioMérieux, Inc., has developed a new Neisseria-Haemophilus (NH) identification card for the Vitek 2 system. The NH card is based on colorimetric technology utilizing dehydrated media containing chromogenic substrates. The card has a database that includes 27 taxa of gram-negative fastidious bacteria and that maintains the predominance of Haemophilus and Neisseria species but also includes Actinobacillus, Campylobacter, Capnocytophaga, Cardiobacterium, Eikenella, Gardnerella, Kingella, Moraxella, Oligella, and Suttonella species. In a recently published investigation from one laboratory, 91% of strains included in the database were identified correctly without the need for additional tests (17). In the present investigation, the quality, reproducibility, and accuracy of this NH card were assessed in three large tertiary care clinical laboratories. Confirmation of identifications was made by 16S rRNA gene sequencing as a quality assurance step to assess the performance of the NH card.
|
|
|---|
Each clinical isolate at all three study sites was stored as a heavy suspension at 4°C in sterile physiological saline. These strains were then shipped directly to MIDI Labs (Newark, DE) for 16S rRNA gene sequencing.
Quality control and reproducibility testing. Nine microorganisms comprised the quality control and reproducibility set of organisms. The strains tested were Eikenella corrodens ATCC BAA-1152, Campylobacter jejuni subsp. jejuni ATCC BAA-1153, Neisseria lactamica ATCC 23970, Gardnerella vaginalis ATCC BAA-1154, Haemophilus (Actinobacillus) actinomycetemcomitans ATCC BAA-1155, Neisseria gonorrhoeae ATCC 19424, Haemophilus influenzae ATCC 9007, Oligella urethralis ATCC 17960, and Haemophilus aphrophilus ATCC 33389. The set was tested 20 times over a minimum of 10 days at each of the three clinical trial sites. The results were used to evaluate system and substrate reproducibility. One manufacturing lot of NH cards was used for the entire study.
Evaluation of system reproducibility was based on the number of correct identifications. Correct identification was defined as accurate identification of a microorganism by the system as the only choice (with any level of confidence—excellent, very good, good, or acceptable) or as one of the choices within a multichoice (i.e., low-discrimination) result. The expected system reproducibility performance criterion was set at 95% or greater correct identification within a 95% confidence interval.
Evaluation of substrate reproducibility was based on microorganism/biochemical well combinations with a distinct expected reaction of positive or negative. For individual well reactions with distinct positive or negative reactions, the expected substrate reproducibility performance criterion was agreement of at least 95% within a 95% confidence interval.
Challenge testing. Each clinical study site tested a panel of 30 well-characterized isolates once on the NH card. The strains were supplied to each study site by bioMérieux. The isolates were not among those used to create the NH database and were selected to represent the identification claims of the NH product. Definitions for the levels of correct identification were as follows: overall correct identification, the isolate was accurately identified by the system as the only choice (with any level of confidence) or as one of the choices within a low-discrimination result; low discrimination, the identification contained two or three single or multichoice identification results in contrast to the correct identification with a single choice; incorrect identification, a final identification in which the genus or species was incorrect; unidentified microorganisms, a final identification of "unidentified organism", "inconclusive identification," or "nonreactive biopattern."
The minimum performance requirements for the clinical trial were at least 95% overall correct identification, less than 25% low discrimination, less than 2% incorrect identification, and less than 5% unidentified organisms for the NH card within a 95% confidence interval.
Clinical isolates. A total of 371 recently isolated clinical strains, either fresh or recently frozen (within the past 6 months) and previously identified as one of the "claimed" species, were tested at the three clinical trial sites. All isolates were tested once on the NH card and by 16S rRNA gene sequencing utilizing approximately 500 bp. 16S rRNA gene sequencing was considered the reference method. MIDI Labs (Newark, DE) performed the 16S sequencing and utilized the Applied Biosystems MicroSeq microbial analysis software and database to evaluate genetic similarity (19). When a genetic match was not found in the MicroSeq library, MIDI performed a BLAST search of the GenBank public database (4). Sequences from isolates with a genus level 16S sequencing identification from MIDI or GenBank were also compared to the bioMérieux, Inc., proprietary 16S database. For the purposes of this investigation the 16S sequencing results were interpreted in the context of the NH identification database. A 16S sequencing result consisting of a single-choice, species level genetic match to the NH card identification was a complete, correct result. A 16S sequencing result with no genetic difference between multiple species was confirmed through supplemental biochemical testing if one of the species listed matched the NH identification. A 16S sequencing result with a species level identification that was discrepant with the NH identification was evaluated to determine if the microorganism listed was part of the NH claim list. If it was, the discrepant result was confirmed or resolved by retesting the isolate. If it was not listed on the NH claim list, the isolate was excluded from the data set but the results were reviewed to evaluate issues related to all unclaimed species. A 16S sequencing result reported by MIDI Labs as a genus level identification and/or match to a sequence in the GenBank database was sent to the bioMérieux R&D Microbiology group in La Balme, France, for review to determine the relationship of the sequence result to microorganisms in the database. Definitions for the level of correct identification and the performance criteria were the same as those used for the challenge testing.
Excluded/unclaimed genera/species. Strains that were not included as claimed genera or species by NH card identification were recorded but necessarily excluded from the overall data set compilations. The 16S RNA sequence was used to arbitrate the NH card results for these strains. The criterion for an excluded or unclaimed strain was an identification from the NH card that was not in the database and that was supported by 16S RNA sequencing, no identification given by the NH card and a 16S sequence identification that was not one of the claimed species, or no 16S rRNA gene sequencing identification.
|
|
|---|
|
View this table: [in a new window] |
TABLE 1. NH card quality control and reproducibility results for all sites combined
|
Clinical isolate testing. The clinical isolates tested comprised all 27 of the taxa contained in the NH identification card database. Compared to 16S rRNA gene sequencing as the reference method, 96.5% (358/371) of the clinical isolates were identified correctly by the NH card and met all performance criteria set by the clinical trial protocol across all three sites (Table 2). The cumulative performance by species for all the clinical isolates tested at all three sites is shown in Table 3. The overall correct-identification performance for the new NH card was 96.5%, with only 10.2% low-discrimination results and 2.7% incorrect results. The results also fell within the established confidence intervals.
|
View this table: [in a new window] |
TABLE 2. NH card overall performance summary by site for clinical isolate testing
|
|
View this table: [in a new window] |
TABLE 3. NH card performance summary by species for the clinical isolate test set for all sites combined
|
Incorrect and discrepant identifications of claimed species as judged by comparison with 16S rRNA gene sequencing are listed in Table 4. The overall performance was excellent. Clinically only one result might be considered problematic. An isolate identified as Neisseria cinerea with an acceptable Vitek 2 confidence was identified by gene sequencing as C. coli.
|
View this table: [in a new window] |
TABLE 4. Incorrect or discrepant identifications of the clinical isolate test set for all sites
|
|
|
|---|
With respect to the Neisseria species, the NH card had greater limitations for nonpathogenic Neisseria species than 16S sequence identification. N. sicca is a claimed species but would often be identified as N. flavescens if the gold standard 16S rRNA gene sequencing identification was used. However, in most cases separation as a nonpathogenic Neisseria species would be sufficient to exclude pathogenic strains. This issue is not unique to any of the automated or conventional systems. At the technologist level, the new NH card would misidentify organisms approximately 3% of the time, and currently about 7% of isolates would be unclaimed in the test system.
The NH card was designed to improve significantly the capabilities of the Vitek 2 system to identify a variety of fastidious clinical microorganisms, but particularly Haemophilus and Neisseria species. Clinical trials such as presented here are important to adequately test such new identification systems. The data support the claims of the manufacturer for this new automated identification card.
All microorganism and biochemical combinations having distinct expected reactions of positive or negative in which quality control was evaluated met the performance criteria when measured cumulatively across sites with the exception of four biochemical reactions for three quality control species. As a result of these trials H. actinomycetemcomitans has been removed from the quality control list, leaving only two discrepancies that passed statistical analysis. For the clinical isolates, only a few isolates of some species were included in the evaluation. Haemophilus haemolyticus may be misidentified as H. influenzae or Haemophilus parainfluenzae, unless hemolysis is observed. Also, all four isolates of Kingella denitrificans were identified with low discrimination, although a correct identification was made. Additional strains of some of these species should be evaluated to confirm the conclusions of this investigation.
Overall, the data for which the card has a claim indicate that the NH card meets all performance criteria at greater than 95% confidence. The results from this clinical trial in three large tertiary clinical laboratories indicate that the Vitek 2 NH card is acceptable for routine use in a clinical microbiology laboratory.
This investigation was supported by a grant-in-aid from bioMérieux, Inc., Durham, NC.
Published ahead of print on 25 June 2008. ![]()
|
|
|---|
an, B., S. Asikainen, and H. Jousimies-Somer. 1999. Evaluation of two commercial kits and arbitrarily primed PCR for identification and differentiation of Actinobacillus actinomycetemcomitans, Haemophilus aphrophilus, and Haemophilus paraphrophilus. J. Clin. Microbiol. 37:742-747.
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Copyright © 2009 by the American Society for Microbiology. For an alternate route to Journals.ASM.org, visit: http://intl-journals.asm.org | More Info»