Journal of Clinical Microbiology, January 2009, p. 217-219, Vol. 47, No. 1
0095-1137/09/$08.00+0 doi:10.1128/JCM.01506-08
Copyright © 2009, American Society for Microbiology. All Rights Reserved.

Wisconsin State Laboratory of Hygiene, Communicable Disease Division, 465 Henry Mall, Madison, Wisconsin 53706
Received 5 August 2008/ Accepted 10 November 2008
|
|
|---|
|
|
|---|
The mecA gene confers resistance to methicillin in S. aureus. The gene is located on the staphylococcal chromosome cassette mec and encodes penicillin binding protein 2a (PBP2a). PBP2a is located in the bacterial cell wall and has a low binding affinity for β-lactams (1, 2). This study evaluated the new Clinical Laboratory Standards Institute (CLSI) breakpoints for the cefoxitin disk test for determining mecA-mediated resistance in S. aureus (4).
CLSI recommends usage of cefoxitin instead of oxacillin when using the disk diffusion method to determine resistance against methicillin for S. aureus (4). Cefoxitin results are easier to interpret and are thus more sensitive for the detection of mecA-mediated resistance than oxacillin results (5, 6, 7, 9-11). The recommended resistance and susceptibility breakpoints for the 30-µg cefoxitin disk test used to detect mecA-mediated resistance in S. aureus were changed in January 2007 by CLSI from
19 mm and
20 mm to
21 mm and
22 mm, respectively. This study sampled a large number of recent S. aureus isolates to compare the performance of the cefoxitin disk test at the new breakpoints to that of the 1-µg oxacillin disk test.
Between August and September 2007, a total of 1,611 nonduplicate S. aureus isolates were collected by 53 Wisconsin clinical laboratories (up to 20 consecutive MRSA and methicillin-susceptible S. aureus isolates per laboratory) and submitted to the Wisconsin State Laboratory of Hygiene. Species identification was confirmed by colony morphology, coagulase slide test, subsequent tube test, and biochemicals. The isolation sites of the S. aureus isolates collected were as follows: skin and soft tissue (1,159 isolates; 71.9%), urine (136 isolates; 8.4%), respiratory tract (130 isolates; 8.1%), bloodstream (65 isolates; 4.0%), and other (121 isolates; 7.5%).
Susceptibility to antimicrobial agents for confirmed S. aureus isolates was evaluated by the CLSI disk diffusion method on Mueller-Hinton agar (3). All agar plates and antibiotic disks were obtained from Remel (Lenexa, KS). A direct colony suspension of each S. aureus isolate was prepared to a 0.5 McFarland standard and plated on Mueller-Hinton agar. The zones of inhibition were measured using the Biomic V3 system (Giles Scientific, Inc., Santa Barbara, CA) after 24 h of incubation at 35°C (4). The oxacillin disk was read using transmitted light as the CLSI document recommends. The ATCC 25923 S. aureus strain was included in each test sample set as a quality control. All results were within the acceptable limits for quality control strains according to the 2007 CLSI guidelines.
Figure 1 shows the scatterplot correlation of oxacillin (1 µg) disk diameter with the cefoxitin (30 µg) disk diameter. The correlation coefficient (R2) for the scattergram was determined to be 0.8389. The cefoxitin disk demonstrated a sensitivity and specificity of 97.3% and 100%, respectively, compared to the oxacillin disk.
![]() View larger version (7K): [in a new window] |
FIG. 1. Correlation of oxacillin (1 µg) and cefoxitin (30 µg) disk test zone diameters obtained from 1,611 Staphylococcus aureus isolates. A circled number indicates a PBP2a-negative isolate at that position. The dashed line represents the 2006 cefoxitin CLSI interpretive criteria for S. aureus; the solid vertical line represents the new cefoxitin CLSI interpretive criteria. The solid horizontal lines represent the CLSI interpretive criteria for oxacillin.
|
Cefoxitin incorrectly identified only one isolate as susceptible (zone diameter = 22 mm) that was oxacillin resistant and PBP2a positive. An additional isolate that tested intermediate to oxacillin and resistant to cefoxitin was found to be negative for PBP2a. Four of the isolates testing resistant to oxacillin from this study would have been misinterpreted as susceptible using the previous cefoxitin susceptibility breakpoint of
20 mm.
Cefoxitin will detect only MRSA with a mecA-mediated resistance mechanism (8). There is a CLSI comment in the M100-S17 document warning of this limitation of cefoxitin as a substitute for oxacillin. However, non-mecA-mediated methicillin resistance in S. aureus is a rare occurrence, as evidenced by this study. Even with this limitation, cefoxitin disk diffusion zones are much easier to read than those of oxacillin due to the frequent hazy oxacillin zones, which are commonly misinterpreted as evidence of oxacillin susceptibility. This rate of false susceptibility associated with the oxacillin disk diffusion test has been noted to be as high as 4.4% in some studies (7; Wisconsin State Laboratory of Hygiene, unpublished data), well above the CLSI-recommended acceptability limit of
1.5%. Oxacillin must also be read using transmitted light, unlike most other antimicrobials, including cefoxitin, to ensure correct interpretation.
This large-sample study validated the new CLSI breakpoints for the 30-µg cefoxitin disk test to detect mecA-mediated resistance in S. aureus. We found the cefoxitin disk diffusion test to be a superior test and a surrogate for the oxacillin disk diffusion test due to its ease of reading and higher sensitivity.
The study contents are solely the responsibility of the authors and do not necessarily represent the official views of the CDC.
Published ahead of print on 19 November 2008. ![]()
|
|
|---|
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||