Journal of Clinical Microbiology, February 2009, p. 441-444, Vol. 47, No. 2
0095-1137/09/$08.00+0 doi:10.1128/JCM.01989-08
Copyright © 2009, American Society for Microbiology. All Rights Reserved.

Technical University Dresden, Medical Faculty Carl Gustav Carus, Institute of Medical Microbiology and Hygiene, Fetscherstrasse 74, D-01307 Dresden, Germany
Received 15 October 2008/ Returned for modification 20 November 2008/ Accepted 5 December 2008
|
|
|---|
|
|
|---|
In recent years, different PCR systems targeting a wide range of genes of M. pneumoniae (e.g., P1 adhesin, 16S rRNA, and ATPase operon gene) have been developed (16). Real-time PCR especially is characterized by rapidity, practicability, reduced risk of contamination, and high specificity and sensitivity of detection (3, 7, 9, 13, 15, 19, 23, 24, 25). However, the increasing number of real-time PCR approaches stresses the need for comparative validation of the performance of the different test procedures. Up to now, studies dealing with the evaluation of more than a single quantitative PCR system to detect M. pneumoniae have been very rare (7, 25) and have not included commercial test kits. To our knowledge, commercial quantitative PCR systems for the detection of M. pneumoniae are not available in the United States (2), whereas kits from different manufacturers are widely used in Europe. The aim of the present study was to investigate the sensitivities of different commercial and in-house real-time PCR approaches for the detection of M. pneumoniae under standardized conditions.
M. pneumoniae reference strains M129 (subtype 1; ATCC 29342) and FH (subtype 2; ATCC 15531) and patient isolates M3896 (subtype 3; kindly provided by S. Dégrange, Université Victor Segalen, Bordeaux, France), 4817 (variant 1), and ST (variant 2a) were propagated as described previously (5). For calculation of the standard curves, a freshly grown culture of M. pneumoniae M129 was harvested and sheared through a 27-gauge needle to reduce bacterial aggregates. Aliquots of the suspension were used to prepare the DNA and to calculate the CFU by spreading 10-fold phosphate-buffered saline dilutions in triplicate on PPLO agar (Becton Dickinson, Sparks, MD). M. pneumoniae-positive clinical samples were obtained from a collection of respiratory tract specimens (bronchoalveolar lavage fluids, nasopharyngeal or pharyngeal swabs, or sputa) of pneumonia patients in Germany and Switzerland. DNA from all M. pneumoniae culture and patient material samples (sample volume, 200 µl each) was extracted with a QIAamp DNA mini kit (Qiagen, Hilden, Germany) according to the instructions of the manufacturer (protocol for blood and body fluids; elution volume, 150 to 200 µl). The DNA concentration was measured photometrically. DNA of patient samples was pretested by using the real-time PCR described recently in reference 7. M. pneumoniae bacteria in specimens confirmed as positive were subtyped by P1 sequencing according to the method of Dumke et al. (6). The DNAs of quantified dilutions of the M. pneumoniae M129 stock, of the different M. pneumoniae subtypes and variants, and of the M. pneumoniae-positive patient samples used for amplification with the different real-time PCR approaches were aliquoted and stored at –20°C until use. For each quantitative PCR test, freshly thawed DNA aliquots of the quantification standards, of the M. pneumoniae subtypes and variants, and of the patient materials were used in order to avoid an influence of DNA degradation on the crossing thresholds.
Five real-time PCR assays were selected for parallel testing of the M. pneumoniae-positive DNA samples. The three in-house real-time PCR systems were chosen according to the target used for amplification and the appropriate validation of the procedure in the literature. The RepMp1-based system (7) amplifies an 89-bp product within the 14 repetitive elements of RepMp1 scattered all over the genome of the sequenced M. pneumoniae strain M129 (20) and represents a multicopy target assay. The second approach targets a 76-bp part of the ATPase operon gene (MPN592) of M. pneumoniae (3), whereas the CARDS Tx assay (25) detected M. pneumoniae DNA by amplification of a 73-bp region located in the recently described community-acquired respiratory distress syndrome toxin gene (10). Primer and probes (Biomers, Ulm, Germany) were exactly as published. Probes were labeled with 6-carboxyfluorescein (5') and 6-carboxytetramethylrhodamine (3'). The different real-time PCR assays were performed by using a LightCycler 1.5 instrument (Roche, Mannheim, Germany) with a final volume of 20.0 µl containing 4.6 µl water (PCR grade, Roche), 2.4 µl MgCl2 (25 mM, Roche), 2.0 µl LightCycler FastStart DNA master HybProbe mix (Roche), 2.0 µl of each primer (5 pmol), 2.0 µl of probe (2 pmol), and 5.0 µl of DNA. The glass capillaries were incubated with the following cycling conditions: preincubation at 95°C (10 min), 45 cycles of denaturation at 95°C (10 s), hybridization at 53°C (RepMp1; 10 s) or 60°C (ATPase and CARDS Tx; 10 s), and elongation at 72°C (30 s). The reaction mixtures were cooled down to 40°C for 1 min. The commercial quantitative PCR assays selected for the study are the artus M. pneumoniae LC PCR kit (Qiagen), also targeting the different RepMp1 copies of M. pneumoniae, and the Venor Mp-QP M. pneumoniae diagnostic kit (Minerva Biolabs, Berlin, Germany), amplifying a part of the main P1 adhesin of M. pneumoniae. Both of the commercial quantitative PCR systems are expressly recommended for use with the LightCycler instrument. Performance of the commercial real-time PCR assays was carried out strictly according to the instructions of the manufacturer. This included the use of the recommended MB Taq DNA polymerase, which is not provided with the kit, in the Venor quantitative PCR. The data from all amplifications were analyzed with the LightCycler software, version 3.5 (Roche).
The in-house and commercial real-time PCR approaches were extensively tested with the DNA of related mycoplasma species and different agents commonly detected in patients with pneumonia and were found to be specific for the amplification of M. pneumoniae. According to the results in the published studies and the data in the handbooks of the manufacturers, the proven detection limits of the procedures investigated ranged from at least 0.2 copies or CFU per assay (RepMp1 and artus) to 1 to 5 CFU per assay (CARDS Tx) and 10 copies or CFU per assay (ATPase and Venor). The DNA samples (with one exception that was due to the limited volume of a single respiratory tract specimen) were tested in triplicate with each of the five real-time PCR approaches investigated. Parallel testing of all materials with the different procedures was carried out within three weeks to reduce the influence of long-term sample storage on the results.
Investigations of collections of patient isolates confirmed that M. pneumoniae is a very homogeneous species from a genetic point of view (5). However, since 1990 (21), two subtypes of M. pneumoniae showing sequence differences in the two copies of the repetitive elements RepMp4 and RepMp2/3 located in the main P1 adhesin have frequently been found in the human population. Furthermore, in recent years, three variants of M. pneumoniae showing sequence variations in one of these two repetitive elements were described (4, 6, 11). Variant 2a strains were reported with increasing numbers in Japan (12) and Europe (6), but only one isolate has been characterized as variant 1. Variant 2b isolates were detected in limited number with a molecular subtyping method in Europe only (6); therefore, DNA samples from cultures of this genotype could not be included in the present study. Recently, a new genetic composition of both repetitive elements in the P1 gene in a strain investigated in France led to the definition of a third subtype of M. pneumoniae (18).
Despite the distinct differences in the quantitative occurrence of the described genotypes, diagnostic real-time PCR procedures have to detect all known subtypes and variants of M. pneumoniae. Possible sequence variations in the target genes in the subtypes and variants are not reported but cannot be excluded. As indicated by the results summarized in Table 1, all real-time PCR approaches investigated demonstrated the occurrence of M. pneumoniae-specific DNA in bouillon cultures of the five available genotypes of M. pneumoniae. The standard curves (crossing thresholds versus log CFU) generated by six 10-fold dilutions of a quantified M. pneumoniae M129 stock (1 to 105 CFU/µl) yielded r2 values of 0.999 (ATPase), 0.998 (RepMp1), 0.997 (CARDS Tx), 0.993 (Venor), and 0.977 (artus), respectively (data not shown). The PCR efficiencies (101/slope) varied between 1.9 (ATPase) and 2.3 (artus). All quantitative PCR procedures included in the study were able to detect at least one CFU per 1-µl sample. The results of testing the samples from the respiratory tract of patients with confirmed M. pneumoniae infection demonstrated that the five real-time PCR procedures investigated also recognized the occurrence of DNA of M. pneumoniae variant 2b strains, which are not available as culture material. The internal control amplification in both of the commercial assays confirmed that PCR inhibition in the tested samples can be excluded.
|
View this table: [in a new window] |
TABLE 1. Results of testing different specimens with commercial and in-house real-time PCR assays for the detection of M. pneumoniae
|
This study minimized sources of variation evident in multicenter investigations in several ways. Unfavorable storage and transportation conditions of the samples, slight variations in the LightCycler instruments used, and different interpersonal sample handling can affect the comparability of the results of real-time procedures (1). These influences are reduced by the standardized performance of the assays tested in the present study. Furthermore, within the test systems investigated, the coefficients of variation of the mean crossing points of the three quantified dilutions of M. pneumoniae DNA used to calculate the concentration of the genome copies in the five independent runs (n = 15) ranged between 0.7 and 3.3% (data not shown), confirming a high level of intralaboratory reproducibility in all assays tested. These facts allow a direct comparison of the crossing thresholds. The results of real-time PCR procedures used to detect M. pneumoniae revealed significant differences in the sensitivities of the test systems. Nevertheless, all assays investigated were able to detect the known genotypes of M. pneumoniae bacteria with a sensitivity of at least one CFU per 1-µl sample and can be recommended for the diagnostic laboratory. Further studies are needed to prove if the higher sensitivity of the procedures amplifying a multicopy target (RepMp1; artus) is of any diagnostic relevance.
Published ahead of print on 10 December 2008. ![]()
|
|
|---|