Journal of Clinical Microbiology, March 2009, p. 645-651, Vol. 47, No. 3
0095-1137/09/$08.00+0 doi:10.1128/JCM.01412-08
Copyright © 2009, American Society for Microbiology. All Rights Reserved.

Special Pathogens Unit, National Institute for Communicable Diseases of the National Health Laboratory Services, Private Bag X4, Sandringham 2131, Republic of South Africa,1 Center for International Collaborative Research, Institute for Tropical Medicine, Nagasaki University, 1-12-4, Sakamoto, Nagasaki 852-8523, Japan,2 Department of Microbiology and Plant Pathology, Faculty of Natural and Agriculture Sciences, University of Pretoria, Pretoria 0002, Republic of South Africa,3 Department of Virology, Institute of Tropical Medicine, Nagasaki University, 1-12-4, Sakamoto, Nagasaki 852-8523, Japan,4 Division of Virology and Communicable Disease Surveillence, School of Pathology, University of the Witwatersrand, Johannesburg, South Africa5
Received 23 July 2008/ Returned for modification 28 August 2008/ Accepted 12 December 2008
|
|
|---|
|
|
|---|
Techniques for the diagnosis of RVF include virus isolation, detection of specific antibody responses, and molecular assays. RVFV can be isolated from serum or whole blood during the febrile stage of the disease as well as from the livers, spleens, and brains of animals who have succumbed to the disease or aborted fetuses. Isolation of the virus is achieved in hamsters, infant or adult mice, and various cell cultures (35). Various serological assays are used to detect antibodies against RVFV, but the virus neutralization test is regarded as the gold standard (23). It is highly accurate, with little or no cross-neutralization with other phleboviruses (33), but since it requires live virus, it can be done only in biocontainment facilities. Enzyme-linked immunosorbent assays, based on inactivated sucrose-acetone-extracted antigens derived from tissue culture or mouse brain, have been extensively validated for the serodiagnosis of RVF (22, 23). However, their production requires biocontainment facilities to limit the risk of exposure of laboratory personnel to infection (9, 29). An indirect enzyme-linked immunosorbent assay based on the recombinant nucleocapsid protein of RVFV has been recently developed for the detection of specific antibodies in human and animal sera (6, 24, 25). Highly sensitive PCR assays for the detection and quantification of RVFV have been reported, including reverse transcriptase PCR (3, 28) and real-time detection PCR (RTD-PCR) based on TaqMan probe technology (2), but they have not been validated for routine diagnostic use.
The loop-mediated isothermal amplification (LAMP) method has been shown to be highly accurate for the detection of DNA (4, 20) and RNA (14, 32, 36) viruses, differentiation of viral serotypes and subtypes (17, 21), and rapid diagnosis of bacterial infections (5). LAMP amplifies target nucleic acid under isothermal conditions, usually between 60°C and 65°C (19). Hence, only simple equipment, such as a heating block or water bath, is required, obviating the need for expensive automated thermal cyclers. It relies on autocycling strand displacement DNA synthesis by Bst DNA polymerase and a set of four or six primers. Two inner primers and two outer primers define the target region, and an additional set of primers, termed loop primers, can be added to increase the sensitivity of the assay. The final products of the LAMP reaction are DNA molecules with a cauliflower-like structure of multiple loops consisting of repeats of the target sequence. The products can be analyzed by real-time monitoring of the turbidity resulting from the production of magnesium pyrophosphate precipitate during the DNA amplification reaction, by using an intercalating fluorescent dye and visualization by natural light or with the aid of UV irradiation, or by agarose gel electrophoresis (15, 16).
This paper reports the development and diagnostic evaluation of a real-time reverse transcription-LAMP assay (RT-LAMP) targeting the large RNA segment in a wide spectrum of RVFV isolates recovered in the field over a period of 50 years (1944 to 2007) and various types of clinical specimens from sheep experimentally infected with wild-type virus and from humans and animals naturally infected during recent outbreaks of the disease in Africa.
|
|
|---|
|
View this table: [in a new window] |
TABLE 1. Primers used for RT-LAMP of the polymerase gene (L segment) of RVFV
|
Cell culture and viruses. The analytical sensitivity of RT-LAMP was evaluated by testing 17 RVFV strains representing a wide geographic and historic spectrum of isolates recovered in the field over the last 50 years (1944 to 2007) (Table 2), with titers ranging from 105.2 50% tissue culture infective doses (TCID50)/ml to 107.3 TCID50/ml. Five selected isolates representing the phylogenetic spectrum of RVFV, namely, those from West Africa, Egypt, and central, eastern, and southern Africa (Table 2), were tested in the range of 0.0065 to 65,000 TCID50 per reaction volume. The analytical specificity of RT-LAMP was evaluated by testing six African phleboviruses related to RVFV, Akabane (106.8 TCID50/ml), Bunyamwera (107.8 TCID50/ml), Gabek Forest (107.0 TCID50/ml), Saint Floris (105.8 TCID50/ml), Arumowot (104.8 TCID50/ml), and Gordil (105.8 TCID50/ml) viruses and six other unrelated arboviruses, yellow fever (106.0 TCID50/ml), dengue I (105.5 TCID50/ml), West Nile lineage 1 (107.8 TCID50/ml), West Nile lineage II (106.0 TCID50/ml), chikungunya (107.5 TCID50/ml), and Crimean-Congo hemorrhagic fever (104.3 TCID50/ml) viruses. All of these virus isolates were obtained from the Special Pathogens Unit of the National Institute for Communicable Diseases, Sandringham, South Africa. All the viruses were amplified either in Vero cells (CCL-81) or in suckling mice by using standard procedures (35).
|
View this table: [in a new window] |
TABLE 2. Identification, year of isolation, source, and origin of RVFV isolates
|
|
View this table: [in a new window] |
TABLE 4. Comparison of RT-LAMP, TaqMan RT-PCR, and virus isolation results for the detection of RVFV in various clinical specimens
|
RNA extraction. Animal tissues were homogenized as 10% (wt/vol) suspensions in EMEM supplemented with antibiotics as described above. After centrifugation at 3,000 x g, supernatants free of cells were harvested and stored at –70°C. Genomic viral RNA was extracted from 140 µl of infected Vero cells tissue culture supernatants, liver and kidney tissue supernatants, or sera by using a QIAamp viral RNA mini kit (Qiagen, Germany) in accordance with the manufacturer's protocol. The extracted RNA was eluted in a total volume of 60 µl of buffer AVE and stored at –70°C until use.
TaqMan RTD-PCR. The TaqMan-RTD PCR assay was performed using a LightCycler RNA amplification kit HybProbe (Roche Diagnostics, Germany) and a Roche LightCycler instrument. Amplifications were carried out in 20-µl reaction mixtures containing 5 µl of the target virus RNA or the in vitro-transcribed RNA standard, 1 µM concentrations of each of the sense and antisense primers, and 5 mM MgCl2. Cycling profiles, primers, and a TaqMan probe targeting a region of the G2 glycoprotein as described by Drosten et al. (2) were used.
RT-LAMP. RT-LAMP was carried out in a final reaction volume of 25 µl using a Loopamp RNA amplification kit (Eiken Chemical Co., Ltd. Tokyo, Japan) with 5 pmol (each) of the primers F3 and B3, 20 pmol (each) of the primers LF and LB, and 40 pmol (each) of the primers FIP and BIP. Five microliters of the extracted RNA was used as template per reaction. For real-time monitoring, the RT-LAMP reactions were incubated at 61°C for 60 min with a LA-200 Loopamp Realtime Turbidimeter (Teramecs, Japan) and inactivated at 80°C for 5 min. Positive and negative controls were included in each run of the assay.
Analysis of RT-LAMP product. (i) Real-time monitoring. A real-time turbidimeter was used to spectrophotometrically monitor every 6 s the accumulation of magnesium pyrophosphate at 400 nm. The cutoff values for positive samples were determined over time, when turbidity increased above the threshold value, which was fixed at 0.1. Results were analyzed using the LA-200E software package (Teramecs, Japan).
(ii) Agarose gel analysis. Five microliters of the RT-LAMP products were electrophoresed on a 2% molecular-grade agarose gel prepared in 0.5 x Tris-borate-EDTA buffer stained with 0.5 µg/ml ethidium bromide. The amplification products were visualized using a transilluminator with UV light at 302 nm.
(iii) Visualization with the naked eye. Reaction tubes were pulse centrifuged to deposit the magnesium pyrophosphate in the bottom of the tubes to detect amplification with the naked eye. Alternatively, 1 µl of fluorescent detection reagent (FDR; Eiken Chemical Co., Ltd., Tokyo, Japan) was added to the LAMP reaction. The FDR contains calcein complexed with manganese ions. Fluorescence occurs when pyrophosphate ions (formed as a by-product during LAMP amplification) removes the manganese ions from the calcein and binds to the calcein in the FDR. Fluorescence is further intensified when the calcein binds to magnesium ions in the reaction (16). For a positive reaction, orange changes to yellow fluorescence that can be detected with the naked eye or to green fluorescence under UV irradiation, while a negative reaction remains the orange color of the unbound dye. The color change can be observed by the naked eye or with the aid of UV light at 302 nm, and the results are captured photographically.
(iv) Restriction endonuclease digestion. Amplification products were digested with BstXI (Roche, Germany) as suggested by the manufacturer and analyzed with a 2% agarose gel as described above.
In vitro transcription and quantification. To obtain a quantitative RNA standard, the diagnostic target region was amplified by standard reverse transcriptase PCR using RVFV RNA prepared from infected tissue culture supernatant and transcribed in vitro. The primers for the RVF RTD-PCR or RVF-LAMP outer primers were used, respectively, to generate amplicons representative of the RTD-PCR or LAMP targets. The target region PCR products were cloned into T7/Sp6 polymerase expression vector pCRII-TOPO (Invitrogen). The inserts were then amplified with vector-specific, universal M13 primers using standard PCR. The PCR products were purified using a Wizard SV Gel and PCR Clean Up system (Promega) and then in vitro transcribed and DNase digested using a MegaScript Sp6 kit (Ambion) according to the manufacturer's protocol. The RNA was purified according to the instructions of the RNeasy Protect kit (Qiagen, Germany) and quantified spectrophotometrically. The target RNA copy number was calculated, and serial dilutions ranging from 100 to 106 RNA copies were used to determine the range of quantification.
Analytical sensitivities of TaqMan RTD-PCR and RT-LAMP. To compare the analytical sensitivities of RT-LAMP and TaqMan RTD-PCR in the detection of decreasing number of RNA copies, tenfold dilution series of the RNA standard, ranging from 100 to 106 per reaction, were tested in six separate runs of each assay.
|
|
|---|
45 min and turbidity above the threshold value of
0.1 were considered a positive result (19). Real-time monitoring of turbidity allowed for the detection of amplification products as early as 16 min after initiation of the reaction, with the majority of positive specimens detected in less than 30 min.
Analytical sensitivity and specificity of RT-LAMP.
When the sensitivity of the RT-LAMP was compared with that of a TaqMan RTD-PCR by testing tenfold serial dilutions of RNA prepared from infective tissue culture supernatant containing 106.8 TCID50/ml of the AR20368 RSA 81 isolate of RVFV, the two assays were equally sensitive, with a detection limit of 0.065 TCID50 units per reaction volume (Fig. 1 A and B). The RT-LAMP detection limit of 0.065 TCID50 units per reaction volume was further confirmed by testing tenfold serial dilutions of four additional strains of RVFV (Table 2), representing a phylogenetic spectrum of the virus (results not shown). High levels of analytical sensitivity for both assays were also demonstrated by measuring decreasing numbers of RNA copies. The RNA standard dose-response curves for TaqMan RTD-PCR and RT-LAMP were very similar to one another (R2 of 0.8386 and 0.8042, respectively) and had the expected characteristic slope (Fig. 2). The borderline analytical detection limit as measured by number of RNA copies per reaction was 10 copies for both assays. During six runs, TaqMan RTD-PCR detected 10 RNA copies/reaction on four occasions (66.7%) and the RT-LAMP on three occasions (50%); however, this difference was not statistically significant (P = 1.0000). Both assays had 100% sensitivity in detecting
100 RNA copies/reaction (Fig. 2). Despite distinct geographic and historic origins, all the RVFV isolates tested in this study were easily detectable by the RT-LAMP (Table 2), further confirming its high level of analytical sensitivity. The high level of analytical specificity of the assay was confirmed by the absence of amplification products when using the RVFV-specific L primer set with RNA extracted from highly concentrated stocks of six African phleboviruses related to RVFV and six other, unrelated arboviruses (results not shown). In addition, the specificity of the RT-LAMP amplification product was confirmed by restriction endonuclease digestion with BstXI, resulting in a product with the expected size of 209 bp (result not shown).
![]() View larger version (25K): [in a new window] |
FIG. 1. Real-time kinetics of RT-LAMP (A) and TaqMan RTD-PCR (B) in tenfold serial dilutions of RNA prepared from infective tissue culture supernatant containing 106.8 TCID50/ml of the AR20368 RSA 81 isolate of RVFV.
|
![]() View larger version (16K): [in a new window] |
FIG. 2. Analytical sensitivity of RT-LAMP and TaqMan RTD-PCR as measured by the detection of numbers of RNA copies. Data were derived from testing tenfold dilution series of RNA standard during six runs of both assays. Logarithmic trend lines are added and correlation coefficients are calculated to determine the strength of linear data sets.
|
![]() View larger version (52K): [in a new window] |
FIG. 3. Agarose gel electrophoresis profile of RT-LAMP products in serially diluted RNA of RVFV. Lane M, 100-bp molecular weight marker. Lanes 1 to 8, RT-LAMP products yielded from tenfold serial dilutions of RNA prepared from infective tissue culture supernatant containing 106.8 TCID50/ml of the AR20368 RSA 81 isolate of RVFV. Lanes 1 to 7, from left to right, 65,000 TCID50 to 0.065 TCID50/reaction volumes. Lane 8, 0.0065 TCID50/reaction volume; not detected. Lane 9, negative control.
|
![]() View larger version (78K): [in a new window] |
FIG. 4. Visual detection of RT-LAMP amplification products in tenfold dilutions of RVFV RNA. Tube 1 to 8 from left to right, RT-LAMP products yielded from tenfold serial dilutions of RNA prepared from infective tissue culture supernatant containing 106.8 TCID50/ml of the AR20368 RSA 81 isolate of RVFV; tube 9, negative control. Detection of the white precipitate (magnesium pyrophosphate) at the bottom of the tubes indicates a positive reaction, while the absence of precipitate indicates a negative reaction. After the addition of FDR to the reaction mix, fluorescence is detectable by the naked eye as a yellow color, while an orange color indicates a negative reaction (A); under UV irradiation, a positive reaction is indicated by a fluorescent bright green color (B).
|
|
View this table: [in a new window] |
TABLE 3. Monitoring of viremia in sheep experimentally infected with wild-type RVFV by RT-LAMP, TaqMan RTD-PCR, and virus titration
|
|
|
|---|
An array of molecular techniques for rapid detection and identification of RVFV have been published, but data on their routine diagnostic performance are not available (2, 3). In this study, the utility of the LAMP technique for rapid and accurate detection of RVFV RNA in clinical specimens was investigated. Including the time required for the extraction of nucleic acid, detection of RNA in different clinical specimens of human and animal origin could be achieved within 2 h after arrival of the samples in the laboratory. Apart from the high levels of analytical and diagnostic accuracy and speed of detection, another important practical advantage of the LAMP technique is that it utilizes simple and relatively inexpensive equipment, which renders it promising for use in resource-poor settings. In addition, only basic molecular and technical skills are required for execution of the assay procedure, and interpretation of the results may be as simple as a visual evaluation of color change in the reaction mix. However, primer design for LAMP is more complex than for the conventional PCR-based assays, and specialized training and software are required for their design. Moreover, the development of LAMP requires the use of a set of multiple primers spanning a highly conserved 300-bp genomic region. For RVFV, this is easily achievable because there is no evidence of serological subgroups or major antigenic variation between isolates of disparate chronologic or geographic origins (30), which is due to their low level of genetic diversity, irrespective of the genome segments analyzed (1). In the present work, we demonstrate that the set of LAMP primers targeting the RNA-dependent RNA-polymerase gene of the L segment of RVFV is well designed to detect all strains of the virus.
Results of the study show that the sensitivities of the RT-LAMP and TaqMan RTD-PCR assays are similar and that there is no cross-reactivity of the primers with the genes of related and unrelated arboviruses. Both assays had high levels of analytical sensitivity as measured by the detection of a known number of virus infectious doses and RNA copies of the in vitro-transcribed RNA standard. Similar analytical sensitivity for the TaqMan RTD-PCR based on the Superscript reverse transcriptase-Platinum Taq polymerase enzyme mixture was reported by Drosten et al. (2). By direct comparison of the two nucleic acid procedures with virus isolation results for clinical specimens from animals and humans, the diagnostic performance of the TaqMan RTD-PCR assay for the detection of RVFV is also reported here for the first time. Specific nucleic acid targets in all positive specimens could be detected in this study in less than 45 min, with the majority detected in less than 30 min. This result confirms that the assay allows for rapid confirmation of clinical cases and early recognition of outbreaks. Visualization of amplification products with the naked eye, fluorescence, or agarose gel electrophoresis may be appropriate for most laboratory settings, while real-time monitoring of the accumulation of magnesium pyrophosphate in the reaction mix potentiates quantification of the assay. One has to emphasize, however, that definitive diagnosis or exclusion of RVF, as for any other suspected case of VHF, should not rely on a single PCR result. The LAMP should be run in parallel with additional tests, including the detection of type-specific antibodies to RVFV. In this context, it is important to note that viremia in RVFV-infected individuals is of very short duration (30) and most infected humans and adult ruminants undergo subclinical or mild infections, but immunoglobulin M and immunoglobulin G antibodies are easily demonstrable shortly after exposure to the virus (22, 23).
|
|
|---|
We thank the following employees of the National Health Laboratory Service, South Africa: Pat Leman for technical laboratory assistance and Guy Hall for photographical assistance.
Published ahead of print on 24 December 2008. ![]()
|
|
|---|
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||