Journal of Clinical Microbiology, March 2009, p. 652-659, Vol. 47, No. 3
0095-1137/09/$08.00+0 doi:10.1128/JCM.01900-08
Copyright © 2009, American Society for Microbiology. All Rights Reserved.

Department of Medical Microbiology and Infectious Diseases, Erasmus Medical Center, Rotterdam, The Netherlands,1 Center for Optical Diagnostics and Therapy, Department of Dermatology, Erasmus Medical Center, Rotterdam, The Netherlands,2 Department of Medical Microbiology, Maastricht Infection Center (MINC), University Hospital Maastricht, Maastricht, The Netherlands3
Received 1 October 2008/ Returned for modification 28 November 2008/ Accepted 15 December 2008
|
|
|---|
|
|
|---|
Up to 30% of HAI are caused by transmission and can therefore be prevented by means of infection control strategies (6). Several studies have emphasized the importance of active screening of patients in order to identify transmission of microorganisms. This detects possible cases of cross-transmission and thereby indicates the infection control measures to be taken, including isolation of patients (1, 2). Rapid assessment of microbial clonal relationships (typing) enables tracking of the spread of pathogens, which may be used to significantly decrease the number of HAI and additional costs, even in high-risk situations (14, 17).
Methicillin-resistant Staphylococcus aureus (MRSA) is a leading cause of HAI, and numerous DNA-based typing methods have been developed for the organism. Therefore, we used MRSA as an example to demonstrate the capabilities of a new typing method.
Pulsed-field gel electrophoresis (PFGE) is generally the most preferred typing method (9), followed by multilocus sequence typing (MLST) and spa typing. All these techniques require dedicated facilities, are labor-intensive, and have a long turnaround time (48 to 72 h when starting from a pure culture) and a low sample throughput. Therefore, they are unsuitable for routine use in hospitals and, consequently, are mostly applied in retrospective analysis of outbreak situations.
An ideal typing method should be rapid and simple and have high throughput and good discriminatory power and should provide results that are in concordance with the epidemiological data (19). We present a novel approach to meet these requirements, based on Raman spectroscopy. Raman spectra of bacteria are representations of their overall molecular composition and can be used as highly specific spectroscopic fingerprints. We used four different collections of methicillin-sensitive S. aureus (MSSA) and MRSA isolates to demonstrate the effectiveness of Raman spectroscopy as a typing tool that could be used in epidemiological surveillance studies and concluded that this technique is an easy-to-use and rapid alternative in the battle against MRSA.
|
|
|---|
Collection I was used as a reference collection and contained 20 well-characterized MRSA isolates. These isolates have previously been analyzed using multiple typing techniques (22) and have been used to determine the interlaboratory reproducibility of PFGE in a multicenter study (20). This collection contains 5 isolates with identical PFGE patterns, 5 isolates with related PFGE patterns (patterns that differ by
3 bands), and 10 unrelated isolates (with patterns that differ by >3 bands).
Collection II was obtained from the Department of Medical Microbiology and Infectious Diseases of the Erasmus University Medical Center, Rotterdam, The Netherlands (Erasmus MC), and included all 78 MRSA isolates collected in the year 2002 in this tertiary-care Dutch hospital. The isolates were retrieved from patients, as well as from health care workers. All of the isolates have been analyzed independently using PFGE at the Department of Medical Microbiology at the Erasmus MC, as well as at the Dutch National Institute for Public Health and the Environment (RIVM).
Collection III was obtained from the Department of Microbiology at the University Hospital Maastricht, Maastricht, The Netherlands, and contained five isolates retrieved from a MRSA-colonized member of the hospital staff collected over a period of 18 months and four isolates obtained from different patients during an MRSA contact screening.
Collection IV contained four MRSA isolates (one from a patient and three from a colonized staff member) and seven MSSA isolates obtained during an MRSA contact screening in January 2007 at the Erasmus MC.
Storage and culture. The isolates were stored at –80°C in brain heart infusion broth (Becton Dickinson, Franklin Lakes, NJ) containing 10% glycerol until they were used.
For Raman measurements, all isolates were grown overnight on Trypticase soy agar (TSA) (Becton Dickinson, Franklin Lakes, NJ). After this first culture was checked for purity, three to five colonies were picked to fill a calibrated 1-µl loop and suspended in 20 µl of sterilized demineralized water (aquadest, prepared in house). This suspension was diluted 100 times in aquadest, and 20 µl was plated on a TSA plate. These plates were incubated for 20 h at 35°C to obtain a confluent bacterial layer.
PFGE. PFGE was performed according to the Harmony protocol (3). Briefly, a suspension of bacteria was mixed with 1% InCert agarose (FMC Bioproducts). The agarose plugs were incubated with lysostaphine (Sigma-Aldrich, Zwijndrecht, The Netherlands), and spheroplasts were lysed using proteinase K (Sigma-Aldrich). DNA was digested by SmaI (Fermentas, St. Leon-Rot, Germany). Macro-restriction fragments were separated using a Bio-Rad Chef Mapper (Bio-Rad, Veenendaal, The Netherlands) with a total run time of 20 h. PFGE patterns were assigned based on unique patterns (19).
spa typing. Real-time amplification of the spa locus, followed by sequencing, was performed as described previously (15). The spa types were clustered into spa clonal complexes (CCs) using the algorithm Based upon Repeat Pattern (BURP) with the Ridom StaphType version 1.5 software package (http://www.ridom.de). The default settings recommended by the manufacturer were used. The associated CCs, as determined by MLST, were allocated through the Ridom SpaServer (http://spaserver.ridom.de).
Sample preparation for Raman spectroscopy. For sample preparation, a calibrated 1-µl loop was filled with biomass and suspended in 5 µl of demineralized water. After a centrifugation step for 1 min at 12,000 x g to remove air bubbles, the pellet was resuspended and transferred to a fused silica slide (Hellma Benelux BV, Aartselaar, Belgium). On this slide, a removable silicone isolator (Sigma-Aldrich, Zwijndrecht, The Netherlands) containing 24 wells was placed. Samples were allowed to dry for 20 min at 35°C.
Raman spectroscopy. A scheme of the custom-built Raman setup that was used for measuring Raman spectra of the bacterial samples is shown in Fig. 1. The instrument consisted of a Model 2500 High Performance Raman Module (River Diagnostics BV, Rotterdam, The Netherlands) coupled to a custom-built measurement compartment. The measurement compartment was equipped with an automated x- and y-axis-positioning stage, which held the fused silica slide with bacterial samples and a custom-designed near-infrared-optimized microscope objective (numerical aperture, 0.7). The objective was used to focus laser light emitted by the Model 2500 High Performance Raman Module in the samples on the fused silica slide and to collect Raman scattered light from the samples. Approximately 220 mW of laser light at 785 nm was used to illuminate the samples (Sacher Tiger; Sacher Lasertechnik, Marburg, Germany). The spectrometer was calibrated according to the manufacturer's guidelines. In each sample, 10 to 60 spectra at a 1-s collection time were measured at different positions.
![]() View larger version (16K): [in a new window] |
FIG. 1. Overview of Raman procedure and spectrometer. Biomass from a bacterial culture (a) on TSA medium is collected using a 1-µl inoculation loop and suspended in 5 µl of demineralized water (b). After a brief centrifugation step to remove air bubbles, the wet pellet is transferred onto a fused silica slide (c), where it is allowed to dry (a typical slide holds 24 samples). The slide with the dried biomass is placed in the measurement stage (d), where the samples are illuminated with laser light (e). The Raman signal generated is collected along the same optical path and separated from the laser light using an optical filter (f) that only reflects light of a higher wavelength than the laser. The laser light is passed through. The wavelength of the Raman signal is dispersed on an optical grating (g) and collected using a near-infrared-optimized charge-coupled device detector (h). The Raman spectra are gathered, stored, and analyzed on a personal computer (i).
|
The software scripts for spectrum pretreatment and data analysis were written in MATLAB version 7.1 (Mathworks, Natick, MA).
(ii) Calculation of similarities between spectra. The similarity between pairs of spectra was calculated using the squared Pearson correlation coefficient (R2) and multiplied by 100 to be expressed as percentages.
(iii) Hierarchical cluster analysis. Cluster analysis of sets of spectra was performed using the pairwise similarities as a distance matrix, in combination with Ward's cluster algorithm. This resulted in a dendrogram in which each node represented the lowest correlation coefficient (or similarity) between all isolates combined in the cluster defined by this node.
For each isolate collection, the five genetically identical isolates of collection I were measured and analyzed simultaneously and used to indicate clusters, or Raman types.
The individual steps of this process were as follows. In step 1, the node where the five identical reference isolates of collection I combined in the dendrogram was defined. In step 2, the correlation coefficient represented by this node was used as a cutoff value in the dendrogram. In step 3, all isolates were combined in a cluster in which, if the value of the node was higher than the correlation coefficient found for the reference, the isolates were considered to be indistinguishable.
All clusters defined by this procedure could then be validated using previously known typing results of the isolates, epidemiological data, or a collection of reference isolates.
(iv) Discriminatory power. The discriminatory power of a typing method is defined as its ability to assign a different type to two unrelated strains (19). To quantify the discriminatory power of Raman spectroscopy, the numerical index of diversity (D) was calculated as described by Hunter and Gaston (7).
Experiments applied to evaluate Raman spectroscopy. (i) Reproducibility and influence of culturing time. Raman spectra represent the overall molecular composition of bacterial cells. Since differences in culture conditions may influence the metabolic pathways and therefore the molecular makeup of bacterial cells, changes in growth conditions may lead to changes in spectra and negatively affect the reproducibility of the Raman procedure.
Two MRSA isolates with identical PFGE patterns and three MRSA isolates with unique PFGE patterns were selected from collection I to evaluate the reproducibility of the Raman technique. All isolates were cultured in five independent sessions, and in each session, samples were prepared and measured after 18 h, 20 h, 22 h, and 24 h of incubation time, yielding a total of 20 samples for each isolate.
(ii) Reproducibility and concordance with other typing techniques. PFGE is considered the gold standard for typing of S. aureus isolates. Therefore, the 20 MRSA isolates of reference collection I were used to compare the results of the Raman typing with previously obtained typing results (22).
Experiments to evaluate the application of Raman spectroscopy in epidemiological surveillance. (i) Retrospective study: concordance with clinical epidemiology. The 78 clinical MRSA isolates of collection II were used to validate the concordance of Raman typing with clinical epidemiological data.
(ii) Historical isolates. Collection III contains MRSA isolates obtained from a staff member that were collected over a period of 18 months. These isolates were used to evaluate the ability of Raman spectroscopy to recognize historical and recent isolates from a colonized person.
(iii) Prospective study: comparative typing. Collection IV represents isolates obtained during an MRSA contact screening and was used to validate Raman spectroscopy to assess relatedness within a small set of isolates (comparative typing) in a possible outbreak situation.
|
|
|---|
Evaluation of Raman spectroscopy. (i) Reproducibility and influence of culturing time. To evaluate the influence of culturing time on the reproducibility of Raman spectra, five isolates of reference collection I were cultured five times and spectra were measured after 18 h, 20 h, 22 h, and 24 h of incubation time.
As expected, based on their PFGE patterns, isolates W01 and W02 were found in one distinct cluster, indicating that they were indistinguishable using Raman spectroscopy (Fig. 2). For the unrelated isolates W11, W12, and W13, Raman clustering was isolate specific, since three clusters containing the spectra of only one isolate were formed. For all isolates, the Raman clustering was equal for each of the five repeats and independent of culturing times that varied from 18 h to 24 h.
![]() View larger version (10K): [in a new window] |
FIG. 2. Reproducibility of Raman spectroscopy. Shown are the results of hierarchical cluster analysis of the repeated measurements of five MRSA isolates measured after different culturing times. Each isolate is indicated with a colored label. a.u., arbitrary units.
|
3 band differences between patterns (20).
![]() View larger version (40K): [in a new window] |
FIG. 3. Hierarchical cluster analysis of the MRSA isolates of collection I. Clusters are indicated in red and are based on the correlation coefficient of the five isolates with identical PFGE patterns (pattern A). For each node with a correlation coefficient lower than the cutoff, the corresponding value is indicated. a.u., arbitrary units.
|
Application of Raman spectroscopy. (i) Retrospective study: concordance with clinical epidemiology. A collection (collection II) of 78 MRSA isolates for which the clinical and epidemiological data were known was evaluated. All isolates were analyzed in house using a standardized PFGE protocol (13). PFGE results for the same isolates were obtained from the RIVM. Both the Raman and PFGE results are summarized in Table 1.
|
View this table: [in a new window] |
TABLE 1. Results of the Raman-based typing and PFGE analysis of the 78 isolates of collection II
|
In total, three discrepancies were documented between the different typing methods. The first discrepancy involved isolates from a patient and a staff member that shared the same PFGE pattern according to the in-house PFGE (both pattern J) but had unique Raman types (type 9 and type 26), as well as unique RIVM-PFGE patterns (cl23 and cl68). No epidemiological connection between these two people was documented. Although this does not exclude the possibility that these patients were colonized or infected with the same MRSA strain, as suggested by the in-house PFGE data, the epidemiological data are in agreement with both the Raman typing and the RIVM-PFGE results.
Thirteen isolates obtained from 11 patients and two staff members were found to share a single Raman type (type 5), but two different PFGE patterns were defined for these isolates (pattern AN/cl37 and pattern C/cl38). No epidemiological information is available to establish a relationship between these two groups.
The last discrepancy involved two staff members with Raman types 24 and 28. According to PFGE, their isolates belonged to a large group of isolates with type AP1 or cl18. The epidemiological data also suggested that these isolates were obtained during the same outbreak, but each had a unique Raman type.
(ii) Historical isolates. During a contact screening at the University hospital Maastricht, one staff member and four patients were found to be positive for MRSA. The staff member was known to carry MRSA for a longer period of time, and isolates from an 18-month period were included in the study. The isolates were previously analyzed using spa typing and belonged to the same spa CC. MLST data were associated through the SpaServer and revealed that all the isolates also belonged to the same MLST CC (Table 2).
|
View this table: [in a new window] |
TABLE 2. Result of the Raman based typing of the MRSA and MSSA isolates of collections III and IVa
|
(iii) Prospective study: comparative typing. During a contact screening at the Erasmus MC, four MRSA isolates (one from the index patient and three from colonized staff member 1) and seven MSSA isolates were obtained. As shown in Table 2, the MRSA isolate from the index patient was defined as Raman type 8, while the three MRSA isolates from staff member 1 were defined as Raman type 9. These findings suggested that there was no transmission between the index patient and the staff member. PFGE confirmation of these results was obtained 4 days later (Fig. 4).
|
View larger version (8K): [in a new window] |
FIG. 4. Contact screening of January 2007. Directly after the first MRSA isolate was isolated, the contact screening was started. Two days later, a second MRSA isolate was found and sent to the RIVM for analysis. At the same time, all isolates were analyzed using Raman spectroscopy. Within 24 h, the Raman results showed that the two MRSA isolates were not identical. This was confirmed by PFGE 4 days later.
|
|
|
|---|
Previous research showed that data obtained with Raman spectroscopy can be seen as spectroscopic fingerprints and used for identification to the species and subspecies levels (10, 11). Recently, we transformed Raman spectroscopy into a rapid and easy-to-use typing technique. The current progress entails simplification of the instrument and improved sensitivity and robustness of the spectrometer. Combined with the use of more powerful lasers, this has resulted in a significant reduction in signal collection time. Previously, a signal collection time of 30 s per spectrum was needed; now, a spectrum of the same quality can be obtained with a signal collection time of 1 s. Furthermore, a general procedure has been developed that is similar to the procedures used today in diagnostic microbiology laboratories. The protocols can be applied to a wide range of organisms, since there is no need for specific labels or dyes or special laboratory equipment, other than the Raman spectrometer. Starting from positive microbial cultures, sample preparation of 24 isolates takes approximately 45 min, including 30 min hands-on time. Raman spectra are collected automatically, and this usually takes 10 s to 1 minute per sample.
To validate typing techniques, a combination of typing data with epidemiological data is needed to confirm an outbreak situation. The Raman clustering obtained for the isolates of collection II revealed good agreement with the epidemiological data. In 2002, five outbreaks were reported in the Erasmus MC that were confirmed by PFGE. Raman spectroscopy recognized these five clusters of isolates and identified the majority of unique PFGE isolates as unique Raman types. Both techniques classified a number of isolates as identical, although no epidemiological link could be found.
One of the main reasons for typing is outbreak investigation. Dutch hospitals use an active "search-and-destroy" policy for MRSA control (24). During such a contact screening, a considerable number of samples have to be screened for the presence of MRSA. Typing methods play an important role in this strategy in order to confirm transmission of isolates (comparative typing). The results obtained with collection III show that Raman spectroscopy enabled historical isolates from an MRSA-colonized individual to be classified as identical, strongly demonstrating a long-time colonization of the staff member and a possible epidemiological link between the staff member and four different patients. Furthermore, both collections III and IV show that, for the contact screenings described here, Raman spectroscopy yielded the same results as established genotyping techniques, but several days earlier.
We conclude that Raman spectroscopy is a highly reproducible, easy-to-use, and rapid alternative to the currently used typing techniques that is suitable for real-time typing in clinical diagnostic laboratories. Such a technique will give infection prevention teams a tool for the continuous monitoring of isolates in their hospitals and will alert them immediately when corrective actions should be taken. This will lead to accurate real-time, rather than retrospective, surveillance and a new approach to combating HAI.
This study was funded by the Dutch Technology Foundation (STW) under project number 06311.
River Diagnostics developed and provided the custom-made Raman equipment for this project. On the basis of the results of this project, as reported here, River Diagnostics is currently pursuing the commercialization of Raman technology for microbial typing. Diana Willemse-Erix, Jan-Willem Jachtenberg, and Kees Maquelin were full-time employees of the Erasmus MC when the research described was performed and are currently part-time (D.W.-E. and K.M.) and full-time (J.-W.J.) employees of River Diagnostics. Tom Bakker Schut holds part-time positions at the Erasmus MC and at River Diagnostics. Gerwin Puppels is the Chief Executive Officer of River Diagnostics and is an associate professor at the Erasmus MC.