Journal of Clinical Microbiology, March 2009, p. 765-770, Vol. 47, No. 3
0095-1137/09/$08.00+0 doi:10.1128/JCM.01721-08
Copyright © 2009, American Society for Microbiology. All Rights Reserved.

West Of Scotland Specialist Virology Centre, a Member of the UK Clinical Virology Network, Gartnavel General Hospital, Great Western Road, Glasgow, G12 OYN, Scotland
Received 5 September 2008/ Returned for modification 16 October 2008/ Accepted 30 December 2008
|
|
|---|
|
|
|---|
Real-time PCR has become the standard test for these purposes, because it is fast, sensitive, and quantitative (10, 14, 17). In most cases, singleplex real-time PCR assays are used, although multiplex PCR assays that allow the detection of more than one target have also been described (6, 24). Multiplex PCR assays have many advantages over singleplex assays. For example, with regard to laboratory service, multiplexing reduces test costs, improves turnaround times, and increases test throughput (12, 18, 23, 25). These benefits have a positive effect on clinical service, allowing clinicians to tailor patient management or to initiate antiviral therapy more promptly. However, such assays need careful optimization in order to avoid competition and maintain sensitivity when more than one pathogen is present in the patient sample.
This paper describes a real-time multiplex PCR assay that can simultaneously detect and quantitate CMV, EBV, and adenovirus from plasma specimens. Most published assays quantitate by use of a standard curve formed by at least five standards of known concentrations. Further, each set of viral standards is tested in a different reaction. The standards and samples are often tested in triplicate in order to provide an accurate quantitative result. By multiplexing, the number of standards has been reduced from 15 to 5, each containing known concentrations of all three viruses: adenovirus, CMV, and EBV. We also show that testing both the standards and samples singly rather than in triplicate is acceptable.
|
|
|---|
Real-time PCR conditions. The primers and TaqMan probes for all three viruses have been published elsewhere (14, 20, 21). Each probe was labeled at the 5' end with a different fluorophore: 6-carboxyfluorescein, VIC, or Cy5 (Table 1). The primers and probes were obtained from Operon and ABI. The primers and probes were optimized prior to use by a previously published optimization technique (11), and each primer and probe was used at a final concentration of 100 µM or 20 µM, respectively.
|
View this table: [in a new window] |
TABLE 1. Sequences of the primers and probes used in this study
|
Multiplex real-time PCR method. The multiplex assay utilized the Qiagen QuantiTect multiplex kit with the following amplification profile: 2 min at 50°C; 95°C for 15 min; and 40 cycles of 95°C for 60 s and 60°C for 60 s. The multiplex assay was also assessed using the Invitrogen Platinum Quantitative PCR SuperMix-UDG kit with the profile outlined above. The real-time assay used a total reaction volume of 15 µl, where 6 µl was the DNA extract. This is an amendment to the recommended protocol.
Development of pooled standards for multiplex quantification. A single set of five standards was developed using plasmid and viral DNAs. The concentrations of each are shown (Table 2). The adenovirus standard was a commercial DNA obtained from Invitrogen. It is an adenovirus serotype 5. The copy number (per milliliter) was provided by the manufacturer. The CMV standard was derived from a CMV plasmid and had previously been standardized against reference material provided by QCMD. The EBV standard was generated from the Namwali EBV-infected cell line, and the copy number was determined by the provider (R. Jarrett, University of Glasgow). The same viral DNA controls at the same five concentrations had been used in the singleplex adenovirus, CMV, and EBV assays.
|
View this table: [in a new window] |
TABLE 2. Quantitation standards pooled to generate standard curves for each virus
|
(ii) Comparison of the singleplex and multiplex assays using a dilution series of a positive sample. The endpoints of the multiplex assay were compared to those of singleplex assays by using a series of 10-fold dilutions of plasma samples known to contain adenovirus, CMV, and EBV. The nucleic acid was extracted as outlined above, and each dilution series was tested in triplicate and the endpoints compared.
(iii) Precision of the multiplex assay. The precision of the assay was assessed by repeat testing of the standards with low (P1), moderate (P3), and high (P5) concentrations of viruses on 20 different occasions. The standard deviation (SD) was examined for each standard to determine test precision.
(iv) Linearity and efficiency of quantification using standards tested in singleplex assays. The five pooled standards were tested once on 20 separate occasions, and the standard curve was examined. The linearity of the curve and its efficiency were examined for each run. For accurate and reproducible quantification, it is ideal to have a standard curve with a linearity between –3.1 and –3.6 (the linearity of an ideal curve is –3.33) and a reaction efficiency near 100%. Such efficiency values represent a twofold increase in the level of the amplicon after each cycle.
(v) Assessment of the quantitation limits of the multiplex assay. In order to determine the quantitation limit of the assay (i.e., the levels above and below which quantitation becomes inaccurate), the five standards were tested on 20 occasions as unknown samples (i.e., given no quantitation value). The mean quantitation value attributed to the standards by the assay was then compared to the known value. The quantitation limit was the lowest standard for which the value attributed was found to be less than 50% or more than 200% of the known value. This experiment would also determine the viral load above which the assay would become inaccurate.
(vi) Whole-system reproducibility of the multiplex assay. The reproducibility of the assay was assessed by repeat testing (from the point of extraction) and by quantifying single controls known to contain 5,000 copies/ml of adenovirus, CMV, or EBV. These controls are derived from clinical samples containing known levels of adenovirus, CMV, or EBV. These samples were diluted to the appropriate concentration using viral transport medium. Although we report positive results below 5,000 copies/ml, we do not provide quantitative results for samples testing positive but below this cutoff. This cutoff has been routinely used and is based on the theory that the system is likely to become less accurate below this level.
DNAs from the 5,000-copies/ml controls were extracted and tested on 20 different occasions. The mean and standard variation were examined for each control.
(vii) Accuracy of multiplex detection of more than one pathogen. The accuracy of multiplex detection of more than one pathogen was assessed using a chessboard technique, designed to provide a large number of wells, each containing a known concentration of two PCR targets at different ratios. The method used for this assessment is described in detail elsewhere (13). These experiments were designed to determine whether competition is likely to be an issue and can also measure the ratio of each target above which competition can be expected.
In total, three panels of simulated samples were generated using five dilutions manufactured from the strongest quantitation standards: one containing various concentrations of EBV and CMV, one containing adenovirus and CMV, and one containing EBV and adenovirus. The dilution series from which these panels were manufactured were then tested using the singleplex assays. The results are presented as means (unless they were positive/negative).
Three clinical samples known to contain three pathogens (as determined by singleplex testing) were also tested. To assess the benefit of the Qiagen QuantiTect multiplex kit, both the panels of simulated mixed samples and the clinical samples were tested using both the Qiagen QuantiTect multiplex kit and the Invitrogen Platinum Quantitative PCR SuperMix-UDG kit.
(viii) Accuracy of multiplex quantification of clinical samples in comparison to singleplex assays. A total of 30 citrated blood samples, previously shown to be positive and quantified for either adenovirus, EBV, or CMV by singleplex real-time PCR, were used to assess whether the quantitative values provided by the multiplex assay were accurate. These samples were tested by the multiplex assay, and the results were compared to those obtained by the singleplex assays.
|
|
|---|
|
View this table: [in a new window] |
TABLE 3. Comparison of the endpoint detection limits of the multiplex and singleplex PCR assays
|
|
View this table: [in a new window] |
TABLE 4. Precision and reproducibility of the multiplex assay
|
|
View this table: [in a new window] |
TABLE 5. Linearity and efficiencya of the standard curves produced by the pooled standards
|
|
View this table: [in a new window] |
TABLE 6. Quantitation values attributed to the five pooled standards when tested as unknown samplesa
|
|
View this table: [in a new window] |
TABLE 7. Whole-system reproducibility of the multiplex assay
|
|
View this table: [in a new window] |
TABLE 8. CTs (in duplicate) obtained by testing of simulated samples containing both CMV and EBV with the Invitrogen Platinum Quantitative PCR SuperMix-UDG kita
|
|
View this table: [in a new window] |
TABLE 9. CTs (in duplicate) obtained by testing of simulated samples containing both CMV and EBV by the Qiagen QuantiTect multiplex kita
|
|
View this table: [in a new window] |
TABLE 10. Comparison of the Qiagen QuantiTect multiplex kit and the Invitrogen Platinum Quantitative PCR SuperMix-UDG kit on clinical samples containing >1 target
|
|
View this table: [in a new window] |
TABLE 11. Comparison of the quantitation values provided by the singleplex and the multiplex assay on 30 clinical samples
|
|
|
|---|
The quantitation is achieved using a single set of five quantitation standards, each containing a different concentration of adenovirus, CMV, and EBV. The presence of more than one target within each individual standard had no effect on the performance of the individual test components. Consequently, the standard curves produced were consistently linear and efficient.
The ability to detect and accurately quantitate >1 target is directly related to the use of the Qiagen QuantiTect kit. As mentioned above, this kit is especially designed for multiplexing and thus reduces test interaction and competition. Consequently, real-time PCR assays can be simply multiplexed with little optimization required (5, 7, 15, 16, 22).
The use of this kit also ensured that the presence of as many as three targets in the simulated and clinical samples had no effect on the performance of any aspect of the assay. Although the clinical significance of mixed infections in immunocompromised patients is unclear, dual and triple infections do occur. A recent audit of patients from the Bone Marrow Transplant Unit at the Glasgow Royal Infirmary showed that 20% had CMV, EBV, or adenovirus present in the samples tested. Of these patients, 13.2% had dual infections and 1.3% had triple infections. Although such infections would have been detected through the use of single assays, multiplexing using alternative PCR kits may have resulted in false-negative results or inaccurate quantitation, and this may, in turn, have had implications for clinical management.
The multiplex assay will have several positive outcomes for routine service (2, 12). For example, each sample is now tested by only one multiplex PCR assay rather than by three separate PCR tests. This will reduce the cost of the service significantly (because smaller amounts of PCR master mix, controls, and standards are needed) and will also improve turnaround times. Since implementing this assay in our laboratory, we have seen a reduction of at least a day in turnaround times, despite receiving an increased number of samples and despite the fact that all samples are now quantified (previously our laboratory quantified only CMV- and EBV-positive samples). The use of only 5 standards per PCR run (instead of 15) and the fact that all samples are now tested singly (instead of triplicate) will allow more clinical samples to be tested per PCR run. This will reduce test costs and turnaround times further. Both these examples highlight how the developments described here have simplified service, which in turn should reduce the number of technical errors occurring. Clinical service will also be improved, since the assay ensures that all relevant results are available immediately rather than sequentially. Consequently, patient management can be tailored more rapidly. The use of the multiplex assay will also free up laboratory equipment for other assays, which may, in turn, reduce their turnaround time.
The benefits of the Qiagen QuantiTect kit could also be applied to the use of other real-time PCR tests, particularly in situations where mixed infections are common. For example, the Qiagen QuantiTect kit would prove useful for respiratory and gastroenteritis multiplex assays. The use of the kit could also aid the development of internal-control systems, since internal-control PCR assays can now be more easily added to existing assays. We have now added an internal control to the triplex assay described here (9). The internal-control assay detects murine CMV, which is added to the samples prior to extraction. The addition of this assay has had no effect on the performance of the multiplex kit and now allows detection in samples containing inhibitors that may have led to false-negative results or inaccurate quantification (data not shown).
Published ahead of print on 14 January 2009. ![]()
|
|
|---|
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Copyright © 2009 by the American Society for Microbiology. For an alternate route to Journals.ASM.org, visit: http://intl-journals.asm.org | More Info»