Our laboratory uses a CE-marked real-time PCR assay (RealTime HCV; Abbott Molecular, Inc., Des Plaines, IL) that was originally evaluated with a magnetically based manual extraction procedure and, more recently, with different extractors, including Abbott m1000 and m2000sp, Tecan Freedom EVO Clinical, and Qiagen BioRobot EZ1 (7, 8, 10), but not easyMAG (bioMérieux, Boxtel, The Netherlands) (4). We have been using this instrument with success for other microorganisms and different clinical specimens. Since we wanted to change from manual to automated extraction without introducing other extractors, the aim of our study was to evaluate the easyMAG for the quantification of HCV load using the RealTime HCV assay in comparison with manual extraction.
The correlation between manual and automated extraction was compared in four runs, each including nine different sera from HCV-infected patients and three controls (negative, low positive, and high positive) for a total of 36 determinations. Sera were retrospectively selected spanning a wide HCV RNA virus load range, from <30 IU/ml—the lower quantification limit—up to 50,000,000 IU/ml, and negative specimens. The Clinical Research Ethics Committee at our institution approved this study. Several analytical parameters were assessed for the real-time PCR assay in combination with the extractor. (i) The first was lower detection limit. Twenty replicates of 25-, 12.5-, and 6.25-IU/ml dilutions obtained from the HCV RNA high-positive control (2 x 106 IU/ml; AcroMetrix Corporation, Benicia, CA) in negative plasma were tested per run. (ii) The second was linearity. Five replicates of five 10-fold dilutions (102 to 106 IU/ml) from the same standard were tested in one run. (iii) The third was intra-assay reproducibility. Five replicates of calibrators A and B (1,000 and 1.07 x 107 IU/ml, respectively; Abbott Molecular) and a dilution of the high-positive control (50,000 IU/ml) were tested. (iv) The fourth was interassay reproducibility. Five additional replicates of the same three dilutions were tested in another two runs performed on different days, adding up to a total of three runs. Negative, low-positive, and high-positive controls were included in each run.
Manual RNA extraction was performed as recommended by the manufacturer (Abbott RealTime HCV, Ref. 4J86; Abbott Molecular, Inc., Des Plaines, IL) from 200 µl of serum and elution in 88 µl. The internal control provided (unrelated armored RNA) was added to the lysis buffer at a proportion of 17 µl per sample. EasyMAG reagents are based on Boom's method for total nucleic acid extraction (1). Extraction was performed as recommended by the manufacturer (NucliSENS easyMAG user manual, v 1.1; BioMérieux, Boxtel, The Netherlands). A 250-µl sample volume and 110-µl elution volume were used to maintain the same sample/elution volume ratio as well as the proportion of internal control per sample. (A total of 235.7 µl of internal control was added to 314.3 µl of elution buffer and 550 µl of silica, and 100 µl of this premix was used per sample.)
Virus load quantification was performed using the RealTime HCV assay according to the manufacturer's instructions with the Abbott m2000rt instrument (Abbott RealTime HCV, Ref. 4J86). Separate calibration curves were obtained for manual and automated extraction methods with triplicate calibrators A and B.
Correlation between extraction methods was assessed through Deming regression and Bland-Altman plots with MedCalc software v9.6.2.0 (Mariakerke, Belgium). All other analyses were performed with SPSS v14.0 (SPSS, Inc., Chicago, IL), with a significance level of 0.05. Probit analysis was used to determine the lower limit of detection. Linearity was assessed through linear regression between expected and observed values and the Pearson correlation coefficient (r).
For the correlation study, a total of 48 sera from HCV-infected patients and controls were extracted in parallel by both manual and automated methods. Three clinical specimens were negative by both methods, as were the four negative controls, which confirmed the absence of carryover contamination. Another three specimens were positive at <30 IU/ml by manual extraction; one of them was also positive at <30 IU/ml, and the other two had virus loads of 55 and 58 IU/ml by automatic extraction. The real-time PCR was inhibited in two specimens: one of them after both manual and automated extractions and the other only after automatic extraction. A total of 36 specimens and controls with viral loads quantified after both manual and automatic extractions were included in the regression analysis. Deming regression showed that the correlation between the two methods was remarkable (y = 0.01379 + 1.0163 x, r = 0.993, P < 0.01): they did not differ by a constant amount (y intercept not significantly different from 0; 95% confidence interval, –0.2135 to 0.2411), and there was not a proportional difference between them either (slope not significantly different from 1; 95% confidence interval, 0.9749 to 1.0577). Bland-Altman analysis (Fig. 1) only identified two outliers, possibly due to a sporadic technical error; the mean difference between the two methods differed from the line of equality by 0.12 log10 IU/ml, which does not have a significant effect on therapeutic decisions.
![]() View larger version (19K): [in a new window] |
FIG. 1. Bland-Altman plot of differences versus average of values obtained after automatic and manual extraction with 95% limits of agreement and the line of equality (dotted line).
|
Results for the inter- and intra-assay reproducibility experiments are shown in Table 1. The variation observed after automated extraction was similar to that reported by the manufacturer for the real-time PCR technique with manual extraction over the range of viral loads tested (0.05 to 0.09 and 0.01 to 0.08 standard deviations in log10 IU/ml for the intra-assay and interassay reproducibility studies, respectively) (Abbott RealTime HCV, Ref. 4J86). Our results were comparable to those obtained with m1000 and EZ1 extractors in recent studies (7, 8), with intra- and interassay coefficients of variation of
2.4% for both instruments.
|
View this table: [in a new window] |
TABLE 1. Intra- and interassay reproducibility of the real-time PCR method using specimens extracted with easyMAG
|