Previous Article | Next Article 
Journal of Clinical Microbiology, April 2009, p. 1031-1036, Vol. 47, No. 4
0095-1137/09/$08.00+0 doi:10.1128/JCM.02099-08
Copyright © 2009, American Society for Microbiology. All Rights Reserved.
Correlation between Human Immunodeficiency Virus Type 1 (HIV-1) RNA Measurements Obtained with Dried Blood Spots and Those Obtained with Plasma by Use of Nuclisens EasyQ HIV-1 and Abbott RealTime HIV Load Tests 
Carolina Garrido,
Natalia Zahonero,
Angélica Corral,
Miguel Arredondo,
Vincent Soriano, and
Carmen de Mendoza*
Infectious Diseases Department, Hospital Carlos III, Madrid, Spain
Received 1 November 2008/
Accepted 26 January 2009

ABSTRACT
The plasma human immunodeficiency virus (HIV) RNA load is used
in the clinical routine for the monitoring of HIV infection
and the patient's response to antiretroviral therapy. Other
body fluids or dried blood spots (DBS) can be used, however,
to assess the level of viremia. The use of DBS may be especially
helpful for the monitoring of HIV-infected patients in resource-poor
settings, where access to adequate laboratory facilities is
often difficult. However, the correlation between the HIV RNA
levels in plasma and those in DBSs has not been well established.
Paired plasma and DBS samples obtained from HIV type 1 (HIV-1)-infected
patients were tested for HIV RNA copy numbers by using two different
commercial assays, the Nuclisens EasyQ HIV-1 (version 1.1) test
(the Nuclisens test; Biomerieux) and the m2000rt RealTime HIV
test (the m2000rt test; Abbott). Nucleic acid extraction was
performed manually by using either the Nuclisens isolation kit
(which uses the Boom methodology) or the m2000rt sample preparation
kit (an iron particle-based method). A total of 103 paired plasma
and DBS samples were tested. Viral load results were obtained
for 97 (94.2%) samples with the Nuclisens isolation kit and
81 (78.6%) samples with the m2000rt kit. The overall correlation
between the RNA loads in plasma and DBS was good, although better
results were obtained by the Nuclisens test (
R2 = 0.87,
P <
0.001) than by the m2000rt test (
R2 = 0.70,
P < 0.001). While
the specificities were excellent and similar for both the Nuclisens
and the m2000rt tests (97.1% and 100%, respectively), the sensitivity
was greater by the Nuclisens test than by the m2000rt test (75.8%
and 56.6%, respectively). Overall, the viral loads in DBS tended
to be lower than those in plasma, with mean differences of 0.3
log unit (standard deviation, 0.5 log unit) and 0.76 log unit
(standard deviation, 0.8 log unit) for the Nuclisens and the
m2000rt tests, respectively. The levels of agreement between
the measurements in plasma and DBS were assessed by using the
Bland-Altman plot for each assay. The Nuclisens test gave results
within its defined limits (–0.65 to 1.26) for 95.9% of
the samples, while the m2000rt test gave results within its
limits (–0.83 to 2.33) for 100% of the samples. In summary,
the HIV-1 load can accurately be quantified by testing DBS by
either the Nuclisens or the m2000rt test, although the Nuclisens
test may outperform the m2000rt test when nucleic acids are
extracted manually.

INTRODUCTION
Viral load measurement has turned out to be the major indicator
of antiretroviral treatment success or failure in human immunodeficiency
virus (HIV)-infected patients (
16). Accordingly, periodic viral
load monitoring is recommended for HIV-infected patients receiving
antiretroviral therapy. Several commercial tests have been approved
for the quantitation of HIV RNA in plasma, including (i) the
Cobas TaqMan HIV type 1 (HIV-1) test (Roche Molecular Systems,
Basel, Switzerland) (
24), (ii) the Nuclisens EasyQ HIV-1 (version
1.1) test (the Nuclisens test; Biomerieux, Durham, NC) (
12),
(iii) the branched DNA Versant HIV-1 RNA (version 3.0) assay
(Siemens, Berkeley, CA) (
14), and (iv) the m2000rt RealTime
HIV-1 test (the m2000rt test; Abbott, Chicago, IL) (
13,
30).
The implementation of antiretroviral therapy in developing countries through the Toward Universal Access program, carried out by the World Health Organization (35) in collaboration with UNAIDS and UNICEF (33), is making the availability of viral load testing of HIV-infected patients in resource-limited settings urgently necessary. Since viral load testing is relatively expensive and requires adequate facilities with complex equipment and specialized personnel, the shipment of plasma samples to reference laboratories has been proposed as an alternative to the development of new facilities with the necessary equipment and personnel. However, that option is complicated because of the requirement of cool temperatures for storage and transportation. Therefore, the use of dried blood spots (DBS) is increasingly being considered as an alternative that can be used to overcome the troubles inherent to the testing of plasma. Patient's blood placed on DBS can be shipped at room temperature to specialized reference laboratories, where the specimens may be subject to HIV RNA quantitation (1, 2, 3, 8, 11, 22), HIV subtyping, and/or drug resistance testing (5, 15, 19, 28). DBS have also been successfully used to assist with the confirmation of a diagnosis of HIV infection (10, 18, 27), measurement of CD4 counts (23), and even analysis of antiretroviral drug levels (17). Nevertheless, the sensitivity and the degree of correlation between the viral loads obtained with DBS and those obtained with plasma have been a matter of concern. We report here on the correlation between the viral load in plasma and that in DBS, as well as the lower limit of detection in DBS, obtained by two commercial assays, the Nuclisens assay (Biomerieux, Durham, NC) and the m2000rt assay (Abbott, Chicago, IL).

MATERIALS AND METHODS
Specimens.
A total of 103 whole-blood samples were collected from HIV-infected
patients receiving care at the Hospital Carlos III in Madrid,
Spain, and placed in tubes containing EDTA. DBS were prepared
by dropping 50 µl of blood in each circle (five spots
per card) of filter paper (S&S903; Schleicher & Schull,
BioScience GmbH, Barcelona, Spain) and were left to dry overnight
at room temperature. They were then stored at 4°C until
they were processed. Plasma aliquots were obtained by centrifugation
of the whole blood, and the supernatant was stored at –20°C
until use.
Before nucleic acid isolation, blood was eluted from the filter papers. For elution, scissors were used to cut one spot from the filter paper containing DBS for each patient (50 µl), and the filter paper spot was left in 2 ml of Nuclisens lysis buffer (Biomerieux, Madrid, Spain) for 2 h with gentle rotation. The filter paper was then removed from the tube, and the solution containing the material eluted from the DBS was further processed according to the manufacturer's instructions.
Viral load determination by Nuclisens assay.
Nucleic acid isolation was carried out by the method of Boom et al. (7), which is a silica-based technology. For DBS samples, extraction was performed manually with the Nuclisens isolation kit, whereas the Nuclisens EasyMag automatic extractor was used for the plasma specimens. Viral load quantitation was performed according to the manufacturer's instructions. The linear dynamic range of this assay ranges from 50 to 3,000,000 HIV RNA copies/ml when 1 ml of plasma is used (12).
Viral load determination by Abbott m2000rt assay.
RNA was extracted manually from both plasma and DBS samples by using the Abbott mSample preparation system, which uses an iron particle-based method. The viral load was then measured according to the manufacturer's instructions. The linear dynamic range of the assay ranges from 40 to 10,000,000 HIV RNA copies/ml when 1 ml of plasma is used (30).
Statistical analysis.
All values were log10 transformed before they were analyzed. Paired sample t test statistics were used for comparisons of the loads obtained with plasma and those obtained with DBS, as well as for comparison of the results obtained by the two distinct assays. For analysis of the concordance between the results obtained with plasma and those obtained with DBS, the method of Bland and Altman (6) was used. All analyses were performed by using SPSS statistical software (version 15.0; SPSS Inc., Chicago, IL). Differences were considered significant when P values were less than 0.05.

RESULTS
Correlation between viral load values obtained with DBS and those obtained with plasma by Nuclisens assay.
Viral load measurements were originally intended to be made
for 103 samples, but paired viral load results for DBS and plasma
could be obtained for only 97 samples (94.2%). Invalid results
for DBS were repeatedly obtained with four pairs of samples
and for both DBS and plasma for another two pairs of specimens.
Figure 1a shows the correlation between the viral load determinations obtained with plasma and those obtained with DBS. The overall R2 value was 0.87 (P < 0.001). Agreement between the undetectable and the detectable viral load results for plasma and DBS was seen for 83.5% of the samples (the true-positive rate), and the kappa coefficient was 0.67 (P < 0.001). This means that 15.8% of the DBS for which undetectable viremia was reported had measurable HIV RNA levels in plasma.
A deeper examination of these false-negative results obtained
with DBS showed a relationship of the results to a lower detection
limit threshold for DBS than for plasma. Overall, 97.7% of the
samples (42/43) with more than 3.48 log HIV RNA copies/ml in
plasma yielded detectable viremia by the use of DBS, while levels
over a range of 3.08 to 3.48 log HIV RNA copies/ml measurable
in plasma were measurable in only 60% (3/5) of the DBS samples.
The Bland-Altman plot (Fig. 2a) shows that the mean difference between the values obtained with DBS and those obtained with plasma by the Nuclisens assay was 0.31, with two times the standard deviation ranging from 0.65 to 1.27. All but four samples provided results within these limits. Overall, differences in viral load values of greater than 0.5 log unit obtained by testing DBS and plasma were seen for 33% of the specimens by the Nuclisens assay.
Correlation between viral load values obtained with DBS and those obtained with plasma by m2000rt assay.
Pairwise determinations of the viral loads in plasma and DBS
could be obtained for 81 of 103 samples (78.6%) that had been
intended to be analyzed. Invalid results were repeatedly obtained
for 13 DBS specimens and 8 plasma samples. Moreover, one sample
pair yielded invalid results with both the DBS and plasma. The
overall
R2 value between the measures obtained with DBS and
those obtained with plasma was 0.70 (
P < 0.001) (Fig.
1b).
Agreement between undetectable and detectable viral load results in plasma and DBS was recognized for 75.3% of the samples. Up to 23.8% of the specimens with measurable viremia in plasma yielded undetectable values when DBS were used. The kappa agreement of coefficient was 0.53 (P < 0.001).
Stratification of the samples into different plasma viral load intervals showed that viral loads greater than 3.72 log HIV RNA copies/ml were detectable by the use of DBS for 100% (22/22) of the specimens. In contrast, only 42.9% (three of seven) of the specimens with plasma viral loads between 3.23 and 3.72 log HIV RNA copies/ml had measurable viremia in DBS.
The Bland-Altman plot (Fig. 2b) showed that two times the standard deviation ranged from –0.80 to 2.34 for the Abbott m2000rt assay. All samples gave results within the limits defined by the standard deviation. The mean difference between measures obtained with DBS and those obtained with plasma was 0.77. Overall, 48.1% of the samples had values with differences greater than 0.5 log HIV RNA copies/ml when the levels obtained with DBS and those obtained with plasma were compared.
Correlation between Nuclisens and Abbott m2000rt assay results.
A comparison of the viral load results obtained by the Nuclisens and m2000rt assays showed a greater correlation when plasma was tested than when DBS were tested. The Pearson value for plasma determinations was 0.960 (P < 0.001) and the R2 value was 0.921, while for the determinations with DBS, the Pearson value was 0.845 (P < 0.001) and the R2 value was 0.714.

DISCUSSION
More than two-thirds of the estimated 34 million people living
with HIV infection worldwide reside in low-income countries.
About 3 million individuals in these countries currently have
access to antiretroviral treatment (
35). In this context, the
availability of viral load testing is crucial to ensure the
early recognition of treatment failure, which in turn favors
the success of rescue interventions and avoids the accumulation
of drug resistance mutations (
4,
9,
20,
26). Since viral load
testing requires specialized facilities and experienced personnel,
alternative strategies have been explored. Recent reports have
suggested that DBS may be suitable for use for the testing of
HIV RNA levels, although most studies have so far examined only
small sets of samples (
1,
2,
25,
32,
34). Both the reliability
and the sensitivity of tests that use DBS compared to those
that use plasma have been matters of concern.
In the present study, we analyzed the reliability of two commercial real-time amplification assays used for the determination of viral loads, the Nuclisens and m2000rt assays, with more than 100 paired DBS and plasma specimens. In all instances, the corresponding manual methods for nucleic acid extraction were used. Both the Nuclisens and the m2000rt assays provided results with DBSs comparable to those provided with plasma, although the Nuclisens assay tended to be more accurate. In a similar study conducted by Zazzi and colleagues (18a), the preliminary results obtained with 175 paired samples tested by the m2000rt assay were better than the results obtained in our study. It should be noted, however, that nucleic acid isolation was performed automatically in their study, while we used a manual protocol. Since almost all commercial viral load tests have automated extraction methods, it seems reasonable to conduct studies like ours by using those updated procedures. The Nuclisens automated nucleic acid system is called EasyMag and is based on the same chemical principle as the manual procedure, the method of Boom et al. (7). For the m2000rt viral load test (7), the use of an automated method for the extraction of nucleic acids may improve the performance over that obtained by manual isolation, which is particularly tedious.
The correlation between the results obtained with plasma and those obtained with DBS by either the Nuclisens or the m2000rt assay was satisfactory. Moreover, the R2 values between the two assays were 0.96 (P < 0.001) when plasma specimens were tested and 0.92 when DBS were tested. These results reinforce the suggestion that the measurement of viral loads in DBS by the use of either of these two commercial assays is reliable and may facilitate the more adequate monitoring of patients receiving antiretroviral treatment in resource-limited settings. A suspicion of antiretroviral treatment failure on the basis of only clinical manifestations or CD4 counts is clearly insufficient. A note of caution must also be kept in mind for the testing of HIV-1 non-B subtypes, which are the most prevalent in resource-limited settings (31), since the reliability of HIV RNA measurements in non-B subtypes for different HIV viral load assays is not uniform (12, 29).
The lower limit of detection for HIV RNA in DBS by either of the two commercial assays was greater than 3.5 log HIV RNA copies/ml. This threshold could be considered unsatisfactory, as the occurrence of a level of plasma viremia of greater than 3 log units while a patient is receiving antiretroviral therapy is well established to be associated with the selection of drug resistance and even clinical progression (21). Moreover, the World Health Organization has advised that drug resistance testing be performed for all patients with plasma HIV RNA loads greater than 1,000 copies/ml (36). The maintenance of a good correlation between the sensitivity obtained by the use of DBS and that obtained by the use of plasma in the lower viral load range seems to be important to support the use of DBS for viral load testing. Our study was carried out with just one blood spot for each determination, which means that the amount of blood used in the assay was only 50 µl. It would be interesting to examine whether the use of two blood spots for nucleic acid extraction would increase the amount of RNA and enhance the sensitivity of viral load testing with DBS. Moreover, the evaluation of viral load assays by the use of automated nucleic acid extraction protocols is warranted.
In summary, HIV RNA can be reliably quantified in DBS for samples with more than 3 log units by using either the Nuclisens or the m2000rt assay, although the first methodology seems to be the most precise, at least when manual nucleic acid extraction protocols are used. Further improvements in sensitivity are warranted to avoid the misrecognition of virological failures in patients who are receiving antiretroviral therapy and who have low-level viremia.

ACKNOWLEDGMENTS
This work was supported in part by grants from the Fundación
Investigación y Educación en SIDA, the Red de
Investigación en SIDA (project RD06/006), the Fondo de
Investigación Sanitaria (project CP06/0284 and PI06/1826),
and the Agencia Laín Entralgo.

FOOTNOTES
* Corresponding author. Mailing address: Department of Infectious Diseases, Hospital Carlos III, Calle Sinesio Delgado 10, Madrid 28029, Spain. Phone: 34 91 4532500. Fax: 34 91 7336614. E-mail:
cmendoza{at}teleline.es 
Published ahead of print on 4 February 2009. 

REFERENCES
1 - Alvarez-Muñoz, M. T., S. Zaragoza-Rodriguez, O. Rojas-Montes, G. Palacios-Saucedo, G. Vazquez-Rosales, A. Gomez-Delgado, J. Torres, and O. Muñoz. 2005. High correlation of human immunodeficiency virus type-1 viral load measured in dried blood spot samples and in plasma under different storage conditions. Arch. Med. Res. 36:382-386.[CrossRef][Medline]
2 - Amellal, B., C. Katlama, and V. Calvez. 2007. Evaluation of the use of dried spots and of different storage conditions of plasma for HIV-1 RNA quantification. HIV Med. 8:396-400.[CrossRef][Medline]
3 - Ayele, W., R. Schuurman, T. Messele, W. Dorigo-Zetsma, Y. Mengistu, J. Goudsmit, W. Paxton, M. P. de Baar, and G. Pollakis. 2007. Use of dried spots of whole blood, plasma, and mother's milk collected on filter paper for measurement of human immunodeficiency virus type 1 burden. J. Clin. Microbiol. 45:891-896.[Abstract/Free Full Text]
4 - Barth, R. E., A. M. Wensing, H. A. Tempelman, R. Moraba, R. Schuurman, and A. I. Hoepelman. 2008. Rapid accumulation of nonnucleoside reverse transcriptase inhibitor-associated resistance: evidence of transmitted resistance in rural South Africa. AIDS 22:2210-2212.[CrossRef][Medline]
5 - Bertagnolio, S., L. Soto-Ramirez, R. Pilon, R. Rodriguez, M. Viveros, L. Fuentes, P. R. Harrigan, T. Mo, D. Sutherland, and P. Sandstrom. 2007. HIV-1 drug resistance surveillance using dried whole blood spots. Antivir. Ther. 12:107-113.[Medline]
6 - Bland, J., and D. Altman. 1986. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet i:307-310.
7 - Boom, R., C. J. Sol, M. Salimans, C. Jansen, P. Wertheim-van Dillen, and J. van der Noordaa. 1990. Rapid and simple method for purification of nucleic acids. J. Clin. Microbiol. 28:495-503.[Abstract/Free Full Text]
8 - Brambilla, D., C. Jennings, G. Aldrovandi, J. Bremer, A. Comeau, S. Cassol, R. Dickover, J. Jackson, J. Pitt, J. Sullivan, A. Butcher, L. Grosso, P. Reichelderfer, and S. Fiscus. 2003. Multicenter evaluation of use of dried blood and plasma spot specimens in quantitative assays for human immunodeficiency virus RNA: measurement, precision, and RNA stability. J. Clin. Microbiol. 41:1888-1893.[Abstract/Free Full Text]
9 - Calmy, A., N. Ford, B. Hirschel, S. J. Reynolds, L. Lynen, E. Goemaere, F. Garcia de la Vega, L. Perrin, and W. Rodriguez. 2007. HIV viral load monitoring in resource-limited regions: optional or necessary? Clin. Infect. Dis. 44:128-134.[CrossRef][Medline]
10 - Cassol, S., N. Lapointe, T. Salas, C. Hankins, M. Arella, M. Fauvel, G. Delage, M. Boucher, J. Samson, and J. Charest. 1992. Diagnosis of vertical HIV-1 transmission using the polymerase chain reaction and dried blood spot specimens. J. Acquir. Immune Defic. Syndr. 5:113-119.[Medline]
11 - Cassol, S., M. J. Gill, R. Pilon, M. Cormier, R. Voigt, B. Willoughby, and J. Forbes. 1997. Quantification of human immunodeficiency virus type 1 RNA from dried plasma spots collected on filter paper. J. Clin. Microbiol. 35:2795-2801.[Abstract]
12 - De Mendoza, C., M. Koppelman, B. Montes, V. Ferre, V. Soriano, H. Cuypers, M. Segondy, and T. Oosterlaken. 2005. Multicenter evaluation of the NucliSens EasyQ HIV-1 v1.1 assay for the quantitative detection of HIV-1 RNA in plasma. J. Virol. Methods 127:54-59.[CrossRef][Medline]
13 - Garcia-Diaz, A., G. Clewley, C. L. Booth, W. Labett, N. McAllister, and A. M. Geretti. 2006. Comparative evaluation of the performance of the Abbott real-time human immunodeficiency virus type 1 (HIV-1) assay for measurement of HIV-1 plasma viral load following automated specimen preparation. J. Clin. Microbiol. 44:1788-1791.[Abstract/Free Full Text]
14 - Gomes, P., A. Palma, J. Cabanas, A. Abecasis, A. Carvalho, R. Ziermann, I. Diogo, F. Gonçalves, C. Lobo, and R. Camacho. 2006. Comparison of the COBAS TAQMAN HIV-1 HPS with VERSANT HIV-1 RNA 3.0 assay (bDNA) for plasma RNA quantitation in different HIV-1 subtypes. J. Virol. Methods 135:223-228.[CrossRef][Medline]
15 - Hallack, R., L. Doherty, J. Wethers, and M. Parker. 2008. Evaluation of dried blood spots specimens for HIV-1 drug resistance testing using the Trugene HIV-1 genotyping assay. J. Clin. Virol. 41:283-287.[CrossRef][Medline]
16 - Hammer, S., J. Eron, P. Reiss, R. Schooley, M. Thompson, S. Walmsley, P. Cahn, M. A. Fischl, J. M. Gatell, M. S. Hirsch, D. M. Jacobsen, J. S. Montaner, D. D. Richman, P. G. Yeni, and P. A. Volberding. 2008. Antiretroviral treatment of adult HIV infection. 2008 recommendations of the International AIDS Society—USA Panel. JAMA 300:555-570.[Abstract/Free Full Text]
17 - Koal, T., H. Burhenne, R. Römling, M. Svoboda, K. Resch, and V. Kaever. 2005. Quantification of antiretroviral drugs in dried blood spot samples by means of liquid chromatography/tandem mass spectrometry. Rapid Commun. Mass Spectrom. 19:2995-3001.[CrossRef][Medline]
18 - Lakshmi, V., T. Sudha, M. Bhanurekha, and L. Dandona. 2007. Evaluation of the Murex HIV Ag/Ab Combination assay when used with dried blood spots. Clin. Microbiol. Infect. 13:1134-1136.[CrossRef][Medline]
18 - Marconi, A., M. Balestrieri, G. Comastri, F. Pulvirenti, W. Gennari, S. Tagliazucchi, M. Pecorari, V. Borghi, D. Marri, and M. Zazzi. 2009. Evaluation of the Abbott Real-Time HIV-1 quantitative assay with dried blood spot specimens. Clin. Microbiol. Infect. 15:93-97.[CrossRef][Medline]
19 - Masciotra, S., C. Garrido, A. Youngpairoj, A. McNulty, N. Zahonero, A. Corral, W. Heneine, C. de Mendoza, and J. García-Lerma. 2007. High concordance between HIV-1 drug resistance genotypes generated from plasma and dried blood spots in antiretroviral-experienced patients. AIDS 21:2503-2511.[Medline]
20 - Mee, P., K. Fielding, S. Charalambous, G. Churchyard, and A. Grant. 2008. Evaluation of the WHO criteria for antiretroviral treatment failure among adults in South Africa. AIDS 22:1971-1977.[CrossRef][Medline]
21 - Murri, R., A. C. Lepri, P. Cicconi, A. Poggio, M. Arlotti, G. Tositti, D. Santoro, M. L. Soranzo, G. Rizzardini, V. Colangeli, M. Montroni, A. D. Monforte, and the ICoNA Study Group. 2006. Is moderate HIV viremia associated with a higher risk of clinical progression in HIV-infected people treated with highly active antiretroviral therapy: evidence from the Italian cohort of antiretroviral-naive patients study. J. Acquir. Immune Defic. Syndr. 41:23-30.[CrossRef][Medline]
22 - Mwaba, P., S. Cassol, A. Nunn, R. Pilon, C. Chintu, M. Janes, and A. Zumla. 2003. Whole blood versus plasma spots for measurement of HIV-1 viral load in HIV-infected African patients. Lancet 362:2067-2068.[CrossRef][Medline]
23 - Mwaba, P., S. Cassol, R. Pilon, C. Chintu, M. Janes, A. Nunn, and A. Zumla. 2003. Use of dried whole blood spots to measure CD4+ lymphocyte counts in HIV-1-infected patients. Lancet 362:1459-1460.[CrossRef][Medline]
24 - Oliver, A., S. Pereira, and D. Clark. 2007. Comparative evaluation of the automated Roche TaqMan real-time quantitative human immunodeficiency virus type 1 RNA PCR assay and the Roche AMPLICOR version 1.5 conventional PCR assay. J. Clin. Microbiol. 45:3616-3619.[Abstract/Free Full Text]
25 - O'Shea, S., J. Mullen, K. Corbett, I. Chrystie, M. L. Newell, and J. Banatvala. 1999. Use of dried whole blood spots for quantification of HIV-1 RNA. AIDS 13:630-631.[CrossRef][Medline]
26 - Petersen, M. L., M. J. van der Laan, S. Napravnik, J. J. Eron, R. D. Moore, and S. G. Deeks. 2008. Long-term consequences of the delay between virologic failure of highly active antiretroviral therapy and regimen modification. AIDS 22:2097-2106.[CrossRef][Medline]
27 - Sarge-Njie, R., M. Schim Van Der Loeff, S. Ceesay, D. Cubitt, S. Sabally, T. Corrah, and H. Whittle. 2006. Evaluation of the dried blood spot filter paper technology and five testing strategies of HIV-1 and HIV-2 infections in West Africa. Scand. J. Infect. Dis. 38:1050-1056.[CrossRef][Medline]
28 - Steegen, K., S. Luchters, E. Demecheleer, K. Dauwe, K. Mandaliya, W. Jaoko, J. Plum, M. Temmerman, and C. Verhofstede. 2007. Feasibility of detecting HIV-1 drug resistance in DNA extracted from whole blood or dried blood spots. J. Clin. Microbiol. 45:3342-3351.[Abstract/Free Full Text]
29 - Swanson, P., C. de Mendoza, Y. Joshi, A. Golden, R. Hodinka, V. Soriano, S. Devare, and J. Hackett. 2005. Impact of human immunodeficiency virus type 1 (HIV-1) genetic diversity on performance of four commercial viral load assays: LCx HIV RNA Quantitative, AMPLICOR HIV-1 MONITOR v1.5, VERSANT HIV-1 RNA 3.0, and NucliSens HIV-1 QT. J. Clin. Microbiol. 43:3860-3868.[Abstract/Free Full Text]
30 - Swanson, P., V. Holzmayer, S. Huang, P. Hay, A. Adebiyi, P. Rice, K. Abravaya, S. Thamm, S. Devare, and J. Hackett. 2006. Performance of the automated Abbott Real Time HIV-1 assay on a genetically diverse panel of specimens from London: comparison to VERSANT HIV-1 RNA 3.0, AMPLICOR HIV-1 MONITOR v1.5, and LCx HIV RNA quantitative assays. J. Virol. Methods 137:184-192.[CrossRef][Medline]
31 - Taylor, B., M. Sobieszczyk, F. McCutchan, and S. Hammer. 2008. The challenge of HIV-1 subtype diversity. N. Engl. J. Med. 358:1590-1602.[Free Full Text]
32 - Toure-Kane, C., H. Ndiaye, S. Diallo, I. Ndiaye, A. Wade, P. Diaw, A. Gaye-Diallo, and S. Mboup. 2008. Quantitation of HIV-1 RNA in dried blood spots by the real-time Nuclisens EasyQ HIV-1 assay in Senegal. J. Virol. Methods 148:291-295.[CrossRef][Medline]
33 - United Nations General Assembly. 2006. Political declaration on HIV/AIDS. United Nations General Assembly document 60/262. United Nations, New York, NY. http://www.unaids.org/en/AboutUNAIDS/Goals/UNGASS.
34 - Uttayamakul, S., S. Likanonsakul, R. Sunthornkachit, K. Kuntiranont, S. Louisirirotchanakul, A. Chaovavanich, V. Thiamchai, S. Tanprasertsuk, and R. Sutthent. 2005. Usage of dried blood spots for molecular diagnosis and monitoring HIV-1 infection. J. Virol. Methods 128:128-134.[CrossRef][Medline]
35 - World Health Organization. 2008. Towards Universal Access. Scaling up priority HIV/AIDS interventions in the health sector. Progress report June 2008. World Health Organization, Geneva, Switzerland. http://www.who.int/hiv/pub/2008progressreport/en/index.html.
36 - World Health Organization. 2006. Antiretroviral therapy for HIV infection in adults and adolescents: recommendations for a public health approach. 2006 revision. World Health Organization, Geneva, Switzerland. http://www.who.int/hiv/pub/guidelines/adult/en/index.html.
Journal of Clinical Microbiology, April 2009, p. 1031-1036, Vol. 47, No. 4
0095-1137/09/$08.00+0 doi:10.1128/JCM.02099-08
Copyright © 2009, American Society for Microbiology. All Rights Reserved.
This article has been cited by other articles:
-
Johannessen, A., Troseid, M., Calmy, A.
(2009). Dried blood spots can expand access to virological monitoring of HIV treatment in resource-limited settings. J Antimicrob Chemother
64: 1126-1129
[Abstract]
[Full Text]