Previous Article | Next Article ![]()
Journal of Clinical Microbiology, Apr 1997, 976-977, Vol 35, No. 4
HH Kessler, B Santner, H Rabenau, A Berger, A Vince, C Lewinski, B Weber, K Pierer, D Stuenzner, E Marth and HW Doerr
The AMPLICOR Enterovirus Test was evaluated with 103 cerebrospinal fluid
(CSF) specimens. Twenty-seven CSF specimens were culture positive. With the
AMPLICOR test, enterovirus RNA was detected in 34 specimens. Compared with
culture, the AMPLICOR test gave a sensitivity of 96.3% and a specificity of
100%. The sensitivity of culture was 79.4% in comparison with the AMPLICOR
test.
Copyright © 1997 by the American Society for Microbiology. All rights reserved.
Rapid diagnosis of enterovirus infection by a new one-step reverse transcription-PCR assay
Institute of Hygiene, KF-University Graz, Austria. harald.kessler@kfunigraz.ac.at
This article has been cited by other articles:
Copyright © 2009 by the American Society for Microbiology. For an alternate route to Journals.ASM.org, visit: http://intl-journals.asm.org | More Info»