This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowReprints and Permissions
Right arrow Copyright Information
Right arrow Books from ASM Press
Right arrow MicrobeWorld
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Zaidi, A. K.
Right arrow Articles by Reller, L. B.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Zaidi, A. K.
Right arrow Articles by Reller, L. B.

 Previous Article  |  Next Article 

Journal of Clinical Microbiology, Aug 1997, 2007-2012, Vol 35, No. 8
Copyright © 1997 by the American Society for Microbiology. All rights reserved.

Controlled comparison of bioMerieux VITAL and BACTEC NR-660 systems for detection of bacteremia and fungemia in pediatric patients

AK Zaidi, S Mirrett, JC McDonald, EE Rubin, LC McDonald, MP Weinstein, M Gupta and LB Reller
Clinical Microbiology Laboratory, Duke University Medical Center, Durham, North Carolina 27710, USA.

The bioMerieux VITAL automated blood culture system measures a decrease in fluorescence to detect the presence of microorganisms in blood. To assess the performance of VITAL with AER aerobic medium versus that of the nonradiometric BACTEC NR-660 PEDS PLUS medium for the detection of sepsis in children, a total of 12,146 blood specimens were collected at three university medical centers and inoculated into AER and PEDS PLUS bottles that were weighed before and after filling. The sample volumes were considered adequate in 6,276 bottle pairs. The total yield of isolates was 629, of which 489 (78%) were judged to be the cause of true infections. Staphylococci (P < 0.001) and yeasts (P < 0.05) were detected more often in PEDS PLUS bottles, as were all microorganisms combined (P < 0.001). The improved detection in the PEDS PLUS medium was most marked for patients on antimicrobial therapy (P < 0.001), but remained statistically significant even for patients not on therapy (P < 0.025). There were 431 episodes of sepsis, including 407 considered adequate for analysis. Of the 363 unimicrobial episodes, 278 were detected by both bottles, 64 were detected by PEDS PLUS bottles only, and 21 were detected by AER bottles only (P < 0.01). No false-negative cultures were detected by terminal subculture of the PEDS PLUS bottles when the companion AER bottle was positive. However, there were 14 false-negative cultures (7 yeasts, 5 staphylococci, 1 Enterococcus faecalis, and 1 Enterobacter sp.) on terminal subculture of the AER bottles when the companion PEDS PLUS bottle was positive. When both systems were positive, the VITAL system detected bacteria earlier than did the BACTEC system by a mean of 1.6 h. Also, false-positive signals were less common with the VITAL system. We conclude that the VITAL system with AER medium must be modified to improve the detection of clinically important staphylococci and yeasts if it is to perform comparably to the BACTEC NR-660 nonradiometric system with PEDS PLUS medium for a pediatric population.


This article has been cited by other articles:

  • Reller, M. E., Zaidi, A. K. M., Sultana, S., Azeem, S., Hanif, B., Qureshi, S., Hasan, R., Bhutta, Z., Akhter, R., Goldmann, D. A. (2009). Controlled Evaluation of Bactec Peds Plus/F and Bactec Lytic/10 Anaerobic/F Media for Isolation of Salmonella enterica Serovars Typhi and Paratyphi A from Blood. J. Clin. Microbiol. 47: 245-246 [Abstract] [Full Text]  