Journal of Clinical Microbiology, Aug 1997, 2007-2012, Vol 35, No. 8
AK Zaidi, S Mirrett, JC McDonald, EE Rubin, LC McDonald, MP Weinstein, M Gupta and LB Reller
The bioMerieux VITAL automated blood culture system measures a decrease in
fluorescence to detect the presence of microorganisms in blood. To assess
the performance of VITAL with AER aerobic medium versus that of the
nonradiometric BACTEC NR-660 PEDS PLUS medium for the detection of sepsis
in children, a total of 12,146 blood specimens were collected at three
university medical centers and inoculated into AER and PEDS PLUS bottles
that were weighed before and after filling. The sample volumes were
considered adequate in 6,276 bottle pairs. The total yield of isolates was
629, of which 489 (78%) were judged to be the cause of true infections.
Staphylococci (P < 0.001) and yeasts (P < 0.05) were detected more
often in PEDS PLUS bottles, as were all microorganisms combined (P <
0.001). The improved detection in the PEDS PLUS medium was most marked for
patients on antimicrobial therapy (P < 0.001), but remained
statistically significant even for patients not on therapy (P < 0.025).
There were 431 episodes of sepsis, including 407 considered adequate for
analysis. Of the 363 unimicrobial episodes, 278 were detected by both
bottles, 64 were detected by PEDS PLUS bottles only, and 21 were detected
by AER bottles only (P < 0.01). No false-negative cultures were detected
by terminal subculture of the PEDS PLUS bottles when the companion AER
bottle was positive. However, there were 14 false-negative cultures (7
yeasts, 5 staphylococci, 1 Enterococcus faecalis, and 1 Enterobacter sp.)
on terminal subculture of the AER bottles when the companion PEDS PLUS
bottle was positive. When both systems were positive, the VITAL system
detected bacteria earlier than did the BACTEC system by a mean of 1.6 h.
Also, false-positive signals were less common with the VITAL system. We
conclude that the VITAL system with AER medium must be modified to improve
the detection of clinically important staphylococci and yeasts if it is to
perform comparably to the BACTEC NR-660 nonradiometric system with PEDS
PLUS medium for a pediatric population.
Copyright © 1997 by the American Society for Microbiology. All rights reserved.
Controlled comparison of bioMerieux VITAL and BACTEC NR-660 systems for detection of bacteremia and fungemia in pediatric patients
Clinical Microbiology Laboratory, Duke University Medical Center, Durham, North Carolina 27710, USA.
This article has been cited by other articles:
Copyright © 2009 by the American Society for Microbiology. For an alternate route to Journals.ASM.org, visit: http://intl-journals.asm.org | More Info»