Previous Article | Next Article 
Journal of Clinical Microbiology, November 1998, p. 3372-3374, Vol. 36, No. 11
0095-1137/98/$04.00+0
Copyright © 1998, American Society for Microbiology. All rights reserved.
Evaluation of the Nuclisens HIV-1 QT Assay for
Quantitation of Human Immunodeficiency Virus Type 1 RNA Levels
in Plasma
Michel
Segondy,1,*
Thoai-Duong
Ly,2
Martine
Lapeyre,2 and
Brigitte
Montes1
Laboratoire de Virologie, Centre Hospitalier
Universitaire de Montpellier, Montpellier,1
and
Institut Fournier, Paris,2 France
Received 20 March 1998/Accepted 10 August 1998
Nuclisens HIV-1 QT is a new version of the NASBA HIV-1 QT assay for
quantitation of human immunodeficiency virus type 1 (HIV-1) RNA in
plasma. The specificity of this assay was 100% in one laboratory and
99%
with nonrepeatability of the initial false positive
in another.
The test was linear between 2.0 and 6.0 log RNA copies per ml.
According to the input HIV-1 RNA concentration, accuracy varied from
0.11 to +0.10 log RNA copy per ml and precision varied from 0.66 to
0.14 log RNA copy per ml. Reproducibility decreased when the HIV-1 RNA
level was near the lower limit of quantitation of the test. HIV-1 RNA
could be quantitated by Nuclisens HIV-1 QT in 36% (laboratory 1) and
24% (laboratory 2) of clinical samples with HIV-1 RNA levels lower
than the lower limit of quantitation by NASBA HIV-1 QT. Nuclisens HIV-1
QT was not suitable for measurement of RNA from clade G and group O
HIV-1 strains.
*
Corresponding author. Mailing address: Laboratoire de
Virologie, Hôpital Saint-Eloi, Centre Hospitalier Universitaire,
34295 Montpellier Cedex 5, France. Phone: (33) 467-337127. Fax: (33) 467-337623. E-mail: msegondy{at}worldnet.fr.
Journal of Clinical Microbiology, November 1998, p. 3372-3374, Vol. 36, No. 11
0095-1137/98/$04.00+0
Copyright © 1998, American Society for Microbiology. All rights reserved.
This article has been cited by other articles:
-
Foulongne, V., Montes, B., Didelot-Rousseau, M.-N., Segondy, M.
(2006). Comparison of the LCx Human Immunodeficiency Virus (HIV) RNA Quantitative, RealTime HIV, and COBAS AmpliPrep-COBAS TaqMan Assays for Quantitation of HIV Type 1 RNA in Plasma.. J. Clin. Microbiol.
44: 2963-2966
[Abstract]
[Full Text]
-
Swanson, P., de Mendoza, C., Joshi, Y., Golden, A., Hodinka, R. L., Soriano, V., Devare, S. G., Hackett, J. Jr.
(2005). Impact of Human Immunodeficiency Virus Type 1 (HIV-1) Genetic Diversity on Performance of Four Commercial Viral Load Assays: LCx HIV RNA Quantitative, AMPLICOR HIV-1 MONITOR v1.5, VERSANT HIV-1 RNA 3.0, and NucliSens HIV-1 QT. J. Clin. Microbiol.
43: 3860-3868
[Abstract]
[Full Text]
-
Antunes, R., Figueiredo, S., Bartolo, I., Pinheiro, M., Rosado, L., Soares, I., Lourenco, H., Taveira, N.
(2003). Evaluation of the Clinical Sensitivities of Three Viral Load Assays with Plasma Samples from a Pediatric Population Predominantly Infected with Human Immunodeficiency Virus Type 1 Subtype G and BG Recombinant Forms. J. Clin. Microbiol.
41: 3361-3367
[Abstract]
[Full Text]
-
de Baar, M. P., van Dooren, M. W., de Rooij, E., Bakker, M., van Gemen, B., Goudsmit, J., de Ronde, A.
(2001). Single Rapid Real-Time Monitored Isothermal RNA Amplification Assay for Quantification of Human Immunodeficiency Virus Type 1 Isolates from Groups M, N, and O. J. Clin. Microbiol.
39: 1378-1384
[Abstract]
[Full Text]
-
Tang, Y.-W., Huong, J. T.-J., Lloyd, R. M. Jr., Spearman, P., Haas, D. W.
(2000). Comparison of Human Immunodeficiency Virus Type 1 RNA Sequence Heterogeneity in Cerebrospinal Fluid and Plasma. J. Clin. Microbiol.
38: 4637-4639
[Abstract]
[Full Text]
-
Murphy, D. G., Côté, L., Fauvel, M., René, P., Vincelette, J.
(2000). Multicenter Comparison of Roche COBAS AMPLICOR MONITOR Version 1.5, Organon Teknika NucliSens QT with Extractor, and Bayer Quantiplex Version 3.0 for Quantification of Human Immunodeficiency Virus Type 1 RNA in Plasma. J. Clin. Microbiol.
38: 4034-4041
[Abstract]
[Full Text]
-
Zanchetta, N., Nardi, G., Tocalli, L., Drago, L., Bossi, C., Pulvirenti, F. R., Galli, C., Gismondo, M. R.
(2000). Evaluation of the Abbott LCx HIV-1 RNA Quantitative, a New Assay for Quantitative Determination of Human Immunodeficiency Virus Type 1 RNA. J. Clin. Microbiol.
38: 3882-3886
[Abstract]
[Full Text]
-
Lerma, J. G. G., Soriano, V., Mas, A., Quiñones-Mateu, M. E., Arts, E. J., Heneine, W.
(2000). Quantitation of Human Immunodeficiency Virus Type 1 Group O Load in Plasma by Measuring Reverse Transcriptase Activity. J. Clin. Microbiol.
38: 402-405
[Abstract]
[Full Text]