Previous Article | Next Article 
Journal of Clinical Microbiology, March 1999, p. 792-795, Vol. 37, No. 3
0095-1137/99/$04.00+0
Copyright © 1999, American Society for Microbiology. All rights reserved.
Evaluation of the Ultrasensitive Roche Amplicor HIV-1 Monitor
Assay for Quantitation of Human Immunodeficiency Virus Type 1 RNA
Maria
Erali1 and
David R.
Hillyard1,2,*
ARUP Institute for Clinical and Experimental
Pathology,1 and the
Department of
Pathology, University of Utah Health Sciences
Center,2 Salt Lake City, Utah 84108
Received 21 July 1998/Returned for modification 28 September
1998/Accepted 8 December 1998
The ultrasensitive Amplicor HIV-1 Monitor test (Roche Diagnostic
Systems) was evaluated for precision, linearity, and sensitivity and
was compared to the standard Amplicor assay. The ultrasensitive assay
reliably quantified samples in the range from 50 to 50,000 human
immunodeficiency virus type 1 RNA copies/ml with acceptable correlation
with the standard Amplicor test.
*
Corresponding author. Mailing address: ARUP
Laboratories, 500 Chipeta Way, Salt Lake City, UT 84108. Phone: (801)
583-2787, ext. 2202. Fax: (801) 584-5207. E-mail:
hillyadr{at}arup-lab.com.
Journal of Clinical Microbiology, March 1999, p. 792-795, Vol. 37, No. 3
0095-1137/99/$04.00+0
Copyright © 1999, American Society for Microbiology. All rights reserved.
This article has been cited by other articles:
-
Foulongne, V., Montes, B., Didelot-Rousseau, M.-N., Segondy, M.
(2006). Comparison of the LCx Human Immunodeficiency Virus (HIV) RNA Quantitative, RealTime HIV, and COBAS AmpliPrep-COBAS TaqMan Assays for Quantitation of HIV Type 1 RNA in Plasma.. J. Clin. Microbiol.
44: 2963-2966
[Abstract]
[Full Text]
-
Salimnia, H., Moore, E. C., Crane, L. R., MacArthur, R. D., Fairfax, M. R.
(2005). Discordance between Viral Loads Determined by Roche COBAS AMPLICOR Human Immunodeficiency Virus Type 1 Monitor (Version 1.5) Standard and Ultrasensitive Assays Caused by Freezing Patient Plasma in Centrifuged Becton-Dickinson Vacutainer Brand Plasma Preparation Tubes. J. Clin. Microbiol.
43: 4635-4639
[Abstract]
[Full Text]
-
Sivapalasingam, S., Essajee, S., Nyambi, P. N., Itri, V., Hanna, B., Holzman, R., Valentine, F.
(2005). Human Immunodeficiency Virus (HIV) Reverse Transcriptase Activity Correlates with HIV RNA Load: Implications for Resource-Limited Settings. J. Clin. Microbiol.
43: 3793-3796
[Abstract]
[Full Text]
-
Swanson, P., de Mendoza, C., Joshi, Y., Golden, A., Hodinka, R. L., Soriano, V., Devare, S. G., Hackett, J. Jr.
(2005). Impact of Human Immunodeficiency Virus Type 1 (HIV-1) Genetic Diversity on Performance of Four Commercial Viral Load Assays: LCx HIV RNA Quantitative, AMPLICOR HIV-1 MONITOR v1.5, VERSANT HIV-1 RNA 3.0, and NucliSens HIV-1 QT. J. Clin. Microbiol.
43: 3860-3868
[Abstract]
[Full Text]
-
Muller, Z., Stelzl, E., Bozic, M., Haas, J., Marth, E., Kessler, H. H.
(2004). Evaluation of Automated Sample Preparation and Quantitative PCR LCx Assay for Determination of Human Immunodeficiency Virus Type 1 RNA. J. Clin. Microbiol.
42: 1439-1443
[Abstract]
[Full Text]
-
Holzl, G., Stocher, M., Leb, V., Stekel, H., Berg, J.
(2003). Entirely Automated Quantification of Human Immunodeficiency Virus Type 1 (HIV-1) RNA in Plasma by Using the Ultrasensitive COBAS AMPLICOR HIV-1 Monitor Test and RNA Purification on the MagNA Pure LC Instrument. J. Clin. Microbiol.
41: 1248-1251
[Abstract]
[Full Text]
-
de Baar, M. P., Timmermans, E. C., Bakker, M., de Rooij, E., van Gemen, B., Goudsmit, J.
(2001). One-Tube Real-Time Isothermal Amplification Assay To Identify and Distinguish Human Immunodeficiency Virus Type 1 Subtypes A, B, and C and Circulating Recombinant Forms AE and AG. J. Clin. Microbiol.
39: 1895-1902
[Abstract]
[Full Text]
-
Swanson, P., Soriano, V., Devare, S. G., Hackett, J. Jr.
(2001). Comparative Performance of Three Viral Load Assays on Human Immunodeficiency Virus Type 1 (HIV-1) Isolates Representing Group M (Subtypes A to G) and Group O: LCx HIV RNA Quantitative, AMPLICOR HIV-1 MONITOR Version 1.5, and Quantiplex HIV-1 RNA Version 3.0. J. Clin. Microbiol.
39: 862-870
[Abstract]
[Full Text]
-
Brambilla, D. J., Granger, S., Jennings, C., Bremer, J. W.
(2001). Multisite Comparison of Reproducibility and Recovery from the Standard and Ultrasensitive Roche AMPLICOR HIV-1 MONITOR Assays. J. Clin. Microbiol.
39: 1121-1123
[Abstract]
[Full Text]
-
Erice, A., Brambilla, D., Bremer, J., Jackson, J. B., Kokka, R., Yen-Lieberman, B., Coombs, R. W.
(2000). Performance Characteristics of the QUANTIPLEX HIV-1 RNA 3.0 Assay for Detection and Quantitation of Human Immunodeficiency Virus Type 1 RNA in Plasma. J. Clin. Microbiol.
38: 2837-2845
[Abstract]
[Full Text]
-
Witt, D. J., Kemper, M., Stead, A., Ginocchio, C. C., Caliendo, A. M.
(2000). Relationship of Incremental Specimen Volumes and Enhanced Detection of Human Immunodeficiency Virus Type 1 RNA with Nucleic Acid Amplification Technology. J. Clin. Microbiol.
38: 85-89
[Abstract]
[Full Text]