Previous Article | Next Article 
Journal of Clinical Microbiology, June 2000, p. 2108-2111, Vol. 38, No. 6
0095-1137/00/$04.00+0
Copyright © 2000, American Society for Microbiology. All rights reserved.
Evaluation of a New System, VITEK 2, for
Identification and Antimicrobial Susceptibility Testing of
Enterococci
Fernando
Garcia-Garrote,
Emilia
Cercenado,* and
Emilio
Bouza
Servicio de Microbiología, Hospital
General Universitario "Gregorio Marañón," Madrid, Spain
Received 3 January 2000/Returned for modification 2 February
2000/Accepted 9 March 2000
We evaluated the new automated VITEK 2 system (bioMérieux)
for the identification and antimicrobial susceptibility testing of
enterococci. The results obtained with the VITEK 2 system were compared
to those obtained by reference methods: standard identification by the
scheme of Facklam and Sahm [R. R. Facklam and D. F. Sahm, p.
308-314, in P. R. Murray et al., ed., Manual of
Clinical Microbiology, 6th ed., 1995] and with the API 20 STREP
system and, for antimicrobial susceptibility testing, broth
microdilution and agar dilution methods by the procedures of the
National Committee for Clinical Laboratory Standards. The presence of
vanA and vanB genes was determined by PCR. A
total of 150 clinical isolates were studied, corresponding to 60 Enterococcus faecalis, 55 Enterococcus faecium, 26 Enterococcus gallinarum, 5 Enterococcus
avium, 2 Enterococcus durans, and 2 Enterococcus raffinosus isolates. Among those isolates, 131 (87%) were correctly identified to the species level with the VITEK 2 system. Approximately half of the misidentifications were for E. faecium with low-level resistance to vancomycin, identified as
E. gallinarum or E. casseliflavus; however, a
motility test solved the discrepancies and increased the agreement to
94%. Among the strains studied, 66% were vancomycin resistant (57 VanA, 16 VanB, and 26 VanC strains), 23% were ampicillin resistant
(MICs,
16 µg/ml), 31% were high-level gentamicin resistant, and
45% were high-level streptomycin resistant. Percentages of agreement for susceptibility and resistance to ampicillin, vancomycin, and teicoplanin and for high-level gentamicin resistance and high-level streptomycin resistance were 93, 95, 97, 97, and 96%, respectively. The accuracy of identification and antimicrobial susceptibility testing
of enterococci with the VITEK 2 system, together with the significant
reduction in handling time, will have a positive impact on the work
flow of the clinical microbiology laboratory.
*
Corresponding author. Mailing address: Servicio de
Microbiología, Hospital General Universitario "Gregorio
Marañón," Dr. Esquerdo 46, 28007 Madrid, Spain. Phone:
34-91-586-8459. Fax: 34-91-504-4906. E-mail:
fgarrote{at}efd.net.
Journal of Clinical Microbiology, June 2000, p. 2108-2111, Vol. 38, No. 6
0095-1137/00/$04.00+0
Copyright © 2000, American Society for Microbiology. All rights reserved.
This article has been cited by other articles:
-
Wellinghausen, N., Bartel, M., Essig, A., Poppert, S.
(2007). Rapid Identification of Clinically Relevant Enterococcus Species by Fluorescence In Situ Hybridization. J. Clin. Microbiol.
45: 3424-3426
[Abstract]
[Full Text]
-
Brigante, G., Luzzaro, F., Bettaccini, A., Lombardi, G., Meacci, F., Pini, B., Stefani, S., Toniolo, A.
(2006). Use of the Phoenix Automated System for Identification of Streptococcus and Enterococcus spp.. J. Clin. Microbiol.
44: 3263-3267
[Abstract]
[Full Text]
-
Abele-Horn, M., Hommers, L., Trabold, R., Frosch, M.
(2006). Validation of VITEK 2 Version 4.01 Software for Detection, Identification, and Classification of Glycopeptide-Resistant Enterococci. J. Clin. Microbiol.
44: 71-76
[Abstract]
[Full Text]
-
Kanemitsu, K., Kunishima, H., Hatta, M., Inden, K., Saga, T., Ouchi, H., Ishizawa, K., Harigae, H., Takemura, H., Kaku, M.
(2005). Evaluation of a Fully Automated System (RAISUS) for Rapid Identification and Antimicrobial Susceptibility Testing of Staphylococci. J. Clin. Microbiol.
43: 5808-5810
[Abstract]
[Full Text]
-
Eigner, U., Schmid, A., Wild, U., Bertsch, D., Fahr, A.-M.
(2005). Analysis of the Comparative Workflow and Performance Characteristics of the VITEK 2 and Phoenix Systems. J. Clin. Microbiol.
43: 3829-3834
[Abstract]
[Full Text]
-
Funke, G., Funke-Kissling, P.
(2005). Performance of the New VITEK 2 GP Card for Identification of Medically Relevant Gram-Positive Cocci in a Routine Clinical Laboratory. J. Clin. Microbiol.
43: 84-88
[Abstract]
[Full Text]
-
Tsai, J.-C., Hsueh, P.-R., Lin, H.-M., Chang, H.-J., Ho, S.-W., Teng, L.-J.
(2005). Identification of Clinically Relevant Enterococcus Species by Direct Sequencing of groES and Spacer Region. J. Clin. Microbiol.
43: 235-241
[Abstract]
[Full Text]
-
Kobayashi, I., Muraoka, H., Iyoda, T., Nishida, M., Hasegawa, M., Yamaguchi, K.
(2004). Antimicrobial susceptibility testing of vancomycin-resistant Enterococcus by the VITEK 2 system, and comparison with two NCCLS reference methods. J Med Microbiol
53: 1229-1232
[Abstract]
[Full Text]
-
Jackson, C. R., Fedorka-Cray, P. J., Barrett, J. B.
(2004). Use of a Genus- and Species-Specific Multiplex PCR for Identification of Enterococci. J. Clin. Microbiol.
42: 3558-3565
[Abstract]
[Full Text]
-
de Cueto, M., Ceballos, E., Martinez-Martinez, L., Perea, E. J., Pascual, A.
(2004). Use of Positive Blood Cultures for Direct Identification and Susceptibility Testing with the Vitek 2 System. J. Clin. Microbiol.
42: 3734-3738
[Abstract]
[Full Text]
-
Donay, J.-L., Mathieu, D., Fernandes, P., Pregermain, C., Bruel, P., Wargnier, A., Casin, I., Weill, F. X., Lagrange, P. H., Herrmann, J. L.
(2004). Evaluation of the Automated Phoenix System for Potential Routine Use in the Clinical Microbiology Laboratory. J. Clin. Microbiol.
42: 1542-1546
[Abstract]
[Full Text]
-
Bruins, M. J., Bloembergen, P., Ruijs, G. J. H. M., Wolfhagen, M. J. H. M.
(2004). Identification and Susceptibility Testing of Enterobacteriaceae and Pseudomonas aeruginosa by Direct Inoculation from Positive BACTEC Blood Culture Bottles into Vitek 2. J. Clin. Microbiol.
42: 7-11
[Abstract]
[Full Text]
-
Fahr, A.-M., Eigner, U., Armbrust, M., Caganic, A., Dettori, G., Chezzi, C., Bertoncini, L., Benecchi, M., Menozzi, M. G.
(2003). Two-Center Collaborative Evaluation of the Performance of the BD Phoenix Automated Microbiology System for Identification and Antimicrobial Susceptibility Testing of Enterococcus spp. and Staphylococcus spp.. J. Clin. Microbiol.
41: 1135-1142
[Abstract]
[Full Text]
-
Horstkotte, M. A., Knobloch, J. K.-M., Rohde, H., Dobinsky, S., Mack, D.
(2002). Rapid Detection of Methicillin Resistance in Coagulase-Negative Staphylococci with the VITEK 2 System. J. Clin. Microbiol.
40: 3291-3295
[Abstract]
[Full Text]
-
Ligozzi, M., Bernini, C., Bonora, M. G., de Fatima, M., Zuliani, J., Fontana, R.
(2002). Evaluation of the VITEK 2 System for Identification and Antimicrobial Susceptibility Testing of Medically Relevant Gram-Positive Cocci. J. Clin. Microbiol.
40: 1681-1686
[Abstract]
[Full Text]
-
Teng, L.-J., Hsueh, P.-R., Wang, Y.-H., Lin, H.-M., Luh, K.-T., Ho, S.-W.
(2001). Determination of Enterococcus faecalis groESL Full-Length Sequence and Application for Species Identification. J. Clin. Microbiol.
39: 3326-3331
[Abstract]
[Full Text]
-
van den Braak, N., Goessens, W., van Belkum, A., Verbrugh, H. A., Endtz, H. P.
(2001). Accuracy of the VITEK 2 System To Detect Glycopeptide Resistance in Enterococci. J. Clin. Microbiol.
39: 351-353
[Abstract]
[Full Text]