This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowReprints and Permissions
Right arrow Copyright Information
Right arrow Books from ASM Press
Right arrow MicrobeWorld
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Hill, C. E.
Right arrow Articles by Caliendo, A. M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Hill, C. E.
Right arrow Articles by Caliendo, A. M.

 Previous Article  |  Next Article 

Journal of Clinical Microbiology, January 2004, p. 286-289, Vol. 42, No. 1
0095-1137/04/$08.00+0     DOI: 10.1128/JCM.42.1.286-289.2004
Copyright © 2004, American Society for Microbiology. All Rights Reserved.

Assessment of Agreement between the AMPLICOR HIV-1 MONITOR Test Versions 1.0 and 1.5

Charles E. Hill,1 Alicia M. Green,1 Jessica Ingersoll,1 Kirk A. Easley,2 Frederick S. Nolte,1 and Angela M. Caliendo1*

Emory University Hospital and Emory University School of Medicine,1 Biostatistics, Rollins School of Public Health, Emory University, Atlanta, Georgia2

Received 3 April 2003/ Returned for modification 22 June 2003/ Accepted 25 September 2003

The agreement of the microwell plate AMPLICOR HIV-1 MONITOR version 1.0 (MWP 1.0), the microwell plate AMPLICOR HIV-1 MONITOR version 1.5 (MWP 1.5), and the COBAS AMPLICOR HIV-1 MONITOR version 1.5 (COBAS 1.5) tests was evaluated using clinical specimens and well-characterized control material. Two hundred patient plasma specimens and a panel of known human immunodeficiency virus type 1 (HIV-1) subtypes were tested. All data were log10 transformed prior to analysis. The 95% limits of agreement for the three tests at the average of 3.66 log10 copies/ml were ± 0.28 log10, ± 0.34 log10, and ± 0.34 log10 copies/ml for MWP 1.0-MWP 1.5, MWP 1.0-COBAS 1.5, and MWP 1.5-COBAS 1.5, respectively. Ten specimens (6.1%) had differences exceeding the limits of agreement for the MWP 1.0 and MWP 1.5 tests. Correlation coefficients among the three tests were high (r >= 0.96). The viral-load values obtained with the MWP 1.0 test were only 2.1% higher on average than those measured with the MWP 1.5 test and 1.6% higher than those seen with the COBAS 1.5 test. The MWP 1.5 test values were 0.8% higher than the COBAS 1.5 test values. Overall, there was less agreement among the different tests for viral-load values near the lower limit of quantification. The MWP 1.0 test underquantified subtypes A, E, F, G, and H by 1.0 to 2.0 log10 copies/ml; this problem was not observed with the MWP 1.5 test. The close agreement among the results obtained with the different test versions and formats suggests that it is not necessary to reestablish a baseline viral load when changing AMPLICOR HIV-1 MONITOR tests, unless the patient is known to be infected with a non-B subtype.


* Corresponding author. Mailing address: Clinical Laboratories, Emory University Hospital, H180, 1364 Clifton Rd., NE, Atlanta, GA 30322. Phone: (404) 712-5721. Fax: (404) 727-3133. E-mail: acalien{at}emory.edu.


Journal of Clinical Microbiology, January 2004, p. 286-289, Vol. 42, No. 1
0095-1137/04/$08.00+0     DOI: 10.1128/JCM.42.1.286-289.2004
Copyright © 2004, American Society for Microbiology. All Rights Reserved.




This article has been cited by other articles:

  • Perrin, L., Pawlotsky, J. M., Bouvier-Alias, M., Sarrazin, C., Zeuzem, S., Colucci, G. (2006). Multicenter Performance Evaluation of a New TaqMan PCR Assay for Monitoring Human Immunodeficiency Virus RNA Load. J. Clin. Microbiol. 44: 4371-4375 [Abstract] [Full Text]  
  • Salimnia, H., Moore, E. C., Crane, L. R., MacArthur, R. D., Fairfax, M. R. (2005). Discordance between Viral Loads Determined by Roche COBAS AMPLICOR Human Immunodeficiency Virus Type 1 Monitor (Version 1.5) Standard and Ultrasensitive Assays Caused by Freezing Patient Plasma in Centrifuged Becton-Dickinson Vacutainer Brand Plasma Preparation Tubes. J. Clin. Microbiol. 43: 4635-4639 [Abstract] [Full Text]  