Previous Article | Next Article ![]()
Journal of Clinical Microbiology, December 2005, p. 6144-6146, Vol. 43, No. 12
0095-1137/05/$08.00+0 doi:10.1128/JCM.43.12.6144-6146.2005
Copyright © 2005, American Society for Microbiology. All Rights Reserved.
Department of Infectious Diseases,1 Section of Clinical Microbiology, The Cleveland Clinic Foundation, Cleveland, Ohio2
Received 10 August 2005/ Returned for modification 8 September 2005/ Accepted 5 October 2005
We evaluated the Roche LightCycler Staphylococcus MGRADE kits to differentiate between Staphylococcus aureus and coagulase-negative staphylococci in blood cultures growing clusters of gram-positive cocci. Testing 100 bottles (36 containing S. aureus), the assay was 100% sensitive and 98.44% specific for S. aureus and 100% sensitive and specific for coagulase-negative staphylococci.