Previous Article | Next Article ![]()
Journal of Clinical Microbiology, May 2005, p. 2523-2525, Vol. 43, No. 5
0095-1137/05/$08.00+0 doi:10.1128/JCM.43.5.2523-2525.2005
Copyright © 2005, American Society for Microbiology. All Rights Reserved.
Division of Medical Microbiology, The Johns Hopkins Medical Institutions, Meyer B1-193, 600 N. Wolfe St., Baltimore, Maryland 21287
Received 24 September 2004/ Returned for modification 3 November 2004/ Accepted 23 December 2004
Two methods were compared for decreasing bacterial contamination in the BacT/Alert mycobacterial culture detection system. Two concentrations, 0.5 ml (standard amount) and 1.0 ml, of mycobacterial antibiotic supplement were evaluated. Contamination rates were 14% and 6% for the standard and the doubled concentrations, respectively. This difference was statistically significant (P < 0.005).
Copyright © 2009 by the American Society for Microbiology. For an alternate route to Journals.ASM.org, visit: http://intl-journals.asm.org | More Info»