Previous Article | Next Article 
Journal of Clinical Microbiology, November 2007, p. 3616-3619, Vol. 45, No. 11
0095-1137/07/$08.00+0 doi:10.1128/JCM.00221-07
Copyright © 2007, American Society for Microbiology. All Rights Reserved.
Comparative Evaluation of the Automated Roche TaqMan Real-Time Quantitative Human Immunodeficiency Virus Type 1 RNA PCR Assay and the Roche AMPLICOR Version 1.5 Conventional PCR Assay
Anthony R. Oliver,*
Spyridon F. Pereira, and
Duncan A. Clark
Diagnostic Virology (Division of Infection), Barts and the London NHS Trust, Pathology & Pharmacy Building, 80 Newark Street, London E1 2ES, United Kingdom
Received 29 January 2007/
Returned for modification 10 June 2007/
Accepted 25 August 2007
The need to evaluate antiviral treatment response and the emergence of resistance have made the human immunodeficiency virus (HIV) viral load assay a major feature of the diagnostic monitoring of HIV-infected individuals. The objective of this study was to evaluate the utility of the recently In Vitro Diagnostic Medical Devices Directive-approved Roche COBAS AmpliPrep/TaqMan96 real-time PCR assay by comparison with the existing Roche COBAS AmpliPrep/AMPLICOR MONITOR conventional PCR assay. EDTA-treated plasma samples from 191 HIV-1-infected individuals were tested for HIV-1 RNA by the AMPLICOR assay and the TaqMan assay. This was a prospective study using 191 pairs of samples from the same bleed per patient. The correlation coefficient of the assays was 98.08%. The mean difference between the assays was 0.05 log10 copy/ml plasma, with a standard deviation (SD) of 0.27 log10 copy/ml plasma. Thirteen samples gave results with variances greater than 0.5 log10 copy/ml plasma, which is our clinical cutoff. Two samples were more than 3 SD different (0.81 log10 copy/ml plasma). The TaqMan assay appeared to be slightly more sensitive at the lower end of the dynamic range. The assays correlated significantly (P > 0.95) with each other, and the regression analysis was also highly significant (R2 > 0.95).
* Corresponding author. Mailing address: Diagnostic Virology (Division of Infection), Barts and the London NHS Trust, Pathology & Pharmacy Building, 80 Newark St., London E1 2ES, United Kingdom. Phone: 44 203 246 0329. Fax: 44 203 246 0325. E-mail:
tony.oliver{at}bartsandthelondon.nhs.uk
Published ahead of print on 5 September 2007.
Journal of Clinical Microbiology, November 2007, p. 3616-3619, Vol. 45, No. 11
0095-1137/07/$08.00+0 doi:10.1128/JCM.00221-07
Copyright © 2007, American Society for Microbiology. All Rights Reserved.
This article has been cited by other articles:
-
Delaugerre, C., Denis, B., Peytavin, G., Palmer, P., Mourez, T., Le Goff, J., Molina, J.-M., Simon, F.
(2009). Clinical and Resistance Consequences of Misquantification of Plasma and Cerebrospinal Fluid Human Immunodeficiency Virus Type 1 (HIV-1) RNA in Samples from an HIV-1 Subtype G-Infected Patient. J. Clin. Microbiol.
47: 3763-3764
[Abstract]
[Full Text]
-
Scott, L., Carmona, S., Stevens, W.
(2009). Performance of the New Roche Cobas AmpliPrep-Cobas TaqMan Version 2.0 Human Immunodeficiency Virus Type 1 Assay. J. Clin. Microbiol.
47: 3400-3402
[Full Text]
-
Scott, L. E., Noble, L. D., Moloi, J., Erasmus, L., Venter, W. D. F., Stevens, W.
(2009). Evaluation of the Abbott m2000 RealTime Human Immunodeficiency Virus Type 1 (HIV-1) Assay for HIV Load Monitoring in South Africa Compared to the Roche Cobas AmpliPrep-Cobas Amplicor, Roche Cobas AmpliPrep-Cobas TaqMan HIV-1, and BioMerieux NucliSENS EasyQ HIV-1 Assays. J. Clin. Microbiol.
47: 2209-2217
[Abstract]
[Full Text]
-
Sloma, C. R., Germer, J. J., Gerads, T. M., Mandrekar, J. N., Mitchell, P. S., Yao, J. D. C.
(2009). Comparison of the Abbott RealTime Human Immunodeficiency Virus Type 1 (HIV-1) Assay to the Cobas AmpliPrep/Cobas TaqMan HIV-1 Test: Workflow, Reliability, and Direct Costs. J. Clin. Microbiol.
47: 889-895
[Abstract]
[Full Text]
-
Garrido, C., Zahonero, N., Corral, A., Arredondo, M., Soriano, V., de Mendoza, C.
(2009). Correlation between Human Immunodeficiency Virus Type 1 (HIV-1) RNA Measurements Obtained with Dried Blood Spots and Those Obtained with Plasma by Use of Nuclisens EasyQ HIV-1 and Abbott RealTime HIV Load Tests. J. Clin. Microbiol.
47: 1031-1036
[Abstract]
[Full Text]
-
Wei, Y., Yang, C., Wei, B., Huang, J., Wang, L., Meng, S., Zhang, R., Li, J.
(2008). RNase-Resistant Virus-Like Particles Containing Long Chimeric RNA Sequences Produced by Two-Plasmid Coexpression System. J. Clin. Microbiol.
46: 1734-1740
[Abstract]
[Full Text]