This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowReprints and Permissions
Right arrow Copyright Information
Right arrow Books from ASM Press
Right arrow MicrobeWorld
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Wang, X.-P.
Right arrow Articles by Ginocchio, C. C.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Wang, X.-P.
Right arrow Articles by Ginocchio, C. C.

 Previous Article  |  Next Article 

Journal of Clinical Microbiology, May 2009, p. 1546-1548, Vol. 47, No. 5
0095-1137/09/$08.00+0     doi:10.1128/JCM.00114-09
Copyright © 2009, American Society for Microbiology. All Rights Reserved.

Automation of the BD GeneOhm Methicillin-Resistant Staphylococcus aureus Assay for High-Throughput Screening of Nasal Swab Specimens{triangledown}

Xue-Ping Wang and Christine C. Ginocchio*

Molecular Diagnostics, Department of Laboratory Medicine, North Shore—LIJ Health System Laboratories, 10 Nevada Drive, Lake Success, New York 11042

Received 19 January 2009/ Returned for modification 1 March 2009/ Accepted 9 March 2009

This study demonstrated that an automated version of the BD-GeneOhm methicillin-resistant Staphylococcus aureus (MRSA) assay (BD-MRSA), using achromopeptidase sample lysis and PCR setup performed on the Hamilton MICROLAB STARlet (Auto-MRSA), gave results comparable to those obtained with BD-MRSA. The positive- and negative-result concordance rates and overall concordance of BD-MRSA and Auto-MRSA were 98.2, 97.7, and 97.6%, respectively. Auto-MRSA required 60% less technical time than BD-MRSA, and the time required to obtain results was 50% of that required for BD-MRSA. Auto-MRSA provides a reliable, robust system for high-volume clinical use.


* Corresponding author. Mailing address: North Shore—LIJ Health System Laboratories, 10 Nevada Drive, Lake Success, NY 11042. Phone: (516) 719-1079. Fax: (516) 719-1254. E-mail: CGinocch{at}nshs.edu

{triangledown} Published ahead of print on 18 March 2009.


Journal of Clinical Microbiology, May 2009, p. 1546-1548, Vol. 47, No. 5
0095-1137/09/$08.00+0     doi:10.1128/JCM.00114-09