Detection of Significant Bacteriuria by Automated Urinalysis Using Flow Cytometry
ABSTRACT
A new flow cytometry-based automated urine analyzer, the UF-50, was evaluated for its ability to screen urine samples for significant bacteriuria. One hundred eighty-six urine specimens from patients attending an outpatient clinic of a university-based hospital were examined. The results obtained with the UF-50 were compared with those obtained by conventional quantitative urine culture. The UF-50 detected significant bacteriuria with a sensitivity of 83.1%, a specificity of 76.4%, a positive predictive value of 62.0%, a negative predictive value of 90.7%, and an accuracy of 78.5%. These results are comparable to those obtained by previously reported screening procedures. Besides detecting significant bacteriuria, the UF-50 can also perform routine urinalysis, including measurement of concentrations of red blood cells, white blood cells, epithelial cells, and casts, within 70 s. This capability renders this new flow cytometry-based urine analyzer superior to previously reported rapid screening methods.
FOOTNOTES
- Received 3 January 2000.
- Returned for modification 18 March 2000.
- Accepted 29 May 2000.
-
↵* Corresponding author. Mailing address: Department of Urology, Kobe University School of Medicine, 7-5-1, Kusunoki-cho, Chuo-ku, Kobe 650-0017, Japan. Phone: 81 78 382 6155. Fax: 81 78 382 6169. E-mail: okada{at}med.kobe-u.ac.jp.
- Copyright © 2000 American Society for Microbiology











