- TABLE 2
Summary of limit of detection results
- TABLE 3
Clinical performance comparison of four EUA molecular assays for the detection of SARS-CoV-2 in nasopharyngeal swab specimens (n = 104)
↵a The reference standard was defined as the result obtained from at least 3 of the 4 molecular assays.
↵b Kappa values representing levels of agreement are categorized as follows: >0.90, almost perfect; 0.80 to 0.90, strong; 0.60 to 0.79, moderate; 0.40 to 0.59, weak; 0.21 to 0.39, minimal; (0 to 0.20), none.
↵c ± 95% CI, upper/lower 95% confidence interval.
↵d This one sample had cycle threshold (CT) values of 38.9 and 39.6 for N1 and N2, respectively, by the CDC assay.
↵e Cycle threshold (CT) value undetermined.
↵f These two samples had cycle threshold (CT) values of 36.2 and 38.5 by the Hologic assay.
- TABLE 4
Details of discordant sample analysis
↵a Discordant sample results are highlighted in bold. CT, cycle threshold; ID, identifier; NEG, negative; POS, positive.
- TABLE 5
Throughput and workflow evaluation for four EUA molecular SARS-CoV-2 assays
↵a NA, not applicable.